proGAV® FV448TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PROGAV SYS PED.W/SA 20 A.BURRH.RES. · Class 2 · Implant
All models of proGAV (28)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04041906132682), received between Jan 26, 2021 and Apr 11, 2024 — classified as 8 injuries.
The most frequently reported specific device problem is “Infusion or Flow Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem6
- Mechanical Problem3
- Break1
Patient problems
- Hydrocephalus8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryApr 11, 2024Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (#FV448T) WAS IMPLANTED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 14 YEARS. HEIGHT: 150.5 CM. WEIGHT: 45 KG. GENDER: FEMALE. SAME EVENT AS MEDWATCH# 3004721439-2024-00084.
Patient problems: Hydrocephalus
MAUDE report 19086828 ↗InjuryFeb 22, 2024Break; Break
IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (#FV448T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE ABDOMINAL CATHETER IS RUPTURED BEHIND THE SHUNT. THE CATHETER WAS REMOVED AND A NEW CATHETER WAS CONNECTED AND IMPLANTED. THE SHUNT SYSTEM IS STILL IMPLANTED. A PART OF THE COMPLAINED CATHETER WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 3 YEARS, 9 MONTHS. HEIGHT: 103.5 CM. WEIGHT: 16 KG. GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 18762663 ↗InjuryNov 22, 2022Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV SHUNT SYSTEM (PART # FV448T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2011. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 12 YEARS AND 3 MONTHS. HEIGHT: 165 CENTIMETERS (CM). WEIGHT: 61 KILOGRAMS (KG). GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 15849820 ↗InjuryMar 29, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (PART # FV448T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE AND DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON AN UNKNOWN DATE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), HEIGHT: 160 CENTIMETERS (CM), WEIGHT: (B)(6), GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 13944001 ↗InjuryDec 3, 2021Mechanical Problem; Mechanical Problem
ACCORDING TO THE COMPLAINANT A PROGAV OPERATED IN OVERDRAINAGE POSTOPERATIVELY . THE VALVE WAS IMPLANTED ON (B)(6) 2015. IT WAS REMOVED ON (B)(6) 2017 DUE TO THE MALFUNCTION AND RETURNED TO THE MANUFACTURER. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: 120 CM.
Patient problems: Hydrocephalus
MAUDE report 12927878 ↗InjuryOct 13, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FV448T - PROGAV SYS PED.W/SA 20 A.BURRH.RES. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OVERDRAINAGE. THE VALVE WAS EXPLANTED AND RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED.
Patient problems: Hydrocephalus
MAUDE report 12627632 ↗InjuryOct 6, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED TO MIETHKE THAT A PROGAV SYS PED.W/SA 20 A.BURRH.RES. (PART # FV448T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2009. ACCORDING TO THE COMPLAINANT, THE VALVE WAS SUSPECTED OF OPERATING IN OVER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2016. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). WEIGHT: (B)(6). HEIGHT: 149 CENTIMETERS (CM) GENDER: UNKNOWN.
Patient problems: Hydrocephalus
MAUDE report 12586803 ↗InjuryJan 26, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.