proSA® FV706TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PROSA VALVE W/DIST.CATHETER · Class 2 · Implant
All models of proSA (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04041906133559), received between Feb 9, 2022 and Aug 22, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Damaged Prior to Use” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Damaged Prior to Use1
- Mechanical Problem1
Patient problems
- Hydrocephalus2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 22, 2024Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT THE STERILIZATION PACKAGING HAD A POOR SEALING WHEN UNPACKING AND REFUSED TO USE THE DEVICE. THE DEVICE/PACKAGING WAS NOT RETURNED TO THE MANUFACTURER. PICTURES WITH THE PROBLEM WAS SUBMITTED.
Patient problems: Hydrocephalus
MAUDE report 20030805 ↗InjuryFeb 9, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A PROSA VALVE (PART # FV706T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, APPROXIMATELY THREE (3) YEARS AFTER IMPLANTATION, THE PHYSICIAN ATTEMPTED TO ADJUST THE PRESSURE WITH THE PRESSURE REGULATING TOOL, BUT THE VALVE PRESSURE FAILED TO CHANGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 13502144 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page