proGAV 2.0® FX411TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PROGAV 2.0 WITH SHUNTASSISTANT 10 · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04041906136536), received between Nov 10, 2021 and Dec 22, 2022 — classified as 3 injuries.
The most frequently reported specific device problem is “Infusion or Flow Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Infusion or Flow Problem3
Patient problems
- Hydrocephalus3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 22, 2022Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 (PART # FX411T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNSPECIFIED DATE IN 2019. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATING IN OVER-DRAINAGE AND WAS NON-ADJUSTABLE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 16038066 ↗InjuryDec 21, 2022Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 (PART#: FX411T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNSPECIFIED DATE IN 2019. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATING IN OVER-DRAINAGE AND WAS NON-ADJUSTABLE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 16029922 ↗InjuryNov 10, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED MIETHKE THAT A PROGAV 2.0 WITH SHUNTASSISTANT 10 (PART # FX411T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE PATIENT RETURNED FOR SURGERY APPROXIMATELY TWELVE (12) MONTHS AFTER THE VALVE WAS IMPLANTED. THE PHYSICIAN SUSPECTED A DISTAL OR PROXIMAL OCCLUSION. REPORTEDLY, A MANOMETER WAS USED ON EACH SIDE OF THE VALVE, THE VALVE WAS THEN CHECKED WITH A CHECKMATE TO CONFIRM ITS SETTING WAS AT TEN (10). ONCE THE SETTING WAS CONFIRMED, THE PHYSICIAN INJECTED SALINE, WHICH REVEALED THAT THE PROGAV ALLOWED COMPLETE FLOW THROUGH WITHOUT THE SETTING HAVING ANY AFFECT ON FLOW. AT THIS TIME, THE DECISION WAS MADE TO REPLACE THE VALVE. THE EXISTING RESERVOIR WAS LEFT IN PLACE AND A NEW VALVE (PART # FX412T) WAS IMPLANTED AT A SETTING OF TEN (10). THE PHYSICIAN THEN ABLE TO SUCCESSFULLY PERFORM A FLOW RETEST. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 12787580 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page