Skip to content

proGAV 2.0® FX412TRecalls & reported problems

Christoph Miethke GmbH & Co. KG · PROGAV 2.0 WITH SHUNTASSISTANT 15 · Class 2 · Implant

All models of proGAV 2.0 (49)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04041906136543), received between Aug 19, 2022 and Nov 7, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1
  • Obstruction of Flow1
  • Physical Resistance/Sticking1

Patient problems

  • Hydrocephalus2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2024Obstruction of Flow; Physical Resistance/Sticking; Obstruction of Flow; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX412T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2024. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. AGE: 75 YEARS. GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 20632372 ↗
  • MalfunctionAug 19, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (PART # FX412T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6)2022. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED CHANGE THE PRESSURE SETTING DURING A MRI. THIS PRODUCT WAS IMPLANTED IN A PATIENT AT A SET PRESSURE OF 12 CENTIMETERS WATER (CMH2O). AFTER TRANSFER, THE PATIENT SUFFERED FROM ANOTHER DISEASE AND WHEN AN MRI SCAN WAS PERFORMED, IT WAS CONFIRMED THAT THE PRESSURE SETTING HAD BEEN CHANGED TO AROUND 6 CMH2O. THERE IS NO MENTION OF PRESSURE CHANGE IN THE PATIENT'S MEDICAL RECORD, AND THE DOCTOR AT THE FACILITY ALSO STATED THAT NO PRESSURE CHANGE WAS MADE. THE PATIENT WAS UNDER-DRAINAGE AND THE PRESSURE SETTING NEEDED TO BE LOWERED, SO IT DID NOT CAUSE ANY SERIOUS PROBLEMS. THE SHUNTSYSTEM IS STILL IMPLANTED.

    Patient problems: Hydrocephalus