proGAV 2.0® FX413TRecalls & reported problems

Christoph Miethke GmbH & Co. KG · PROGAV 2.0 WITH SHUNTASSISTANT 20 · Class 2 · Implant

Quick summary

This Christoph Miethke proGAV 2.0 shunt is an implanted valve that drains fluid from the brain in patients with hydrocephalus. The 57 reports mostly describe the valve draining too little or too much fluid, plus flow blockages and valves that stick or are hard to adjust; several patients needed revision surgery in 2025. There are 57 injuries and no deaths. No recalls name this device, and reports dropped from 12 in the previous 12 months to 1 in the latest12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
57
Deaths reported
0
Injuries reported
57
Recalls naming it
0

Summary

The FDA's MAUDE database contains 57 adverse event reports that list this exact device (DI 04041906136550), received between Sep 17, 2020 and Jan 20, 2026 — classified as 57 injuries.

The most frequently reported specific device problem is “Infusion or Flow Problem” (37 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem37
  • Mechanical Problem20
  • Obstruction of Flow8
  • Physical Resistance/Sticking6
  • Insufficient Device Problem Information2
  • Complete Blockage1

Patient problems

  • Hydrocephalus56
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 57 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 20, 2026Obstruction of Flow

    IT WAS REPORTED THAT A PROGAV 2.0 (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE SHUNT CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. SAME EVENT/ PATIENT AS # 3004721439-2026-00012.

    Patient problems: Hydrocephalus

    MAUDE report 24114902 ↗
  • InjuryOct 20, 2025Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2015. ACCORDING TO THE COMPLAINANT, THE ABDOMINAL CATHETER WAS CALCIFIED AND A OVER-DRAINAGE WAS DETERMINED. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 23336056 ↗
  • InjuryAug 29, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2015. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 22921121 ↗
  • InjuryJul 11, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 22478282 ↗
  • InjuryJul 3, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE SHUNT CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 22395116 ↗
  • InjuryJun 23, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED ON (B)(6) 2025. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 22308902 ↗
  • InjuryMar 31, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 21730340 ↗
  • InjuryMar 17, 2025Obstruction of Flow; Physical Resistance/Sticking; Obstruction of Flow; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX423T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2017. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 21617877 ↗
  • InjuryFeb 24, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 (#FX413T) WAS IMPLANTED AS A PART OF A SHUNTSYSTEM ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 21447925 ↗
  • InjuryFeb 10, 2025Obstruction of Flow; Physical Resistance/Sticking; Obstruction of Flow; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 21344636 ↗
  • InjuryJan 15, 2025Mechanical Problem; Obstruction of Flow; Insufficient Device Problem Information; Mechanical Problem; Obstruction of Flow; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX432T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED A MALFUNCTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS NOT YET SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 21157478 ↗
  • InjuryDec 9, 2024Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2017. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2024. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 20877601 ↗
  • InjuryNov 7, 2024Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED AN OVER-DRAINAGE AND ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 20632448 ↗
  • InjuryAug 27, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX442T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 8 YEARS, 1 MONTH. WEIGHT: 34 KG. HEIGHT: 124 CM. GENDER: FEMALE. SAME EVENT AS # 3004721439-2024-00265.

    Patient problems: Hydrocephalus

    MAUDE report 20074707 ↗
  • InjuryAug 7, 2024Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 6 YEARS. HEIGHT: 104 CM. WEIGHT: 15 KG.

    Patient problems: Hydrocephalus

    MAUDE report 19930392 ↗
  • InjuryJul 3, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 1 YEAR, 7 MONTHS. HEIGHT: 76 CM. WEIGHT: 9 KG. GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 19670538 ↗
  • InjuryJun 26, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED AN OVER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 8 YEARS,3 MONTHS WEIGHT: 28 KG HEIGHT: 120 CM GENDER: FEMALE

    Patient problems: Hydrocephalus

    MAUDE report 19613517 ↗
  • InjuryNov 20, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON UNKNOWN. ACCORDING TO THE COMPLAINANT, THE SHUNT SHOWED AN OVER-DRAINAGE AND ADJUSTMENT PROBLEMS. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON UNKNOWN. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 13 YEARS. WEIGHT: 68 KGS. HEIGHT: 140 CM. GENDER: FEMALE. SAME EVENT AS # 3004721439-2023-00351.

    Patient problems: Hydrocephalus

    MAUDE report 18172967 ↗
  • InjuryNov 14, 2023Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 5 YEARS, HEIGHT: 106 CM, WEIGHT: 21.4 KG, GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 18130478 ↗
  • InjuryNov 9, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE SHUNT SHOWED A OVER DRAINAGE AND ADJUSTMENT PROBLEMS. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINANT DEVICE HAS NOT YET RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 12 MONTHS. WEIGHT: 8.2 KILOGRAMS. HEIGHT: 72 CENTIMETERS. GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 18104101 ↗
  • InjuryNov 2, 2023Infusion or Flow Problem; Infusion or Flow Problem

    CORRECTION: THE PROGAV 2.0 SHUNT SYSTEM (FX413T) SHOULD BE IMPLANTED ON (B)(6) 2023. A DELAY OCCURRED BECAUSE THE SHUNT DID NOT DRAIN AS DESIRED. A REPLACEMENT PRODUCT FROM ANOTHER MANUFACTURER WAS USED AND THE SURGERY WAS COMPLETED. NO IMPAIRMENT OF THE PATIENT'S CONDITION IS KNOWN. THE COMPLAINED PRODUCT WAS RETURNED FOR EVALUATION TO THE MANUFACTURER.

    Patient problems: Hydrocephalus

    MAUDE report 18056230 ↗
  • InjuryNov 2, 2023Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN (B)(6) 2019. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 9 YEARS. HEIGHT: 78 CENTIMETERS. WEIGHT: 10,8 KILOGRAMS. GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 18055894 ↗
  • InjuryNov 1, 2023Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SHOWED A UNDER-DRAINAGE AND BLOCKAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINANT DEVICE HAS NOT YET RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 6 MONTHS. WEIGHT: 10.4 KILOGRAMS. HEIGHT: 82.7 CENTIMETERS. GENDER: MALE. SAME EVENT AS # 3004721439-2023-00312 AND # 3004721439-2023-00313.

    Patient problems: Hydrocephalus

    MAUDE report 18049209 ↗
  • InjuryAug 18, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (MATERIAL #FX413T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, FUNCTIONALITY GLITCHES WERE EXPERIENCED. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED. THE COMPLAINT DEVICE HAS NOT YET RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 17578046 ↗
  • InjuryMay 26, 2023Infusion or Flow Problem; Infusion or Flow Problem

    PATIENT INFORMATION: AGE: 74 YEARS. WEIGHT: 67 KG. HEIGHT: 156 CM.

    Patient problems: Hydrocephalus

    MAUDE report 17014540 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →