proGAV 2.0® FX424TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PROGAV 2.0 SYSTEM WITH SPRUNG RESERVOIR · Class 2 · Implant
All models of proGAV 2.0 (49)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04041906136949), received between Aug 17, 2021 and May 4, 2026 — classified as 8 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem6
- Infusion or Flow Problem3
- Fluid/Blood Leak1
- Material Integrity Problem1
- Obstruction of Flow1
- Physical Resistance/Sticking1
Patient problems
- Hydrocephalus9
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 4, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A VETRICULAR CATHETER PART OF A PROGAV 2.0 SHUNT SYSTEM (#FX424T) TORE DURING PRE-SURGERY PREPARATIONS. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25070709 ↗MalfunctionMar 23, 2026Mechanical Problem
IT WAS REPORTED THAT A PROGAV SHUNTSYSTEM (#FV424T) WAS TO BE IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2026. ACCORDING TO THE COMPLAINANT, THE RESERVOIR COULD NOT BE PRIMED AND PRESSED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: Hydrocephalus
MAUDE report 24662273 ↗InjuryFeb 10, 2026Material Integrity Problem
IT WAS REPORTED THAT A PERITONEAL CATHETER FROM A PROGAV 2.0 SHUNT SYSTEM (#FX424T) BROKE INTRAOPERATIVELY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.
Patient problems: Hydrocephalus
MAUDE report 24303636 ↗InjuryAug 11, 2025Obstruction of Flow; Physical Resistance/Sticking; Obstruction of Flow; Physical Resistance/Sticking
IT WAS REPORTED THAT A PROGAV 2.0 (#FX424T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. SAME INCIDENT AS MEDWATCH#3004721439-2025-00228.
Patient problems: Hydrocephalus
MAUDE report 22773359 ↗InjuryJun 10, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A PROGAV 2.0 SYSTEM (#FX424T) WAS IMPLANTED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WILL BE RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 71 YEARS, GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 19499679 ↗InjuryFeb 7, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 (PART # FX424T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATING IN OVER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. REFERENCE ASSOCIATED MEDWATCH REPORT MDR ID # 3004721439-2023-00011.
Patient problems: Hydrocephalus
MAUDE report 16312461 ↗InjurySep 22, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 (PART # FX424T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 7 MONTHS. HEIGHT: 94 CENTIMETERS (CM). WEIGHT: 11 KILOGRAMS (KG). GENDER: MALE. SAME EVENT AS MEDWATCH# 3004721439-2022-00281.
Patient problems: Hydrocephalus
MAUDE report 15465097 ↗InjuryFeb 17, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO AESCULAP AG THAT A PROGAV 2.0 SHUNT SYSTEM (PART # FX424T) WAS IMPLANTED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE SURGEON WAS UNABLE TO ADJUST THE VALVE. REPORTEDLY, THE SURGEON PLACED THE VALVE IN A VERY DIFFICULT (MASTOID BONE) LOCATION, THEREFORE, IT WAS DIFFICULT TO LOCATE AND ADJUST THE VALVE. THE PATIENT FELT "BAD" AND THE ADJUSTMENT WAS WORSENING ACCORDING TO THE X-RAY. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 13552808 ↗InjuryNov 4, 2021Mechanical Problem; Mechanical Problem
ACCORDING TO THE COMPLAINANT A PROGAV WAS BLOCKED POSTOPERATIVELY. THE PATIENT WAS ORIGINALLY IMPLANTED ON (B)(6) 2017; 18 DAYS LATER, THE MALFUNCTION WAS NOTED. THE VALVE WAS EXPLANTED ON (B)(6) 2017 AND RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED.
Patient problems: Hydrocephalus
MAUDE report 12752517 ↗InjuryAug 17, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 (#FX417T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO HAVE A BLOCKAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. NO FURTHER PATIENT DATA ARE AVAILABLE. THE SAME EVENT: MEDWATCH# 3004721439-2021-00222.
Patient problems: Hydrocephalus
MAUDE report 12331155 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page