proGAV 2.0® FX426TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PROGAV 2.0 SYS W/SA15 A.SPRUNG RESERVOIR · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04041906136963), received between Mar 18, 2021 and Dec 21, 2022 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Infusion or Flow Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Infusion or Flow Problem4
- Fluid/Blood Leak1
- Mechanical Problem1
- Obstruction of Flow1
- Optical Obstruction1
Patient problems
- Hydrocephalus6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 21, 2022Fluid/Blood Leak; Infusion or Flow Problem; Fluid/Blood Leak; Infusion or Flow Problem
ACCORDING TO THE COMPLAINANT, PRIOR TO USE, A LEAK ON THE RESERVOIR SIDE OF THE PROGAV 2.0 SHUNTSYSTEM (PART # FX426T) WAS OBSERVED. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED. AGE: UNKNOWN. WEIGHT: 172 KILOGRAMS (KG). HEIGHT: 80 CENTIMETERS (CM). GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 16028916 ↗InjuryDec 6, 2021Obstruction of Flow; Optical Obstruction; Obstruction of Flow; Optical Obstruction
IT WAS REPORTED THAT A PROGAV 2.0 SYS W/SA15 A.SPRUNG RESERVOIR (PART # FX426T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE RESERVOIR WAS NOTED AS BEING FULL OF STEAM. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 12934733 ↗InjuryNov 30, 2021Mechanical Problem; Mechanical Problem
ACCORDING TO THE COMPLAINANT A PROGAV WAS BLOCKED POSTOPERATIVELY. THE PATIENT WAS IMPLANTED ON (B)(6) 2015. ONCE THE PRODUCT WAS IMPLANTED, PRESSURE COULD NOT BE ADJUSTED. HOWEVER, THE PRODUCT WAS REMAINED SINCE THE PATIENT CONDITION WAS GOOD. THE SURGERY FOR BRAIN TUMOR HAD BEEN DONE TWICE SINCE THE PRIMARY SURGERY WITHOUT VALVE FAILURE. HOWEVER THIS TIME, THE REVISION SURGERY WAS NEEDED DUE TO UNDERDRAINAGE. THEREFORE, THE VALVE INVESTIGATION INCLUDED ADJUSTMENT FAILURE WAS REQUESTED. THE VALVE WAS EXPLANTED ON (B)(6) RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: 172 CM.
Patient problems: Hydrocephalus
MAUDE report 12892808 ↗InjuryAug 19, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (PART #FX426T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO HAVE A BLOCKAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), HEIGHT: (B)(6) CENTIMETERS (CM), WEIGHT: (B)(6) KILOGRAMS (KG), GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 12341775 ↗InjuryMar 18, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT THERE WAS A PROBLEM WITH A SPRUNG RESERVOIR TO THE PROGAV 2.0 SHUNTSYSTEM. ACCORDING TO THE COMPLAINANT,WHEN SHUNT WAS ASSEMBLED ON THE STERILE TABLE THE PHYSICIAN ASSISTANT WAS UNABLE TO PRIME THE SHUNT USING THE CONTROL RESERVOIR. THE PATIENT WAS UNDER ANESTHESIA, THERE WAS NO DELAY IN SURGERY, AND THE SURGEON HAD A BACK-UP SHUNT TO USE INSTEAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED AS OF THIS REPORT. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE.
Patient problems: Hydrocephalus
MAUDE report 11517407 ↗InjuryMar 18, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED TO AESCULAP THAT PROGAV 2.0 SYS W/SA15 A.SPRUNG RESERVOIR WAS USED DURING SHUNT PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, WHEN SHUNT WAS ASSEMBLED ON THE STERILE TABLE THE PHYSICIAN ASSISTANT WAS UNABLE TO PRIME THE SHUNT USING THE CONTROL RESERVOIR. FURTHER INFORMATION WAS RECEIVED ON FEBRUARY 24, 2021 WHICH CONFIRMED THAT THE PATIENT WAS UNDER ANESTHESIA, THERE WAS NO DELAY IN SURGERY, AND THE SURGEON HAD A BACK-UP SHUNT TO USE INSTEAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED AS OF THIS REPORT. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: Hydrocephalus
MAUDE report 11517229 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page