proGAV 2.0® FX432TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PROGAV 2.0 SYS W/SA10 A.CONTROL RESERV. · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04041906137304), received between Sep 21, 2021 and May 20, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Loss of or Failure to Bond2
- Obstruction of Flow2
- Inflation Problem1
- Use of Device Problem1
Patient problems
- Hydrocephalus5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 20, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX432T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNSPECIFIED DATE IN 2023. ACCORDING TO THE COMPLAINANT, THE PATIENT'S STATE OF HEALTH DETERIORATED AND THE VALVE WAS REVISED AS A PRECAUTIONARY MEASURE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS NOT YET RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: Hydrocephalus
MAUDE report 22052700 ↗InjuryJan 13, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT A PROGAV 2.0 (#FX432T) WAS IMPLANTED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2024. THE COMPLAINED PRODUCT WILL BE RETURNED TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 21136966 ↗MalfunctionSep 16, 2024Inflation Problem; Use of Device Problem; Inflation Problem; Use of Device Problem
THE PRODUCT WAS RETURNED IN DRY CONDITION TO THE MANUFACTURER FOR EXAMINATION. IT WAS REPORTED THAT THE DEVICE (#FX432T) COULD NOT BE USED DUE TO THE FACHT THAT THE AIR IN THE RESERVOIR COULD NOT BE EXPELLED BY PUMPING. A REPLACEMENT PRODUCT WAS USED AND THE OPERATION WAS COMPLETED. NO COMPLICATIONS WERE REPORTED ON THE PATIENT.
Patient problems: Hydrocephalus
MAUDE report 20225094 ↗InjurySep 21, 2021Loss of or Failure to Bond; Break; Loss of or Failure to Bond; Break
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX432T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, THE POSTERIOR CATHETER WAS DISCONNECTED. THE PATIENT RETURNED SOMETIME LATER AND COMPLAINED THAT SHE WAS "NOT FEELING RIGHT". A SHUNT SERIES WAS ORDERED AND RESULTS SHOWED THAT THE DISTAL CATHETER WAS FRACTURED. WHEN THE SURGEON EXAMINED THE PATIENT IN THE OPERATING ROOM, THE SHUNT WAS EXPOSED AND THE POSTERIOR CATHETER WAS NOTED TO BE COMPLETELY DISCONNECTED (BELOW THE REINFORCED SECTION DISTAL TO SA). AFTER FINDING THE DISTAL BREAK, THE SYSTEM INCLUDING RESERVOIR, PROGAV, AND SHUNT ASSISTANT WAS EXPLANTED. ANOTHER MIETHKE PRODUCT, FX817T WAS IMPLANTED INSTEAD. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. NO FURTHER PATIENT INFORMATION ARE AVAILABLE.
Patient problems: Hydrocephalus
MAUDE report 12509670 ↗InjurySep 21, 2021Loss of or Failure to Bond; Break; Loss of or Failure to Bond; Break
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX432T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, THE POSTERIOR CATHETER WAS DISCONNECTED. THE PATIENT RETURNED SOMETIME LATER AND COMPLAINED THAT SHE WAS "NOT FEELING RIGHT". A SHUNT SERIES WAS ORDERED AND RESULTS SHOWED THAT THE DISTAL CATHETER WAS FRACTURED. WHEN THE SURGEON EXAMINED THE PATIENT IN THE OPERATING ROOM, THE SHUNT WAS EXPOSED AND THE POSTERIOR CATHETER WAS NOTED TO BE COMPLETELY DISCONNECTED (BELOW THE REINFORCED SECTION DISTAL TO SA). AFTER FINDING THE DISTAL BREAK, THE SYSTEM INCLUDING RESERVOIR, PROGAV, AND SHUNT ASSISTANT WAS EXPLANTED. ANOTHER MIETHKE PRODUCT, FX817T WAS IMPLANTED INSTEAD. THE PATIENT UNDERWENT A REVISION PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. NO FURTHER PATIENT INFROMATION ARE AVAILABLE.
Patient problems: Hydrocephalus
MAUDE report 12509623 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026