The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04041906137335), received between Mar 11, 2021 and Jun 23, 2025 — classified as 9 injuries, 1 malfunction.
The most frequently reported specific device problem is “Infusion or Flow Problem” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem7
- Mechanical Problem5
- Loss of or Failure to Bond1
- Material Integrity Problem1
- Obstruction of Flow1
Patient problems
- Hydrocephalus10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJun 23, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX435T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2018. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 22309092 ↗InjurySep 7, 2022Loss of or Failure to Bond; Material Integrity Problem; Loss of or Failure to Bond; Material Integrity Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (PART # FX410T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE PATIENT GOT WORSE IN (B)(6) 2022. IN THE X-RAY WAS VERIFIED THE VALVE RUPTURE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 15374803 ↗InjuryDec 3, 2021Mechanical Problem; Mechanical Problem
ACCORDING TO THE COMPLAINANT A PROGAV WAS BLOCKED POSTOPERATIVELY. THE VALVE WAS IMPLANTED ON (B)(6) 2017. IT WAS REMOVED ON (B)(6) 2017 DUE TO THE MALFUNCTION AND RETURNED TO THE MANUFACTURER. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: 170 CM.
Patient problems: Hydrocephalus
MAUDE report 12927956 ↗InjuryDec 2, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SYS W/SA25 A.CONTROL RESERV. (#FX435T) WAS IMPLANTED DURING THE PROCEDURE ON (B)(6) 2005. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE DISFUNCTION. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2017 THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. FURTHER DETAILS WERE NOT PROVIDED. HEIGHT: (B)(6) .
Patient problems: Hydrocephalus
MAUDE report 12918744 ↗MalfunctionNov 4, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH FX435T -PROGAV 2.0 SYS W/SA25 A.CONTROL RESERV. ACCORDING TO THE COMPLAINANT, WHEN PHYSICIAN PUSH THE RESERVOIR, IT WAS DIFFICULT FOR SALINE TO COME OUT AND THE CATHETER WAS REPLACED WITH THE DIFFERENT PRODUCT. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: Hydrocephalus
MAUDE report 12751400 ↗InjuryOct 29, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SYS W/SA25 A.CONTROL RESERV..(#FX435T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE BLOCKED POSTOPERATIVELY. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2017. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: Hydrocephalus
MAUDE report 12724199 ↗InjuryAug 26, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATING IN OVERDRAINAGE AND WAS DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6), 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). WEIGHT: (B)(6). HEIGHT: 168 CENTIMETERS (CM). GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 12375052 ↗InjuryMay 4, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (#FX435T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE . ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE CLOGGED. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). GENDER: FEMALE. INVOLVED COMPONENT: 3004721439-2021-00130 (FX800T - M.BLUE 0 VALVE ).
Patient problems: Hydrocephalus
MAUDE report 11770640 ↗InjuryApr 30, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX435T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, THE SHUNTSYSTEM WAS BELIEVED TO BE OPERATED IN UNDER-DRAINAGE. THE VENTRICLE OF THE PATIENT WERE DILATED. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), WEIGHT: (B)(6), HEIGHT: 173 CM, GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 11752869 ↗InjuryMar 11, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗