Skip to content

proGAV 2.0® FX438TRecalls & reported problems

Christoph Miethke GmbH & Co. KG · PROGAV 2.0 SYS PED.WITH PRECHAMBER · Class 2 · Implant

All models of proGAV 2.0 (49)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04041906137434), received between Feb 2, 2018 and Oct 20, 2025 — classified as 8 injuries.

The most frequently reported specific device problem is “Infusion or Flow Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem6
  • Physical Resistance/Sticking3
  • Mechanical Problem2
  • Insufficient Device Problem Information1
  • Obstruction of Flow1
  • Partial Blockage1

Patient problems

  • Hydrocephalus8
  • No Known Impact Or Consequence To Patient1
  • Obstruction/Occlusion1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 20, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER AND THE INVESTIGATION HAS BEEN COMPLETED.

    Patient problems: Hydrocephalus

    MAUDE report 23335443 ↗
  • InjuryAug 4, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 (#FX438T) WAS IMPLANTED TOGETHER WITH AN SHUNTASSISTANT (FV413-T) DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCTS WERE SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. SAME INCIDENT AS MEDWATCH #3004721439-2025-00220.

    Patient problems: Hydrocephalus

    MAUDE report 22698255 ↗
  • InjuryMay 1, 2025Obstruction of Flow; Insufficient Device Problem Information; Physical Resistance/Sticking; Obstruction of Flow; Insufficient Device Problem Information; Physical Resistance/Sticking

    IT WAS REPORTED THAT A PROGAV 2.0 VALVE (#FX438T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN UNDER-DRAINAGE AND HAD ADJUSTMENT PROBLEMS. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS NOW SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 21941692 ↗
  • InjuryApr 28, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 (#FX438T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SHOWED AN OVER-DRAINAGE AND ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 7 MONTHS. WEIGHT: 14 KILOGRAMS (KG). HEIGHT: 93.5 CENTIMETERS (CM). GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 16838793 ↗
  • InjuryJul 14, 2022Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 (#FX438T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE BLOCKED. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 3 YEARS. GENDER: FEMALE. SAME EVENT AS MEDWATCH#3004721439-2022-00200.

    Patient problems: Hydrocephalus

    MAUDE report 15020903 ↗
  • InjuryNov 16, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (PART # FX438T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATING IN OVER AND UNDERDRAINAGE AND WAS NOT ADJUSTABLE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6) YEARS, HEIGHT: 180 CENTIMETERS (CM), WEIGHT: (B)(6) KILOGRAMS (KG), GENDER: MALE. ITS THE SAME PATIENT AS # 3004721439-2021-00309.

    Patient problems: Hydrocephalus

    MAUDE report 12823995 ↗
  • InjuryMar 30, 2021Infusion or Flow Problem; Infusion or Flow Problem

    ADDITIONAL INFOFMATION: AGE: 15 MONTHS HEIGHT: 58 CM WEIGHT: 5.3 KG.

    Patient problems: Hydrocephalus; Obstruction/Occlusion

    MAUDE report 11589357 ↗
  • InjuryFeb 2, 2018Partial Blockage; Infusion or Flow Problem; Partial Blockage; Infusion or Flow Problem

    COUNTRY OF COMPLAINT: (B)(6). IT WAS REPORTED THAT THERE IS BLOCKAGE IN THE VALVE.

    Patient problems: Hydrocephalus; No Known Impact Or Consequence To Patient

    MAUDE report 7236118 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates