proGAV 2.0® FX438TRecalls & reported problems
Christoph Miethke GmbH & Co. KG · PROGAV 2.0 SYS PED.WITH PRECHAMBER · Class 2 · Implant
All models of proGAV 2.0 (49)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04041906137434), received between Feb 2, 2018 and Oct 20, 2025 — classified as 8 injuries.
The most frequently reported specific device problem is “Infusion or Flow Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem6
- Physical Resistance/Sticking3
- Mechanical Problem2
- Insufficient Device Problem Information1
- Obstruction of Flow1
- Partial Blockage1
Patient problems
- Hydrocephalus8
- No Known Impact Or Consequence To Patient1
- Obstruction/Occlusion1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryOct 20, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking
THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER AND THE INVESTIGATION HAS BEEN COMPLETED.
Patient problems: Hydrocephalus
MAUDE report 23335443 ↗InjuryAug 4, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT A PROGAV 2.0 (#FX438T) WAS IMPLANTED TOGETHER WITH AN SHUNTASSISTANT (FV413-T) DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCTS WERE SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. SAME INCIDENT AS MEDWATCH #3004721439-2025-00220.
Patient problems: Hydrocephalus
MAUDE report 22698255 ↗InjuryMay 1, 2025Obstruction of Flow; Insufficient Device Problem Information; Physical Resistance/Sticking; Obstruction of Flow; Insufficient Device Problem Information; Physical Resistance/Sticking
IT WAS REPORTED THAT A PROGAV 2.0 VALVE (#FX438T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN UNDER-DRAINAGE AND HAD ADJUSTMENT PROBLEMS. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS NOW SENT TO THE MANUFACTURER FOR INVESTIGATION.
Patient problems: Hydrocephalus
MAUDE report 21941692 ↗InjuryApr 28, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 (#FX438T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SHOWED AN OVER-DRAINAGE AND ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 7 MONTHS. WEIGHT: 14 KILOGRAMS (KG). HEIGHT: 93.5 CENTIMETERS (CM). GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 16838793 ↗InjuryJul 14, 2022Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 (#FX438T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE BLOCKED. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 3 YEARS. GENDER: FEMALE. SAME EVENT AS MEDWATCH#3004721439-2022-00200.
Patient problems: Hydrocephalus
MAUDE report 15020903 ↗InjuryNov 16, 2021Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (PART # FX438T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATING IN OVER AND UNDERDRAINAGE AND WAS NOT ADJUSTABLE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6) YEARS, HEIGHT: 180 CENTIMETERS (CM), WEIGHT: (B)(6) KILOGRAMS (KG), GENDER: MALE. ITS THE SAME PATIENT AS # 3004721439-2021-00309.
Patient problems: Hydrocephalus
MAUDE report 12823995 ↗InjuryMar 30, 2021Infusion or Flow Problem; Infusion or Flow Problem
ADDITIONAL INFOFMATION: AGE: 15 MONTHS HEIGHT: 58 CM WEIGHT: 5.3 KG.
Patient problems: Hydrocephalus; Obstruction/Occlusion
MAUDE report 11589357 ↗InjuryFeb 2, 2018Partial Blockage; Infusion or Flow Problem; Partial Blockage; Infusion or Flow Problem
COUNTRY OF COMPLAINT: (B)(6). IT WAS REPORTED THAT THERE IS BLOCKAGE IN THE VALVE.
Patient problems: Hydrocephalus; No Known Impact Or Consequence To Patient
MAUDE report 7236118 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page