PROGAV 2.0 FX385TRecalls & reported problems
Aesculap AG · PROGAV 2.0 W/SA2.0 20 & SPRUNG RESERVOIR · Class 2 · Implant
All models of PROGAV 2.0 (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04041906502461), received between Feb 28, 2023 and Jun 15, 2026 — classified as 4 injuries.
The most frequently reported specific device problem is “Calibration Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Calibration Problem1
- Infusion or Flow Problem1
- Mechanical Problem1
- Obstruction of Flow1
- Physical Resistance/Sticking1
Patient problems
- Hydrocephalus3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 15, 2026Mechanical Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNTSYSTEM (#FX385T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE PROGAV 2.0 VALVE HAD UNINTENDED SHUNT ADJUSTMENT. 4 YEARS AGO, THE PATIENT HAVE A PREVIOUSLY SELF- ADJUSTED VALVE, TOO. THE NEUROSURGEON SUSPECTED THAT THE PATIENT ADJUSTED THE SHUNT HIMSELF. THE PATIENT UNDERWENT A REVISION PROCEDURE AND A NEW SINGLE PRESSURE VALVE WAS IMPLANTED. THE SHUNT IS STILL IN TRANSIT TO MANUFACTURER FOR EXAMINATION.
Patient problems: Hydrocephalus
MAUDE report 25505506 ↗InjuryFeb 19, 2025Obstruction of Flow; Physical Resistance/Sticking; Obstruction of Flow; Physical Resistance/Sticking
THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.
Patient problems: Hydrocephalus
MAUDE report 21410548 ↗InjuryMar 8, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (PART # FX385T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, APPROXIMATELY TWO (2) DAYS AFTER BEING IMPLANTED, THE VALVE SETTING WAS NOT ABLE TO BE REPROGRAMMED. THE SURGEON MADE SEVERAL UNSUCCESSFUL ATTEMPTS TO PERFORM THE ADJUSTMENT WITH THE SALES REPRESENTATIVE PRESENT. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINT DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 16507353 ↗InjuryFeb 28, 2023Calibration Problem; Calibration Problem
IT WAS REPORTED TO AESCULAP INC. THAT A PROGAV 2.0 W/SA2.0 20 & SPRUNG RESERVOIR (PART # FX385T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, APPROXIMATELY TWO (2) DAYS AFTER BEING IMPLANTED, THE VALVE SETTING WAS NOT ABLE TO BE REPROGRAMMED. THE SURGEON MADE SEVERAL UNSUCCESSFUL ATTEMPTS TO PERFORM THE ADJUSTMENT WITH THE SALES REPRESENTATIVE PRESENT. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINT DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Insufficient Information
MAUDE report 16453360 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page