Skip to content

PROGAV 2.0 FX667TRecalls & reported problems

Aesculap AG · PROGAV 2.0 W/SA2.0 25 & CONTROL RESERV. · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04041906503116), received between Apr 14, 2022 and Jun 7, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Infusion or Flow Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem4
  • Mechanical Problem1

Patient problems

  • Hydrocephalus4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 7, 2024Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 (# FX667T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. (B)(6).

    Patient problems: Hydrocephalus

    MAUDE report 19489180 ↗
  • InjuryFeb 7, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM(#FX667T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDER-DRAINAGE WITHIN 24 HOURS. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 21YEARS. HEIGHT: 165 CM. WEIGHT: 60 KG. GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 18658258 ↗
  • InjuryDec 21, 2022Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A SHUNTASSISTANT 2.0 (PART # FX667T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATE IN UNDER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 8 YEARS, 5 MONTHS, HEIGHT: 130 CENTIMETERS (CM), WEIGHT: 32 KILOGRAMS (KG), GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 16030154 ↗
  • InjuryApr 14, 2022Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A PROGAV 2.0 SHUNT SYSTEM (PART # (B)(4)) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON AN UNKNOWN DATE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). HEIGHT: 130 CENTIMETERS (CM). WEIGHT: (B)(6) (KG). GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 14112289 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates