M.BLUE FX816TRecalls & reported problems
Aesculap AG · M.BLUE 5 SYS W/PED.CONTROL RESERVOIR · Class 2 · Implant
All models of M.BLUE (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Aesculap M.Blue FX816T valve is an implanted shunt component used to drain fluid from the brain in patients with hydrocephalus. The 28 reports mostly describe the valve draining too much or too little, being hard to adjust, sticking, or becoming infected, and in every case the patient had a revision procedure; there are 28 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 28
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 04041906504526), received between Aug 26, 2021 and Mar 17, 2025 — classified as 28 injuries.
The most frequently reported specific device problem is “Infusion or Flow Problem” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem24
- Mechanical Problem15
- Insufficient Device Problem Information1
- Obstruction of Flow1
- Patient Device Interaction Problem1
- Physical Resistance/Sticking1
Patient problems
- Hydrocephalus28
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
InjuryMar 17, 2025Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT AN M. BLUE VALVE (#FX816T) WAS IMPLANTED ON AN UNSPECIFIED DATE. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.
Patient problems: Hydrocephalus
MAUDE report 21618113 ↗InjuryFeb 3, 2025Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2024. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 21294795 ↗InjuryDec 9, 2024Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT AN M. BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6)2024. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6)2024. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: Hydrocephalus
MAUDE report 20877939 ↗InjuryJul 1, 2024Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M. BLUE (#FX817T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WILL BE RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 1 YEAR, 11 MONTHS HEIGHT: 70 CM WEIGHT: 8 KG GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 19644122 ↗InjuryMay 28, 2024Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 4 YEARS, HEIGHT: 103 CM, WEIGHT: 19,5 KG, GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 19404663 ↗InjuryMay 16, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A M.BLUE SYSTEM (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS REMOVED DUE AN INFECTION. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 14 YEARS WEIGHT: 77 KG GENDER: FEMALE
Patient problems: Hydrocephalus
MAUDE report 19336825 ↗InjuryMar 18, 2024Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE VALVE SHOWED AN UNDER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 1 YEARS, 4 MONTHS. WEIGHT: 10 KGS. HEIGHT: 80 CM. GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 18925248 ↗InjuryMar 15, 2024Infusion or Flow Problem; Insufficient Device Problem Information; Infusion or Flow Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED A MALFUNCTION. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WILL BE RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 18 MONTHS GENDER: FEMALE
Patient problems: Hydrocephalus
MAUDE report 18911891 ↗InjuryNov 14, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 1 YEAR, 9 MONTHS, HEIGHT: 90 CM, WEIGHT: 11.5 KG, GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 18130535 ↗InjuryJul 21, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M. BLUE (PART#: FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2019. ACCORDING TO THE COMPLAINANT, THE VALVE CAUSED AN UNDER- DRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. WHILE REMOVING, THE LID WAS LOOSENED FROM THE VALVE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 4 YEARS, 7 MONTHS, GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 17375289 ↗InjuryApr 14, 2023Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE SYSTEM WAS BELIEVED TO BE OPERATED IN OVER-DRAINAGE AND NON-ADJUSTABLE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 65 YEARS WEIGHT: 63 KILOGRAMS (KG) HEIGHT: 171 CENTIMETERS (CM) GENDER: UNKNOWN.
Patient problems: Hydrocephalus
MAUDE report 16746094 ↗InjuryMar 17, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 8 MONTHS, HEIGHT: 79 CENTIMETERS (CM), WEIGHT: 8.1 KILOGRAMS (KG). GENDER: MALE
Patient problems: Hydrocephalus
MAUDE report 16568697 ↗InjuryMar 9, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE SHUNT SYSTEM (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2023. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 1 YEAR, 6 MONTH, HEIGHT: 80 CENTIMETERS (CM), WEIGHT: 9 KILOGRAMS (KG), GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 16518400 ↗InjuryFeb 3, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATED IN UNDER-DRAINAG. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 16299560 ↗InjuryJan 6, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 13 YEARS, HEIGHT: 170 CENTIMETERS (CM), WEIGHT: 74 KILOGRAMS (KG), GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 16118328 ↗InjuryJan 4, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE AND NON-ADJUSTABILITY. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 16098785 ↗InjuryJan 4, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 16098776 ↗InjuryNov 16, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 2 MONTHS. HEIGHT: 84.5 CENTIMETERS (CM). WEIGHT: 9.8 KILOGRAMS (KG). GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 15809328 ↗InjurySep 29, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE VALVE (#FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 13 MONTHS, HEIGHT: 81 CENTIMETERS (CM), WEIGHT: 10.8 KILOGRAMS (KG), GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 15506222 ↗InjuryAug 10, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE VALVE WAS BELIEVED TO BE NOT ADJUSTABLE. THE PATIENT WILL BE REVIEWED AGAIN IN SIX (6) WEEKS. AT THIS REVIEW, THE SURGEON WILL DECIDE WHETHER THE PRODUCT SHOULD BE EXPLANTED OR NOT.
Patient problems: Hydrocephalus
MAUDE report 15204404 ↗InjuryJul 12, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 14992877 ↗InjuryMar 17, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN OVERDRAINAGE AND WAS DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 13793401 ↗InjuryFeb 10, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATED IN UNDERDRAINAGE AND WAS DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). HEIGHT: 95 CENTIMETERS (CM). WEIGHT: (B)(6). GENDER: MALE.
Patient problems: Hydrocephalus
MAUDE report 13509616 ↗InjuryJan 28, 2022Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX816T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE BLOCKED. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2021. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6). HEIGHT: UNKNOWN. WEIGHT: UNKNOWN. GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 13388678 ↗InjuryJan 6, 2022Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M. BLUE PLUS (# (B)(4)) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNSPECIFIED DATE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM WAS BELIEVED TO BE OPERATING IN UNDERDRAINAGE AND DIFFICULT TO ADJUST. THE PATIENT UNDERWENT A REVISION PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: (B)(6), HEIGHT: UNKNOWN, WEIGHT: (B)(6) KG, GENDER: FEMALE.
Patient problems: Hydrocephalus
MAUDE report 13176959 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page