M.BLUE FX817TRecalls & reported problems
Aesculap AG · M.BLUE 10 SYS W/PED.CONTROL RESERVOIR · Class 2 · Implant
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04041906504533), received between Apr 5, 2021 and Apr 19, 2023 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Infusion or Flow Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Infusion or Flow Problem3
- Mechanical Problem2
Patient problems
- Hydrocephalus4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryApr 19, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED TO AESCULAP INC. THAT A M.BLUE 10 SYS W/PED.CONTROL RESERVOIR (PRODUCT # FX817T) WAS IMPLANTED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE DEVICE WAS BELIEVED TO HAVE A BLOCKAGE. THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2023. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE XC (B)(4).
Patient problems: Hydrocephalus
MAUDE report 16770406 ↗InjuryApr 19, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE SHUNT SYSTEM (PART # FX647T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE SYSTEM WAS BELIEVED TO BE OPERATED IN BLOCKAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 3023. THE COMPLAINT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 16769893 ↗MalfunctionJul 20, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX817T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE MBLUE10 SHUNT HAD INADVERTENTLY REPROGRAMMED. IN EARLY JUNE, THE VALVE WAS ADJUSTED TO EIGHT (8). WHEN THE PATIENT RETURNED ON (B)(6) 2021, THE VALVE VERIFIED AT 16 (SIXTEEN). THE VALVE WAS ADJUSTED BACK TO EIGHT (8) AND THE SETTING WAS VERIFIED WITH BOTH COMPASSES. ON (B)(6) 2022 THE PATIENT RETURNED FOR FOLLOW-UP AND THE VALVE SETTING REMAINED AT EIGHT (8), WHICH WAS THE PRIOR SETTING. NO FURTHER ADJUSTMENTS WERE MADE DURING THIS VISIT. THE PATIENT WAS SCHEDULED TO UNDERGO A T WAS TO UNDERGO A COMPUTED TOMOGRAPHY (CT) SCAN AND SHUNTOGRAM. THE COMPLAINT DEVICE IS STILL IMPLANTED AND NOT YET AVAILABLE FOR FURTHER EVALUATION. THE PATIENT IS SCHEDULED FOR CLINICAL FOLLOW-UP.
Patient problems: Hydrocephalus
MAUDE report 15066852 ↗InjuryJul 20, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A M.BLUE (PART # FX817T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE MBLUE10 SHUNT HAD INADVERTENTLY REPROGRAMMED. IN EARLY (B)(6), THE VALVE WAS ADJUSTED TO EIGHT (8). WHEN THE PATIENT RETURNED ON (B)(6) 2021, THE VALVE VERIFIED AT 16 (SIXTEEN). THE VALVE WAS ADJUSTED BACK TO EIGHT (8) AND THE SETTING WAS VERIFIED WITH BOTH COMPASSES. ON (B)(6) 2022 THE PATIENT RETURNED FOR FOLLOW-UP AND THE VALVE SETTING REMAINED AT EIGHT (8), WHICH WAS THE PRIOR SETTING. NO FURTHER ADJUSTMENTS WERE MADE DURING THIS VISIT. THE PATIENT WAS SCHEDULED TO UNDERGO A COMPUTED TOMOGRAPHY (CT) SCAN AND SHUNTOGRAM. THE COMPLAINT DEVICE IS STILL IMPLANTED AND NOT YET AVAILABLE FOR FURTHER EVALUATION. THE PATIENT IS SCHEDULED FOR CLINICAL FOLLOW-UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15066821 ↗InjuryApr 5, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT A M.BLUE SHUNTSYSTEM WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE VALVE IS REPORTED TO HAVE SHOWN BLOCKAGE DURING THE USE. THE PATIENT UNDERWENT A REVISION PROCEDURE.
Patient problems: Hydrocephalus
MAUDE report 11616230 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.