Skip to content

M.BLUE PLUS FX824TRecalls & reported problems

Aesculap AG · M.BLUE PLUS SYS W/CONTROL RESERVOIR · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Aesculap M.Blue Plus FX824T is an implanted brain shunt system with a control reservoir, used to drain fluid in patients with hydrocephalus. Most of the 51 reports describe drainage flow problems — fluid moving too little or too much, sticking or resisting adjustment, and a catheter that broke off during preparation — often leading to a revision surgery; 48 reports describe injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
51
Deaths reported
0
Injuries reported
48
Recalls naming it
0

Summary

The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 04041906504830), received between May 10, 2021 and Apr 20, 2026 — classified as 48 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Infusion or Flow Problem” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 5 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 4 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem36
  • Mechanical Problem10
  • Obstruction of Flow7
  • Physical Resistance/Sticking7
  • Insufficient Device Problem Information2
  • Defective Component1
  • Human-Device Interface Problem1
  • Material Integrity Problem1

Patient problems

  • Hydrocephalus47
  • No Clinical Signs, Symptoms or Conditions2
  • Failure of Implant1
  • Pain1

Recent adverse event reports

The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 20, 2026Obstruction of Flow

    IT WAS REPORTED THAT A M.BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 24931653 ↗
  • MalfunctionApr 20, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT A M.BLUE PLUS (#FX824T) WAS NOT ADJUSTABLE BEFORE USE. THE PRODUCT IS STILL IN THE STERILE PACKAGING AND NO PATIENT INJURY WAS REPORTED AS A RESULT OF TEH MALFUNCTION. THE SHUNT WAS RETURNED FOR EXAMNIATION TO THE MANUFACTURER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24927157 ↗
  • MalfunctionApr 13, 2026Material Integrity Problem

    IT WAS REPORTED THAT DURING THE PREPARATION THE VENTRICULAR CATHETER (PART OF #FX824T) WAS BROKEN OFF. THE COMPLAINED PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. ANOTHER SHUNT WAS OPENED AND NO DELAY OR OTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24865581 ↗
  • InjuryApr 13, 2026Obstruction of Flow

    IT WAS REPORTED THAT AN M. BLUE PLUS SHUNTSYSTEM (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2026. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2026. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 24860760 ↗
  • InjuryJan 20, 2026Obstruction of Flow; Physical Resistance/Sticking

    IT WAS REPORTED THAT A M. BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2025. ACCORDING TO THE COMPLAINANT, THE SHUNT CAUSED UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE.

    Patient problems: Hydrocephalus

    MAUDE report 24114817 ↗
  • InjuryDec 8, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT AN M.BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNSPECIFIED DATE. THE USER'S INITIAL REPORT DOES NOT INDICATE ANY PRODUCT MALFUNCTION AS THE REASON FOR EXPLANTATION. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 23741999 ↗
  • InjuryAug 25, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT AN M.BLUE PLUS(#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 22882422 ↗
  • InjuryJul 28, 2025Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A M.BLUE VALVE (PART OF AN M.BLUE PLUS SYSTEM) (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON AN UNSPECIFIED DATE. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 22629013 ↗
  • InjuryJul 21, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT AN M. BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 22554184 ↗
  • InjuryMay 12, 2025Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT AN M. BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 22001448 ↗
  • InjuryMar 31, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT AN M. BLUE PLUS(#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2025. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 21730337 ↗
  • InjuryMar 31, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT A M. BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2025. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 21730081 ↗
  • InjuryMar 21, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT AN M. BLUE PLUS (#FX824T) WAS IMPLANTED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 21662389 ↗
  • InjuryMar 10, 2025Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED TO AESCULAP INC. THAT A M. BLUE PLUS SYS W/CONTROL RESERVOIR (PART # FX824T) WAS IMPLANTED ON AN UNSPECIFIED DATE. ACCORDING TO THE COMPLAINANT, THE PATIENT'S VALVE WAS NOT FUNCTIONING PROPERLY. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2024, AND THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE RETURN KIT QUESTIONNAIRE CONTAINED ADDITIONAL INFORMATION. VALVE OVER-DRAINAGE WAS NOTED PRIOR TO THE REVISION. NO HOUSING DAMAGE WAS OBSERVED POST-REVISION, BUT THE VALVE ADJUSTMENT WAS NOT TESTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4).

    Patient problems: Failure of Implant

    MAUDE report 21561497 ↗
  • InjuryMar 7, 2025Infusion or Flow Problem; Insufficient Device Problem Information; Infusion or Flow Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT A M. BLUE PLUS (#FX824T) WAS IMPLANTED. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED A OVER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6)2024. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 21546310 ↗
  • InjuryFeb 24, 2025Obstruction of Flow; Physical Resistance/Sticking; Obstruction of Flow; Physical Resistance/Sticking

    IT WAS REPORTED THAT AN M.BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.

    Patient problems: Hydrocephalus

    MAUDE report 21449707 ↗
  • InjuryDec 9, 2024Infusion or Flow Problem; Physical Resistance/Sticking; Infusion or Flow Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT AN M. BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN OVERDRAINAGE AND ADJUSTMENT DIFFICULTIES THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2024. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. THE COMPLAINED PRODUCT WAS SENT TO THE MANUFACTURER FOR INVESTIGATION.

    Patient problems: Hydrocephalus

    MAUDE report 20878001 ↗
  • InjuryOct 21, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT AN M. BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED AN UNDER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 6 MONTHS WEIGHT: 9 KGS HEIGHT: 80 CM GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 20492574 ↗
  • InjurySep 27, 2024Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A M. BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM CAUSED AN UNDERDRAINAGE AND HAD ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE. THE COMPLAINANT DEVICE WILL BE RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 6 YEARS, 3 MONTHS HEIGHT: 110 CM WEIGHT: 18 KG GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 20320897 ↗
  • InjuryAug 27, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT AN M.BLUE PLUS(#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2024. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED AN UNDER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 65 YEARS GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 20074718 ↗
  • InjuryAug 23, 2024Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT AN M.BLUE PLUS(#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2022. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED AN OVER-DRAINAGE AND ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 2 YEARS, 5 MONTHS WEIGHT: 8 KGS HEIGHT: UNKNOWN. GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 20048582 ↗
  • InjuryJul 25, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT AN M.BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 65 YEARS. WEIGHT: 54 KGS. HEIGHT: 160 CM. GENDER: FEMALE.

    Patient problems: Hydrocephalus

    MAUDE report 19827676 ↗
  • InjuryJul 3, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT AN M.BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED AN OVER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 41 YEARS. WEIGHT: 60 KGS. HEIGHT: 170 CM. GENDER: MALE.

    Patient problems: Hydrocephalus

    MAUDE report 19670488 ↗
  • InjuryMar 18, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT A M.BLUE PLUS (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN 2023. ACCORDING TO THE COMPLAINANT, THE SHUNT SYSTEM SHOWED AN UNDER-DRAINAGE. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 11 MONTHS.

    Patient problems: Hydrocephalus

    MAUDE report 18925322 ↗
  • InjuryFeb 22, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT A M.BLUE (#FX824T) WAS IMPLANTED DURING A PROCEDURE PERFORMED IN (B)(6) 2022. ACCORDING TO THE COMPLAINANT, THE VALVE SHOWED ADJUSTMENT DIFFICULTIES. THE PATIENT UNDERWENT A REVISION PROCEDURE PERFORMED ON (B)(6) 2024. THE COMPLAINANT DEVICE HAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION PROCEDURE. AGE: 77 YEARS WEIGHT: 85 KILOGRAMS HEIGHT: 180 CENTIMETER GENDER: UNKNOWN

    Patient problems: Hydrocephalus

    MAUDE report 18760344 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page