Skip to content

Screw tension device 10033700Recalls & reported problems

MAQUET GmbH · Class 1

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046768000247), received between Apr 5, 2024 and Apr 5, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack1
  • Detachment of Device or Device Component1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2024Crack; Detachment of Device or Device Component; Unintended Movement; Crack; Detachment of Device or Device Component; Unintended Movement

    ON 26TH FEBRUARY 2024, GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR ACCESSORIES - (B)(6) - SCREW TENSION DEVICE USED WITH 141901JC - EXTENSION DEVICE. AS IT WAS STATED AND CONFIRMED WITH THE PHOTOGRAPHIC EVIDENCE, CRACKS IN SCREW FIXING PARTS OF THE HOUSING WERE DISCOVERED. ON (B)(6) 2024, ADDITIONAL INFORMATION WAS PROVIDED. IT HAS BEEN CONFIRMED THAT THE ISSUE OCCURRED DURING HIP JOINT DISPLACEMENT SURGERY ON AN ANESTHETIZED PATIENT. AS IT WAS STATED, THE ACCESSORY DETACHED COMPLETELY LEADING TO SLOW CHANGE IN PATIENT POSITION THAT WAS UNDER STAFF CONTROL. THE ISSUE RESULTED IN A DELAY IN PROCEDURE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATIONS, NAMELY THE DETACHMENT OF THE ACCESSORY LEADING TO CHANGE IN THE PATIENT POSITION AND DELAY IN SURGERY RESULTING IN PROLONGED ANESTHESIA TIME, WERE TO REOCCUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19048417 ↗

Other TABLE AND ATTACHMENTS, OPERATING-ROOM devices

All TABLE AND ATTACHMENTS, OPERATING-ROOM devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.