Screw tension device 10033700Recalls & reported problems
MAQUET GmbH · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046768000247), received between Apr 5, 2024 and Apr 5, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Crack” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
- Detachment of Device or Device Component1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2024Crack; Detachment of Device or Device Component; Unintended Movement; Crack; Detachment of Device or Device Component; Unintended Movement
ON 26TH FEBRUARY 2024, GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF OUR ACCESSORIES - (B)(6) - SCREW TENSION DEVICE USED WITH 141901JC - EXTENSION DEVICE. AS IT WAS STATED AND CONFIRMED WITH THE PHOTOGRAPHIC EVIDENCE, CRACKS IN SCREW FIXING PARTS OF THE HOUSING WERE DISCOVERED. ON (B)(6) 2024, ADDITIONAL INFORMATION WAS PROVIDED. IT HAS BEEN CONFIRMED THAT THE ISSUE OCCURRED DURING HIP JOINT DISPLACEMENT SURGERY ON AN ANESTHETIZED PATIENT. AS IT WAS STATED, THE ACCESSORY DETACHED COMPLETELY LEADING TO SLOW CHANGE IN PATIENT POSITION THAT WAS UNDER STAFF CONTROL. THE ISSUE RESULTED IN A DELAY IN PROCEDURE. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL FOR SERIOUS INJURY IF THE SITUATIONS, NAMELY THE DETACHMENT OF THE ACCESSORY LEADING TO CHANGE IN THE PATIENT POSITION AND DELAY IN SURGERY RESULTING IN PROLONGED ANESTHESIA TIME, WERE TO REOCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19048417 ↗
Other TABLE AND ATTACHMENTS, OPERATING-ROOM devices
- Arthrex® AR-1747-BArthrex, Inc. · TABLE AND ATTACHMENTS, OPERATING-ROOM12 reports
- Arthrex® AR-1644Arthrex, Inc. · Table and attachments, operating-room5 reports
- Leg holder 100380A0MAQUET GmbH · TABLE AND ATTACHMENTS, OPERATING-ROOM4 reports
- Arthrex® AR-1560-02Arthrex, Inc. · TABLE AND ATTACHMENTS, OPERATING-ROOM4 reports
- Arthrex® AR-1500Arthrex, Inc. · TABLE AND ATTACHMENTS, OPERATING-ROOM3 reports
- Pair of leg plates 113373BCMAQUET GmbH · TABLE AND ATTACHMENTS, OPERATING-ROOM3 reports
- Kneeling frame 100704D0MAQUET GmbH · TABLE AND ATTACHMENTS, OPERATING-ROOM2 reports
- Traction boot, adult 10018800MAQUET GmbH · TABLE AND ATTACHMENTS, OPERATING-ROOM2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.