The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04046826385491), received between Nov 9, 2018 and Jan 7, 2019 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fracture3
  • Device Handling Problem2

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Failure of Implant1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2019Fracture; Fracture

    THE ORTHODONTIC SCREW BROKE AFTER SUCCESSFUL IMPLANTATION WHEN THE USER ATTEMPTED TO REMOVE THE IMPLANTATION SLEEVE USING THE DEDICATED SCREWDRIVER. HOWEVER, THE SLEEVE WAS FIRMLY LODGED ON THE SCREW HEAD, THE SCREWDRIVER BLADE BROKE, AND ANOTHER INSTRUMENT WAS REQUIRED TO FORCIBLY REMOVE THE SLEEVE.

    Patient problems: Failure of Implant

    MAUDE report 8221831 ↗
  • MalfunctionNov 9, 2018Fracture; Device Handling Problem; Fracture; Device Handling Problem

    THE ORTHODONTIC SCREW BROKE AS IT WAS BEING INSERTED INTO THE PATIENT'S JAW. THE BROKEN PIECE WAS REMOVED IMMEDIATELY. NO HARM TO PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8055194 ↗
  • MalfunctionNov 9, 2018Fracture; Device Handling Problem; Fracture; Device Handling Problem

    THE ORTHODONTIC SCREW BROKE AS IT WAS BEING INSERTED INTO THE PATIENT'S JAW. THE BROKEN PIECE WAS REMOVED IMMEDIATELY. NO HARM TO PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient