SURECAN SAFETY II 4447006-02Recalls & reported problems
B. Braun Medical Inc. · Surecan™ Safety II Port Access Needle Set 20G x 0.8 in. (20mm) · Class 2
All models of SURECAN SAFETY II (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04046955156276), received between May 17, 2018 and Dec 29, 2021 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Protective Measures Problem” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Protective Measures Problem7
- Fail-Safe Problem5
- Device Handling Problem1
Patient problems
- No Known Impact Or Consequence To Patient5
- Needle Stick/Puncture4
- No Consequences Or Impact To Patient4
- Exposure to Body Fluids1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 29, 2021Device Handling Problem; Device Handling Problem
AS REPORTED BY THE USER FACILITY: 2 NEEDLESTICKS OCCURRED WITH STAFF- ONE WITH CLEAN NEEDLE WHILE TRYING TO REATTACH THE BASEPLATE. ONE WAS WITH NEEDLE THAT HAD BEEN ACCESSED. IT IS REPORTED THAT NURSE STUCK HERSELF WHEN TRYING TO READJUST THE NEEDLE. IT IS UNKNOWN IF THE NEEDLE WAS CLEAN OR HAD ALREADY BEEN USED AT THE TIME OF INJURY.
Patient problems: Needle Stick/Puncture
MAUDE report 13112631 ↗MalfunctionDec 8, 2021Fail-Safe Problem; Fail-Safe Problem
AS REPORTED BY THE USER FACILITY: THE SURECAN NEEDLE DID NOT LOCK IN PLACE RESULTING IN NEEDLESTICK FOR RN.
Patient problems: Needle Stick/Puncture
MAUDE report 12963139 ↗MalfunctionMay 11, 2020Protective Measures Problem; Protective Measures Problem
AS REPORTED BY USER FACILITY: AFTER DEACCESSING A PORT-A-CATH, THE PORT NEEDLE SAFETY ENGAGED, BUT DID NOT FULLY LOCK INTO PLACE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10043332 ↗MalfunctionApr 15, 2020Protective Measures Problem; Protective Measures Problem
AS REPORTED BY USER FACILITY: NEEDLE REMOVED FROM PATIENT AND SAFETY FEATURE MALFUNCTIONED AND CAME OFF DURING REMOVAL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9961810 ↗MalfunctionOct 22, 2019Protective Measures Problem; Protective Measures Problem
AS REPORTED BY USER FACILITY: WHEN REMOVED FROM THE PORT THE SAFETY DID NOT ENGAGE. NO INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9219525 ↗MalfunctionMay 17, 2018Fail-Safe Problem
AS REPORTED BY USER FACILITY: EVENT 1 WHEN THE TECHNICIAN WAS REMOVING THE HUBER NEEDLE FROM THE PATIENT, THE SAFETY MECHANISM DID NOT ACTIVATE. THE SAFETY MECHANISM SLIDES TO THE BOTTOM OF THE NEEDLE BUT DOESN'T ENGAGE. NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7522517 ↗MalfunctionMay 17, 2018Fail-Safe Problem
AS REPORTED BY USER FACILITY: UPON DE-ACCESSING A PATIENT'S PORT, SURECAN SAFETY DID NOT ENGAGE AND CLINICIAN OBTAINED A NEEDLE STICK INJURY. CLINICIAN AND THE SOURCE PATIENT WERE BOTH TESTED FOR (B)(6). CURRENT STATUS OF THE CLINICIAN IS HEALTHY.
Patient problems: Exposure to Body Fluids; Needle Stick/Puncture
MAUDE report 7522515 ↗MalfunctionMay 17, 2018Protective Measures Problem; Protective Measures Problem
AS REPORTED BY USER FACILITY: EVENT 1: WHEN THE TECHNICIAN WAS REMOVING THE HUBER NEEDLE FROM THE PATIENT, THE SAFETY MECHANISM DID NOT ACTIVATE. THE SAFETY MECHANISM SLIDES TO THE BOTTOM OF THE NEEDLE BUT DOESN'T ENGAGE. NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7522513 ↗MalfunctionMay 17, 2018Protective Measures Problem; Protective Measures Problem
AS REPORTED BY USER FACILITY: EVENT 3: WHEN THE TECHNICIAN WAS REMOVING THE HUBER NEEDLE FROM THE PATIENT, THE SAFETY MECHANISM DID NOT ACTIVATE. THE SAFETY MECHANISM SLIDES TO THE BOTTOM OF THE NEEDLE BUT DOESN'T ENGAGE. NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7522508 ↗MalfunctionMay 17, 2018Fail-Safe Problem
AS REPORTED BY USER FACILITY: EVENT 3: WHEN THE TECHNICIAN WAS REMOVING THE HUBER NEEDLE FROM THE PATIENT, THE SAFETY MECHANISM DID NOT ACTIVATE. THE SAFETY MECHANISM SLIDES TO THE BOTTOM OF THE NEEDLE BUT DOESN'T ENGAGE. NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7522503 ↗MalfunctionMay 17, 2018Protective Measures Problem; Protective Measures Problem
AS REPORTED BY USER FACILITY: EVENT 2 WHEN THE TECHNICIAN WAS REMOVING THE HUBER NEEDLE FROM THE PATIENT, THE SAFETY MECHANISM DID NOT ACTIVATE. THE SAFETY MECHANISM SLIDES TO THE BOTTOM OF THE NEEDLE BUT DOESN'T ENGAGE. NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7522502 ↗MalfunctionMay 17, 2018Fail-Safe Problem
AS REPORTED BY USER FACILITY: EVENT 2: WHEN THE TECHNICIAN WAS REMOVING THE HUBER NEEDLE FROM THE PATIENT, THE SAFETY MECHANISM DID NOT ACTIVATE. THE SAFETY MECHANISM SLIDES TO THE BOTTOM OF THE NEEDLE BUT DOESN'T ENGAGE. NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7522498 ↗MalfunctionMay 17, 2018Protective Measures Problem; Protective Measures Problem
AS REPORTED BY USER FACILITY: UPON DE-ACCESSING A PATIENT'S PORT, SURECAN SAFETY DID NOT ENGAGE AND CLINICIAN OBTAINED A NEEDLE STICK INJURY. CLINICIAN AND THE SOURCE PATIENT WERE BOTH TESTED FOR (B)(6 AND (B)(6). CURRENT STATUS OF CLINICIAN IS HEALTHY.
Patient problems: Needle Stick/Puncture
MAUDE report 7522496 ↗
Other Non-Coring (Huber) Needle devices
- SafeStep CPA00049Bard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- SafeStep CPA00040Bard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- SafeStep CPA00042ABard Access Systems, Inc. · Non-Coring (Huber) Needle1 recall0 reports
- INTERA AP-07014USIntera Oncology Inc. · Non-Coring (Huber) Needle37 reports
- SURECAN SAFETY II 4447005-02B. Braun Medical Inc. · Non-Coring (Huber) Needle6 reports
- AllPoints Port Access System SafeStep AN152075MBard Access Systems, Inc. · Non-coring (Huber) needle5 reports
- SURECAN SAFETY II 4447012-02B. Braun Medical Inc. · Non-Coring (Huber) Needle4 reports
- MEDLINE HNX20075Medline Industries, Inc. · Non-coring (Huber) needle4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page