Skip to content

APEX® 601237Recalls & reported problems

B. Braun Medical Inc. · APEX, SCDM CONTROL MODULE, WIN-10 · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046955589524), received between Jun 5, 2024 and Jul 9, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Application Program Problem: Dose Calculation Error” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem: Dose Calculation Error2
  • Application Program Freezes or Fails to Launch1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 9, 2025Application Program Problem: Dose Calculation Error; Application Program Freezes or Fails to Launch; Application Program Problem: Dose Calculation Error; Application Program Freezes or Fails to Launch

    AS REPORTED BY THE USER FACILITY: HEALTH PROFESSIONAL TEAM NOTICED THAT THERE HAVE BEEN A HANDFUL OF ORDERS BEING PUMPED AT BBMUS # (B)(4) THAT ARE HAVING A DISCREPANCY OF EXACTLY 13MLS FOR LINE 1 (STERILE WATER) OF WHAT IS ORDERED AND WHAT IS PUMPED. HEALTH PROFESSIONAL WAS INFORMED THAT AN ERROR OCCURRED ON THAT PUMP A LITTLE AFTER LUNCH (1PM) THAT CAUSED THEM TO RESTART THE PUMP. THE TIME STAMPS OF SOME IMPACTED BAGS ARE 2:34PM AND 3:08PM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22461190 ↗
  • MalfunctionJun 5, 2024Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTED THAT UPON COMPOUNDING IT IS NOTED CERTAIN STATIONS RUNNING DRY AND SHOWING A NEGATIVE VOLUME ON SCREEN. NO NOTIFICATIONS SEEN TO CHANGE A SOLUTION. THIS HAS BEEN AN INTERMITTENT ISSUE INCLUDING THE FOLLOWING OCCURENCES: THIS OCCURS ON STATION 1 WATER WHEN BY ITSELF AND WHEN Y LINKED TO STATION 2. THIS OCCURS ON STATION 22 PLENAMINE WHEN BY ITSELF AND WHEN LINKED TO STATION 23. STATIONS RUN DRY PULLING BUBBLES WITHOUT ANY ALERTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19467398 ↗

Other System/device, pharmacy compounding devices

All System/device, pharmacy compounding devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.