AS Columbus® REVISION F NR071ZRecalls & reported problems
Aesculap AG · AS COLUMBUS REV F TIB.OFFSET CEMENT.T1 · Class 2 · Implant
All models of AS Columbus REVISION F (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04046963175948), received between Oct 11, 2021 and Aug 28, 2025 — classified as 6 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Device Appears to Trigger Rejection2
- Patient-Device Incompatibility2
Patient problems
- Failure of Implant3
- No Clinical Signs, Symptoms or Conditions2
- Implant Pain1
- Metal Related Pathology1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryAug 28, 2025Patient-Device Incompatibility; Patient-Device Incompatibility
THE PATIENT UNDERWENT A LEFT TOTAL KNEE REPLACEMENT ON (B)(6) 2023.
Patient problems: Implant Pain; Metal Related Pathology
MAUDE report 22913044 ↗InjuryApr 2, 2025Patient-Device Incompatibility; Patient-Device Incompatibility
IT WAS REPORTED TO AESCULAP INC. THAT AN AS COLUMBUS REV F TIB.OFFSET CEMENT.T1 (PART# NR071Z) WAS IMPLANTED ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT A SURGERY TO REPLACE THE POLYETHYLENE COMPONENTS HAD BEEN PLANNED FOR (B)(6) 2025; SIZES LARGER THAN 20 MILLIMETER (MM) NEEDED. THE PROCEDURE WOULD BE REVISION RIGHT KNEE TOTAL ARTHROPLASTY (TKA). ADDITIONAL INFORMATION WAS NOT PROVIDED BUT HAS BEEN REQUESTED. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4). ASSOCIATED MEDWATCH REPORTS: 3005673311-2025-00019, 3005673311-2025-00020, 3005673311-2025-00022, 3005673311-2025-00023, 3005673311-2025-00024, 3005673311-2025-00025, 3005673311-2025-00026, 3005673311-2025-00027, 3005673311-2025-00028, 3005673311-2025-00029, 3005673311-2025-00030.
Patient problems: Failure of Implant
MAUDE report 21755137 ↗InjuryMay 20, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
IT WAS REPORTED TO AESCULAP AG THAT A AS TIBIA OFFSET STEM D12X52MM CEMENTED. (PART # NR191Z) WAS IMPLANTED DURING A LEFT KNEE REVISION SURGERY PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, POST-OPERATIVE BONE DETERIORATION OF FEMUR AND TIBIA OCCURRED. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. AN ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. A REVISION IN THE FUTURE IS LIKELY, ACCORDING TO THE PATIENT. ADDITIONAL INFORMATION WAS NOT PROVIDED. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4). ASSOCIATED MEDWATCH REPORTS: (B)(4) - NR191Z (9610612-2022-00142); (B)(4) - NR291Z (9610612-2022-00143); (B)(4) - NR565Z (9610612-2022-00144); (B)(4) - NR565Z (9610612-2022-00145); (B)(4) - NR611M (9610612-2022-00146); (B)(4) - NR005Z (9610612-2022-00148); (B)(4)- NR400Z (9610612-2022-00149).
Patient problems: Failure of Implant
MAUDE report 14456572 ↗InjuryMay 20, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
IT WAS REPORTED TO AESCULAP INC. THAT A AS COLUMBUS REV F TIB. OFFSET CEMENT. T1 (PART # NR071Z) WAS IMPLANTED DURING A LEFT KNEE REVISION SURGERY PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, POST-OPERATIVE BONE DETERIORATION OF FEMUR AND TIBIA OCCURRED. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. AN ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. ACCORDING TO THE PATIENT , A REVISION SURGERY IN THE FUTURE IS LIKELY. ADDITIONAL INFORMATION WAS NOT PROVIDED. THE MALFUNCTION IS FILED UNDER AIC REFERENCE (B)(4). ASSOCIATED MEDWATCH REPORTS: 400554585: NR191Z, (2916714-2022-00055), 400554586: NR291Z, (2916714-2022-00056), 400554587: NR565Z, (2916714-2022-00057), 400554588: NR565Z, (2916714-2022-00058), 400554589: NR611M, (2916714-2022-00059), 400554591: NR005Z, (2916714-2022-00061), 400554592: NR400Z, (2916714-2022-00062).
Patient problems: Failure of Implant
MAUDE report 14456074 ↗InjuryOct 11, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO AESCULAP INC. THAT A COLUMBUS REV F TIB.OFFSET CEMENT.T1 (PART # NR071Z-AS) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE CEMENTING PROCESS OF THE TIBIA TRAY AND STEM, THE CEMENT WORKED ITS WAY INTO THE TIBIA BASEPLATE. THE CEMENT ENTERED THE TIBIA TRAY AT THE JUNCTION OF THE INFERIOR SIDE OF THE TIBIA TRAY. THE CEMENT OBSTRUCTED THE THREADED SCREW HOLE FOR THE TIBIAL INSERT SCREW. REPORTEDLY, THE SURGEON SPENT EXTRA TIME REMOVING THE HARDENED CEMENT FROM THE SMALL SCREW HOLE BEFORE THE SCREW WAS ABLE TO BE PLACED. HOWEVER, THE SURGEON WAS ABLE TO CLEAR THE CEMENT AND INSERT THE SCREW, AND THEN THE PROCEDURE WAS ABLE TO BE CONTINUED AND SUCCESSFULLY COMPLETED. THE COMPLAINT DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AIC REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12610661 ↗InjuryOct 11, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO AESCULAP INC. THAT A COLUMBUS REV F TIB.OFFSET CEMENT.T1 (PART # NR071Z-AS) WAS IMPLANTED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE CEMENTING PROCESS OF THE TIBIA TRAY AND STEM, THE CEMENT WORKED ITS WAY INTO THE TIBIA BASEPLATE. THE CEMENT ENTERED THE TIBIA TRAY AT THE JUNCTION OF THE INFERIOR SIDE OF TIBIA TRAY. THE CEMENT OBSTRUCTED THE THREADED SCREW HOLE FOR THE TIBIAL INSERT SCREW. REPORTEDLY, THE SURGEON SPENT EXTRA TIME REMOVING THE HARDENED CEMENT FROM THE SMALL SCREW HOLE BEFORE THE SCREW WAS ABLE TO BE PLACED. HOWEVER, THE SURGEON WAS ABLE TO CLEAR THE CEMENT AND INSERT THE SCREW. THE COMPLAINT DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605085 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page