- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04046963318185), received between Apr 20, 2018 and Jun 19, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Fail-Safe Problem1
Patient problems
- Foreign Body In Patient5
- Unspecified Infection2
- Needle Stick/Puncture1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJun 19, 2026Break
AS REPORTED IN THE MEDSUN (B)(4). DESCRIBE THE EVENT OR PROBLEM: PT CONSENTED FOR A CAUDAL BLOCK PRIOR TO CHORDEE CORRECTION AND CIRCUMCISION WITH DR. 1 [REDACTED]. AFTER THE BLOCK TIMEOUT AND PREPPING, DR. 2 [REDACTED] STARTED THE CAUDAL BLOCK WITH DR. 3 [REDACTED] BY HER SIDE. CAUDAL ATTEMPTED WITH A 20G CATHETER. ATTEMPT WAS UNSUCCESSFUL AND NEEDLE WITH CATHETER MET RESISTANCE UPON REMOVING. WHEN REMOVING THE IV CATHETER, THE NEEDLE WAS FULLY INTACT, BUT 1.5CM OF THE IV CATHETER WAS MISSING FROM THE CATHETER PULLED OUT. WHAT WAS LEFT OF THE CATHETER WAS FRAYED AT THE END. DR. 3 [REDACTED] DID AN ULTRASOUND IN THE OR AND SAW THE CATHETER LEFT BEHIND SUBCUTANEOUSLY. CARE TEAM MEMBER REACHED OUT TO GENERAL SURGERY MD. DR. 1 [REDACTED] SPOKE WITH IR [INTERVENTIONAL RADIOLOGY] AND PLACED AN ORDER FOR ULTRASOUND IMAGING TO CONFIRM THE LOCATION IS SUBCUTANEOUS. CARE TEAMS CONFIRMED SUBCUTANEOUS LOCATION OF THE 1.3CM CATHETER. FAMILY HAD BEEN NOTIFIED THAT AN ULTRASOUND ORDER WAS PLACED AND WENT TO THE FRONT DESK CONCERNED. FAMILY PLACED IN CONSULT ROOM FOR DR. 1 [REDACTED] TO INFORM THEM OF THE SITUATION. GENERAL SURGERY MD CONSENTED THE FAMILY TO REMOVE THE CATHETER. CATHETER WAS A 20
Patient problems: Foreign Body In Patient
MAUDE report 25563575 ↗InjuryJun 19, 2026Break
AS REPORTED BY THE USER FACILITY: CATHETER BROKEN OFF. DESCRIPTION - PIECE OF THE CATHETER BROKE OFF IN THE PATIENTS ARM AND HAD BEEN SURGICALLY REMOVED. PATIENT WAS ADMITTED TO THE HOSPITAL (B)(6) 2023 WHICH IV WAS STARTED. ON (B)(6) 2023 IV WAS REMOVED ONCE PATIENT WAS RELEASED FROM HOSPITAL, HE WAS SENT HOME AND THE PLACE THE IV WAS GOT INFECTED. THEY HAD AN ULTRASOUND DONE AND FOUND A PIECE OF THE CATHETER IN HIS BASILIC VEIN (RIGHT ARM). ON (B)(6) 2023 SURGERY WAS PERFORMED TO REMOVE THE BROKEN PIECE. PATIENT WAS OKAY. MALE 61 UNABLE TO VERIFY WHICH CATHETER BROKE OFF. PART NUMBERS LISTED ARE 4252527-02 & 4251652-02 IN PATIENT RECORD. AS PART OF AN INTERNAL NONCONFORMANCE, THIS IMPORTER MDR IS BEING SUBMITTED RETROACTIVELY. PLEASE NOTE THAT THE ASSOCIATED MANUFACTURER MDR FOR THIS EVENT, 9610825-2023-00612, HAD ALREADY BEEN PREVIOUSLY SUBMITTED TIMELY ON 18DEC2023. AS B. BRAUN MEDICAL, INC. SUBMITTED THE MANUFACTURER MDR ON BEHALF OF THE MANUFACTURER, THE IMPORTER MDR WAS INADVERTENTLY MISSED.
Patient problems: Foreign Body In Patient; Unspecified Infection
MAUDE report 25563378 ↗MalfunctionJan 6, 2025Fail-Safe Problem; Fail-Safe Problem
AS REPORTED BY MEDSUN FORM 3500A (B)(4): DATE OF EVENT: OCT-2024. DESCRIBE THE EVENT OR PROBLEM: AFTER AN UNSUCCESSFUL IV INSERTION, THE SAFETY MECHANISM DID NOT FUNCTION PROPERLY AND DIDN'T COVER THE SHARP TIP OF THE NEEDLE. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? IV INFUSION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21085857 ↗MalfunctionMay 1, 2024Break; Break
AS REPORTED BY THE USER FACILITY: CUSTOMER COMPLAINT ADVOCATE RECEIVED A LIVE CALL - CUSTOMER REPORTED THAT ON (B)(6) 2024 THE INTROCAN SAFETY FEP 20G, 1.1X45MM-US WAS INSERTED IN A PATIENT DURING IN THE OPERATING ROOM. THE PATIENT WAS THEN TRANSFERRED TO THE FLOOR AND TWO TO THREE LATER A NURSE REMOVED THE INTROCAN SAFETY FEP 20G, 1.1X45MM-US FROM THE PATIENT ARM AND THE DEVICE DETACHED. THE NURSE USED THE PROPER TECHNIQUE WITH ONE HAND APPLIED TO THE PATIENT AND USED THE OTHER HAND TO REMOVE THE DEVICE. THE PINK HUB CAME OUT OF THE PATIENT BUT THE PLASTIC SHEATH THAT IS ATTACHED TO THE PINK HUB DID NOT COME OUT. THE PATIENT HAD TO BE SEDATED TO REMOVED THE DETACHED PIECE STILL IN THE ARM. PATIENT WAS TRANSFERRED BACK THE OPERATING ROOM WHERE THE DEVICE WAS REMOVED. NOT SURE IF AN INCISION WAS REQUIRED. LOCAL ANESTHESIA USED AND LIGHT SEDATION (NON ANESTHESIA). THE DEVICE IS NOT AVAILABLE FOR EVALUATION A PHOTO MAY BE AVAILABLE IN ADDED TO THE PATIENT RECORD.
Patient problems: Foreign Body In Patient
MAUDE report 19220063 ↗InjuryDec 18, 2023Break; Break
AS REPORTED BY THE USER FACILITY: CATHETER BROKEN OFF. DESCRIPTION - PIECE OF THE CATHETER BROKE OFF IN THE PATIENTS ARM AND HAD BEEN SURGICALLY REMOVED. PATIENT WAS ADMITTED TO THE HOSPITAL (B)(6) 2023 WHICH IV WAS STARTED. ON (B)(6) 2023 IV WAS REMOVED. ONCE PATIENT WAS RELEASED FROM HOSPITAL, HE WAS SENT HOME AND THE PLACE THE IV WAS GOT INFECTED. THEY HAD AN ULTRASOUND DONE AND FOUND A PIECE OF THE CATHETER IN HIS BASILIC VEIN (RIGHT ARM). ON (B)(6) 2023 SURGERY WAS PERFORMED TO REMOVE THE BROKEN PIECE. PATIENT WAS OKAY. MALE 61. UNABLE TO VERIFY WHICH CATHETER BROKE OFF. PART NUMBERS LISTED ARE 4252527-02 & 4251652-02 IN PATIENT RECORD.
Patient problems: Foreign Body In Patient; Unspecified Infection
MAUDE report 18347163 ↗MalfunctionFeb 1, 2023Break; Break
AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: "IV CATHETER WILL NOT FLUSH OR DRAW BLOOD BACK AFTER SALINE LOCK APPLIED. SAME APPEARED SHEARED AT THE TIP FOLLOWING REMOVAL." NO INJURY REPORTED.
Patient problems: Needle Stick/Puncture
MAUDE report 16277540 ↗MalfunctionMar 18, 2019Break; Break
AS REPORTED BY USER FACILITY: TIP OF AN INTROCAN IV CATHETER BROKE OFF WHEN THE CATHETER WAS BEING REMOVED AND REMAINED IN PATIENT'S VEIN. CATHETER REMOVAL WAS ATTEMPTED ON (B)(6) 2019. THE CATHETER TIP WAS NOT REMOVED FROM THE PATIENT. A CONSULT WITH A VASCULAR SURGEON WAS COMPLETED AND THE PATIENT WAS PROVIDED THE SURGEONS' OFFICE NUMBER FOR FOLLOW UP IN A WEEK OR TWO AND INSTRUCTED TO CALL IF THERE IS REDNESS OR INCREASED PAIN.
Patient problems: Foreign Body In Patient
MAUDE report 8427974 ↗MalfunctionApr 20, 2018Break; Break
AS REPORTED ON MEDWATCH (B)(4): THE PATIENT WAS TRANSFERRED FROM A OUTSIDE USER FACILITY. THE RN HAD TROUBLE FLUSHING THE IV LINE. ATTEMPT TO FLUSH LINE IN THE RIGHT UPPER ARM CAUSED A LOT OF PAIN FOR THE PATIENT. THE LINE COULD NOT BE FLUSHED. THERE WAS A HARD PLASTIC-LIKE BULGE FELT WHERE THE CATHETER SHOULD BE. THE LINE WAS PULLED AND FOUND TO BE VERY KINKED. THE TIP SEEMED TO BE TORN OFF. THE PROVIDER WAS NOTIFIED. A CONCERN FOR THE RETAINED TIP WAS NOTED AND DISCLOSURE TO THE PATIENT WAS DONE. MULTIPLE DIAGNOSTIC EXAMS WERE PERFORMED TO LOCATE THE TIP OF IV LINE. THERE WAS NO VERIFICATION BY AN EXAM THAT THERE WAS A RETAINED FOREIGN OBJECT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7448735 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page