Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046963809829), received between Feb 25, 2021 and Aug 14, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1
  • Patient-Device Incompatibility1

Patient problems

  • Ambulation or Postural Difficulties1
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 14, 2026Patient-Device Incompatibility

    An outside law firm reported that a patient experienced postoperative loosening following a left total knee arthroplasty. According to the available medical records, the patient underwent a left total knee arthroplasty on (b)(6) 2021. On (b)(6) 2024, the patient underwent revision of the left patellar component for aseptic loosening. The operative report documented that the femoral and tibial components were inspected and appeared to have no loosening. The patellar button was evaluated, and when an osteotome was placed at the medial side of the button, the button reportedly popped out. The remaining cement was removed from the patella and the patella was resurfaced to allow for improved tracking. No intraoperative complications were reported, and the patient was taken to recovery in stable condition. Subsequent records documented further concerns involving the left total knee prosthesis. At a (b)(6) 2025 postoperative evaluation, radiographs demonstrated a posterior-stabilized total knee replacement with the patella resurfaced. The femoral component appeared loose, with severe periprosthetic osteolysis about the femur. The patient was assessed with mechanical loosening and peripros

    Patient problems: Insufficient Information

    MAUDE report 50260356 ↗
  • InjuryFeb 25, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED TO AESCULAP INC. THAT A VEGA KNEE IMPLANT SYSTEM WAS IMPLANTED DURING A PRIMARY SURGERY PERFORMED ON (B)(6) 2015. ACCORDING TO THE COMPLAINANT, FOLLOWING THE PRIMARY SURGERY, THE PATIENT EXPERIENCED PAIN AND DIFFICULTY WALKING OF THE IMPLANT. THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2018. THE COMPLAINT DEVICE WAS NOT AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED AT THIS TIME, IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE COMPLAINT WILL BE UPDATED ACCORDINGLY. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER REFERENCE XC (B)(4). INVOLVED COMPONENTS: NX007Z (PS FEMUR CEMENTED F3N LT); NN260P (PEEK PLUG F/ TIBIA); NX049Z (PS TIBIA CEMENTED T0); NX042 (UNIVERSAL PATELLA P2); NX102 (PS PE INSERT T0/T0+, 14MM). CEMENT USED IS UNKNOWN. ASSOCIATED MEDWATCHES: 2916714-2020-00313, 2916714-2020-00314, 2916714-2020-00315, 2916714-2020-00316.

    Patient problems: Ambulation or Postural Difficulties; Pain

    MAUDE report 11381276 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates