AS VEGA System® PS NX032ZRecalls & reported problems
Aesculap AG · AS VEGA PS FEMORAL COMP.CEMENTED F5R · Class 2 · Implant
All models of AS VEGA System PS (28)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046963810184), received between Oct 14, 2022 and Oct 21, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth2
Patient problems
- Failure of Implant2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 21, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH NX032Z - AS VEGA PS FEMORAL COMP.CEMENTED F5R. ACCORDING TO THE COMPLAINT DESCRIPTION, THE VEGA DEVICES WERE REMOVED. THE ORIGINAL SURGERY WAS ABOUT FIVE (5) YEARS AGO. DURING REMOVAL, IT WAS NOTED THAT CEMENT NO LONGER ADHERED TO IMPLANTS NOR BONE; PICTURES WERE PROVIDED. REPLACEMENTS FOR THE IMPLANTS WERE MOST LIKELY NON-AESCULAP PRODUCTS. A REVISION WAS REQUIRED DUE TO LOOSENING. ADDITIONAL DETAILS HAVE BEEN REQUESTED. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4). ASSOCIATED MEDWATCH REPORTS: (B)(4) (9610612-2022-00294) NX032Z, (B)(4) (9610612-2022-00295) AE-QAS-K521-56, (B)(4) (9610612-2022-00296) AE-QAS-K521-56.
Patient problems: Failure of Implant
MAUDE report 15655160 ↗InjuryOct 14, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED TO AESCULAP INC. THAT AN AS VEGA PS FEMORAL COMP. CEMENTED F5R (PART # NX032Z) WAS IMPLANTED DURING A PRIMARY PROCEDURE PERFORMED APPROXIMATELY FIVE (5) YEARS AGO BY ANOTHER SURGEON. ACCORDING TO THE COMPLAINANT, POSTOPERATIVE LOOSENING OF THE IMPLANT OCCURRED. REPORTEDLY, THE CEMENT DID NOT ADHERE TO THE IMPLANTS OR THE BONE. THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2022. THE COMPLAINT DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION SURGERY. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AIC REFERENCE (B)(4). ASSOCIATED MEDWATCH REPORT: (B)(4) (2916714-2022-00103) AE-QAS-K521-56. (B)(4) (2916714-2022-00104 ) AE-QAS-K521-56.
Patient problems: Failure of Implant
MAUDE report 15605328 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page