Skip to content

AS VEGA System® PS NX059ZRecalls & reported problems

Aesculap AG · AS VEGA PS TIBIAL PLATEAU CEMENTED T5 · Class 2 · Implant

All models of AS VEGA System PS (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046963810535), received between Mar 25, 2021 and Sep 8, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Patient-Device Incompatibility1

Patient problems

  • Failure of Implant1
  • Implant Pain1
  • Joint Laxity1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 8, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    THE PATIENT PRESENTED WITH PAIN AND SWELLING OF THE LEFT KNEE. IMAGING DEMONSTRATED LOOSENING OF THE VEGA KNEE SYSTEM. ON (B)(6) 2025, THE PATIENT UNDERWENT REVISION TOTAL KNEE ARTHROPLASTY OF THE LEFT KNEE FOR LOOSENING OF PRIOR COMPONENTS. PREOPERATIVE EVALUATION INDICATED ASEPTIC LOOSENING; NO INFECTION WAS DETECTED. INTRAOPERATIVELY, THE TIBIAL COMPONENT WAS NOTED TO BE GROSSLY LOOSE WITH NO CEMENT PRESENT ON ANY ASPECT OF THE IMPLANT. THE FEMORAL COMPONENT WAS ALSO LOOSE AND REMOVED EASILY WITH LIGHT TAPPING. THE PATELLAR COMPONENT WAS STABLE WITH NO EVIDENCE OF LOOSENING. FINDINGS WERE CONSISTENT WITH IMPLANT DEBONDING AT THE CEMENT INTERFACE. THE FEMUR AND TIBIA WERE PREPARED, AND REVISION FEMORAL AND TIBIAL COMPONENTS WERE IMPLANTED USING CEMENT FIXATION. A POLYETHYLENE INSERT AND MODULAR AUGMENTATION COMPONENTS WERE PLACED FOR STABILITY. AN ALLOGRAFT AND BONE CEMENT WERE ALSO UTILIZED. FINAL INSPECTION CONFIRMED STABLE FIXATION AND APPROPRIATE ALIGNMENT. THE WOUND WAS IRRIGATED, VANCOMYCIN POWDER WAS APPLIED, AND THE CAPSULE WAS CLOSED. THE PATIENT TOLERATED THE PROCEDURE AND WAS TRANSFERRED TO PACU IN STABLE CONDITION.

    Patient problems: Failure of Implant

    MAUDE report 22994165 ↗
  • InjuryMar 25, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A VEGA KNEE COMPONENTS. AS A RESULT OF HAVING AN AESCULAP PRODUCT IMPLANTED, THE PATIENT HAS EXPERIENCED KNEE PAIN AND LOOSENING OF THE IMPLANT. THE PRIMARY SURGERY OCCURRED ON (B)(6) 2015 , WHICH INCLUDED NON-AESCULAP IMPLANTS. A REVISION TO REPLACE NON-AESCULAP PRODUCTS WITH AESCULAP IMPLANTS OCCURRED ON (B)(6) 2017, AND THERE WAS NO REPORTED SECOND REVISION SURGERY. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED AT THIS TIME, IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE COMPLAINT WILL BE UPDATED ACCORDINGLY. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER REFERENCE (B)(4). INVOLVED COMPONENTS: (B)(4) (PS FEMUR CEMENTED F5N LT), (B)(4)( PS TIBIA CEMENTED T5), (B)(4) (PS PE INSERT T5, 14MM), (B)(4) (TIBIAL OBTURATOR D14MM, L7MM). ASSOCIATED MEDWATCHES: 2916714-2020-00396, 2916714-2020-00398, 2916714-2020-00399.

    Patient problems: Implant Pain; Joint Laxity

    MAUDE report 11568635 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates