Aesculap AG · VEGA PS GLIDING SURFACE T0/0+ 10MM · Class 2 · Implant

All models of VEGA System PS (31)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04046963810702), received between Sep 15, 2020 and Nov 6, 2020 — classified as 3 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Mechanical Problem1

Patient problems

  • Ambulation or Postural Difficulties2
  • Joint Laxity2
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1
  • Synovitis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 6, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH AS VEGA KNEE, LEFT SIDE. IT WAS REPORTED THAT AS A RESULT OF HAVING THE PRODUCT IMPLANTED, THE PATIENT HAS EXPERIENCED KNEE PAIN AND SWELLING, AND NORMAL DAILY ACTIVITIES WERE RESTRICTED DUE TO THE PAIN. THE PRIMARY PROCEDURE OCCURRED ON (B)(6) 2013, AND THE REVISION WAS PERFORMED ON (B)(6) 2016. INTRAOPERATIVE FINDINGS SHOWED LOOSENING OF THE FEMORAL COMPONENT, METALLOSIS, AND DEBRIS IN THE KNEE JOINT, AND REACTIVE SYNOVITIS. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED AT THIS TIME, IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE COMPLAINT WILL BE UPDATED ACCORDINGLY. INVOLVED COMPONENTS: NX009Z (PS FEMUR CEMENTED F4N LT); NX042 (UNIVERSAL PATELLA P2); NX049Z (PS TIBIA CEMENTED T0); NX100 (PS PE INSERT T0/T0+, 10MM); NN260P (PEEK PLUG F/ TIBIA). THE CEMENT USED WAS UNSPECIFIED. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER REFERENCE (B)(4). ASSOCIATED MEDWATCHES: 2916714-2020-00506; 2916714-2020-00507; 2916714-2020-00508; 2916714-2020-00510.

    Patient problems: Ambulation or Postural Difficulties; Pain; Synovitis

    MAUDE report 10802532 ↗
  • InjurySep 29, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH AS VEGA KNEE. AS A RESULT OF HAVING THE PRODUCT IMPLANTED THE PATIENT HAS EXPERIENCED KNEE PAIN, SWELLING, DIFFICULTY WALKING, AND LOOSENING AND INSTABILITY OF THE IMPLANT. THE PRIMARY SURGERY OCCURRED ON (B)(6) 2015 AND THE REVISION SURGERY OCCURRED ON (B)(6) 2016. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THE COMPLAINT WILL BE UPDATED ACCORDINGLY. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER REFERENCE (B)(4). ASSOCIATED MEDWATCHES: 2916714-2020-00411, 2916714-2020-00413, 2916714-2020-00414, 2916714-2020-00415. INVOLVED COMPONENTS: NX009Z (PS FEMUR CEMENTED F4N LT). NX100 (PS PE INSERT T0/T0+, 10MM). NX049Z (PS TIBIA CEMENTED T0). NN260P (PEEK PLUG F/ TIBIA). NX042 (UNIVERSAL PATELLA P2). THE CEMENT USED IS UNIDENTIFIED.

    Patient problems: Ambulation or Postural Difficulties; Joint Laxity

    MAUDE report 10601798 ↗
  • InjurySep 15, 2020Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth; Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth

    NO UPDATES.

    Patient problems: Joint Laxity; No Clinical Signs, Symptoms or Conditions