Skip to content

INTROCAN SAFETY® 3 4251127-02Recalls & reported problems

B. Braun Medical Inc. · Introcan Safety® 3 Closed IV Catheter 24 Ga. x 0.75 in., PUR, Winged · Class 2

All models of INTROCAN SAFETY 3 (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04046964013997), received between Nov 16, 2023 and Sep 29, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash4
  • Contamination /Decontamination Problem1

Patient problems

  • Chemical Exposure3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Contamination /Decontamination Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928021 ↗
  • MalfunctionOct 2, 2024Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: EVENT 2 BRIEF INQUIRY DESCRIPTION MULTIPLE OCCURRENCES OF INTROCAN SAFETY 3 IV S LEAKING AT THE EXTENSION SITE CONNECTION DETAILED INQUIRY DESCRIPTION IVS LEAKED DURING CHEMO ADMINISTRATION AND MEDICATION SPILLED ONTO PATIENTS.

    Patient problems: Chemical Exposure

    MAUDE report 20355353 ↗
  • MalfunctionOct 2, 2024Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: EVENT 3 BRIEF INQUIRY DESCRIPTION MULTIPLE OCCURRENCES OF INTROCAN SAFETY 3 IV S LEAKING AT THE EXTENSION SITE CONNECTION DETAILED INQUIRY DESCRIPTION IVS LEAKED DURING CHEMO ADMINISTRATION AND MEDICATION SPILLED ONTO PATIENTS.

    Patient problems: Chemical Exposure

    MAUDE report 20355309 ↗
  • MalfunctionOct 2, 2024Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: EVENT 1 BRIEF INQUIRY DESCRIPTION MULTIPLE OCCURRENCES OF INTROCAN SAFETY 3 IV S LEAKING AT THE EXTENSION SITE CONNECTION DETAILED INQUIRY DESCRIPTION IVS LEAKED DURING CHEMO ADMINISTRATION AND MEDICATION SPILLED ONTO PATIENTS.

    Patient problems: Chemical Exposure

    MAUDE report 20355029 ↗
  • MalfunctionNov 16, 2023Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: CHEMO LEAK AT CATHETER-JLOOP CONNECTION. DETAILED INQUIRY DESCRIPTION: RN CALLED WITH CONCERN FOR CHEMO LEAK DURING IV INFUSION. SHE STATED IV HAD GOOD BLOOD RETURN AND SALINE FLUSH (VIA IV DRIP) INFUSED WITHOUT ISSUE. CHEMO WAS THEN STARTED VIA PUMP INFUSION. LEAK WAS IDENTIFIED AT CATHETER-JLOOP CONNECTION. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18151503 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page