Skip to content

Discofix® 456003Recalls & reported problems

B. Braun Medical Inc. · Three-way stopcock with SPIN-LOCK® · Class 1

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04046964183577), received between Nov 29, 2016 and Feb 20, 2026 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Operates Differently Than Expected3
  • Leak/Splash3
  • Break2
  • Fluid/Blood Leak2
  • Contamination /Decontamination Problem1
  • Degraded1

Patient problems

  • No Consequences Or Impact To Patient5
  • No Clinical Signs, Symptoms or Conditions4
  • Blood Loss2
  • No Patient Involvement1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026Leak/Splash

    AS REPORTED BY THE USER FACILITY: NICU WAS RUNNING TPN THROUGH THE STOPCOCK. THE NURSE NOTED THERE WAS VISIBLE LEAKING WHEN SHE HOOKED UP THE TPN'S. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405355 ↗
  • MalfunctionNov 11, 2025Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: DURING THERAPY THE STOPCOCK LEAKED. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23526819 ↗
  • MalfunctionNov 6, 2025Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: DURING THERAPY THE STOPCOCK LEAKED. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23491203 ↗
  • MalfunctionJan 29, 2025Degraded; Degraded

    AS REPORTED BY THE USER FACILITY: DETAILS OF COMPLAINT (REPORTED ISSUE): "WE HAVE A GROUP OF THESE 3 WAY STOPCOCKS RECEIVED THAT ARE DEFECTIVE. IF YOU LOOK CLOSELY THERE IS A DEFECT ON THE TIP THAT ATTACHES TO THE PATIENTS IV LINE. THIS IS THE FIRST BATCH WE HAVE SEEN THAT ARE DEFECTIVE." NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21263969 ↗
  • MalfunctionFeb 28, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT AT THE END OF THE MANUFACTURING PROCESS A PARTICULATE WAS OBSERVED IN THE FINAL PRODUCT BAG. SINCE THE DEFECT WAS NOT OBSERVED UNTIL THE END OF THE MANUFACTURING PROCESS, ALL COMPONENTS USED ARE BEING INVESTIGATED.

    Patient problems: No Patient Involvement

    MAUDE report 8379133 ↗
  • MalfunctionNov 20, 2017Break; Break

    AS REPORTED BY THE USER FACILITY (TRANSLATION OF USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN FRANCE: STOPCOCK-DAMAGED-BROKEN. RESULTED IN SIGNIFICANT BLOOD LOSS, AND THE PATIENT HAD TO RECEIVE A TRANSFUSION.

    Patient problems: Blood Loss

    MAUDE report 7044732 ↗
  • InjuryNov 20, 2017Break; Break

    AS REPORTED BY THE USER FACILITY (TRANSLATION OF USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN FRANCE: STOPCOCK-DAMAGED-BROKEN. RESULTED IN SIGNIFICANT BLOOD LOSS, AND THE PATIENT HAD TO RECEIVE A TRANSFUSION.

    Patient problems: Blood Loss

    MAUDE report 7044725 ↗
  • MalfunctionAug 29, 2017Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED BY THE USER FACILITY: LEAKAGE OF CHEMOTHERAPY DURING CHEMO EMBOLIZATION. NO INJURIES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6828343 ↗
  • MalfunctionAug 29, 2017Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED BY THE USER FACILITY: THE STOPCOCK LEAKED BLOOD DURING USE. WHILE RETURNING BLOOD TO PATIENT THROUGH A CLOSED SYSTEM, APPROXIMATELY 2-3 CC OF BLOOD LEAKED OUT OF THE DISTAL STOPCOCK CONNECTION ONTO THE PATIENT AND THE PATIENT'S MOTHER. THE PATIENT'S CLOSED SYSTEM WAS THEN FLUSHED WITH 10 CC OF NORMAL SALINE AND 300 UNITS OF HEPARIN TO THE PATIENT'S PORT AND THEN THE PATIENT WAS DE-ACCESSED. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6828341 ↗
  • MalfunctionNov 29, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected

    AS REPORTED BY THE USER FACILITY: EVENT 2: CUSTOMER REPORTS THAT THE PRODUCT HAS BEEN BREAKING. FOR EXAMPLE, WHEN USING THEM TO INFLATE A BALLOON DURING AN ANGIOPLASTY, THE BLUE SWITCH WILL POP OUT OF THE STOPCOCK, AND CONTRAST WILL SPEW EVERYWHERE. THIS HAS HAPPENED AT LEAST 3 DIFFERENT TIMES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6133856 ↗
  • MalfunctionNov 29, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected

    AS REPORTED BY THE USER FACILITY: EVENT 3: CUSTOMER REPORTS THAT THE PRODUCT HAS BEEN BREAKING. FOR EXAMPLE, WHEN USING THEM TO INFLATE A BALLOON DURING AN ANGIOPLASTY, THE BLUE SWITCH WILL POP OUT OF THE STOPCOCK, AND CONTRAST WILL SPEW EVERYWHERE. THIS HAS HAPPENED AT LEAST 3 DIFFERENT TIMES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6133841 ↗
  • MalfunctionNov 29, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected

    AS REPORTED BY THE USER FACILITY: EVENT 1: CUSTOMER REPORTS THAT THE PRODUCT HAS BEEN BREAKING. FOR EXAMPLE, WHEN USING THEM TO INFLATE A BALLOON DURING AN ANGIOPLASTY, THE BLUE SWITCH WILL POP OUT OF THE STOPCOCK, AND CONTRAST WILL SPEW EVERYWHERE. THIS HAS HAPPENED AT LEAST 3 DIFFERENT TIMES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6133824 ↗

Other ACCESSORIES, CATHETER devices

All ACCESSORIES, CATHETER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.