The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964310997), received between Jun 1, 2017 and Jun 1, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Kinked” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Kinked1
- Material Rupture1
Patient problems
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 1, 2017Kinked; Material Rupture
AS PER HOSPITAL MEDWATCH REPORT: "A MALE PATIENT WITH HISTORY OF BICUSPID AORTIC VALVE DISEASE STATUS POST ROSS PROCEDURE WITH RESULTING STENOTIC PULMONARY HOMOGRAFT ADMITTED FOR TRANSCATHETER PULMONARY VALVE REPLACEMENT. BALLOON SIZING OF THE PULMONARY ARTERY WITH DYNA CT PERFORMED WITH A 30 MM X 5 CM NUMED PERCUTANEOUS TRANSLUMINAL SIZING (PTS) SIZING BALLOON. BALLOON WAS ADVANCED INTO THE RIGHT VENTRICULAR OUTFLOW TRACT (RVOT) AND MAIN PULMONARY ARTERY AND INFLATED UNDER RAPID PACING. THE BALLOON RUPTURED AT LOW PRESSURE REQUIRING REMOVAL OF 16F SHEATH AND REPLACEMENT WITH A 16F SHORT COOK SHEATH AND THE BALLOON WAS SUCCESSFULLY RETRIEVED. THE PROCEDURE WAS ABORTED AND RESCHEDULED." ADDITION INFORMATION PER DISTRIBUTOR: "IT IS UNKNOWN WHETHER AN INFLATION DEVICE WITH PRESSURE GAUGE WAS USED. THE CATHETER SHAFT WAS KINKED. THERE WAS NOTHING UNUSUAL ABOUT THE PATIENT ANATOMY. THE PATIENT WAS STABLE POST PROCEDURE. THE BALLOON RUPTURED CIRCUMFERENTIALLY, IT IS UNKNOWN AT WHAT ATM THE BALLOON RUPTURED - THE REPORT STATES ONLY 'AT LOW PRESSURE'."
Patient problems: No Code Available
MAUDE report 6604958 ↗
Other Catheter, intravascular occluding, temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports