TYSHAK II® 611940Recalls & reported problems
B Braun Interventional Systems, Inc · PDC540 23MM X 4CM BALL CATH TYSHA II STD · Class 2
All models of TYSHAK II (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046964339417), received between Jun 27, 2017 and Aug 3, 2020 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K003052, K931009); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Rupture4
- Degraded1
- Physical Resistance/Sticking1
Patient problems
- No Known Impact Or Consequence To Patient2
- Foreign Body In Patient1
- Ischemia1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2020Degraded; Material Rupture
AS REPORTED TO NUMED FROM DISTRIBUTOR (B)(6) - UPON POSITIVE PREPARATION OF THE BALLOON DILATATION CATHETER OUTSIDE THE PATIENT'S BODY THE EVENT REPORTER STATED THE BALLOON "DISINTEGRATED" WITH AROUND 22ML OF SOLUTION IN THE BALLOON. THE BALLOON RUPTURED. NO PATIENT HARM. UPDATE (B)(6) 2020 FROM (B)(6) - IN A CONVERSATION WITH THE ACCOUNT: USED A 60CC SYRINGE WITH A 3 TO 1 SALINE TO CONTRAST RATIO AND PLACED APPROXIMATELY 21 TO 22 ML OF SOLUTION INTO THE BALLOON DURING POSITIVE PREPARATION. THE BALLOON RUPTURED / DISINTEGRATED. NO GUIDEWIRE WAS UTILIZED DURING PREPARATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10355939 ↗MalfunctionJul 31, 2020Material Rupture
BALLOON BEING PREPPED BEFORE ENTERING THE BODY. BALLOON POPPED DURING PREP TIME WITHOUT FULL AMOUNT OF LIQUID IN BALLOON.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10349511 ↗InjuryNov 20, 2018Material Rupture; Physical Resistance/Sticking
DESCRIPTION TAKEN DIRECTLY FROM A MEDWATCH REPORT RECEIVED DIRECTLY FROM THE USER FACILITY: WITH THE USE OF FLUOROSCOPIC AND VASCULAR ULTRASOUND GUIDANCE, THE RIGHT COMMON FEMORAL ARTERY WAS ACCESSED. ANGIOGRAPHY REVEALED GOOD CATHETER POSITIONING WITH A MICROPUNCTURE KIT AND A SINGLE PERCLOSE PROGLIDE DEVICE WAS PLACED AND THEN A 9 FRENCH SHEATH WITHOUT DIFFICULTY. HEPARIN WAS THEN ADMINISTERED AT FULL DOSAGE. DUE TO THE RIGHT FEMORAL ARTERY SHEATH, A PIGTAIL CATHETER WAS ADVANCED TO THE LEVEL OF THE EDWARDS SAPIEN VALVE AND THE VALVE WAS CROSSED WITHOUT DIFFICULTY. THROUGH THE PIGTAIL CATHETER, A SAFARI EXTRA SMALL WIRE WAS PLACED IN THE LV APEX AND THE PIGTAIL CATHETER WAS SUBSEQUENTLY REMOVED. OVER THIS WIRE, A TYSHAK 23 MM BALLOON WAS PASSED AND ACROSS THE NATIVE VALVE WITHOUT DIFFICULTY. RAPID VENTRICULAR PACING WAS PERFORMED WITH THE ASSISTANCE OF A PACEMAKER REP WITH A MAXIMUM RATE OF 170 WITH GOOD DROP IN SYSTOLIC PRESSURE AND VERY LOW PULSE PRESSURE. THE BALLOON IS FULLY INFLATED AND AT MAXIMUM INFLATION, THE BALLOON DID RUPTURE. THE BALLOON CATHETER WAS WITHDRAWN TO THE LEVEL OF THE SHEATH MAKING SURE THAT THE BALLOON MARKER ARE IN PLACE. THERE WAS SOME RESISTANCE TO REM
Patient problems: Foreign Body In Patient; Ischemia
MAUDE report 8090911 ↗MalfunctionJun 27, 2017Material Rupture
AS PER THE REPORT FROM BIS - "DURING A VALVULOPLASTY PROCEDURE, PHYSICIAN HAD THE BALLOON RUPTURE UPON INFLATION. THERE WAS NOT HARM CAUSED TO THE PATIENT AND THE PROCEDURE ENDED UP WITH A GOOD RESULT, DESPITE THIS RUPTURE. THE CARDIAC OUTPUT PRIOR TO THE PROCEDURE WAS A 1.6 AND AFTER THE PROCEDURE IT WAS AN AVG OF 1.95. CALCIFICATION WAS NOTED AT THE PROCEDURE SITE. AN INFLATION DEVICE WITH PRESSURE GAUGE WAS NOT USED. THE ATM AT WHICH THE BALLOON RUPTURED IS UNKNOWN BECAUSE A SYRINGE WAS USED. THE CATHETER SHAFT WAS NOT KINKED. A 9F TERUMO PINNACLE SHEATH WAS USED. NO UNUSUAL PATIENT ANATOMY WAS REPORTED."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6669240 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K003052, K931009. One submission often covers many models, so check the product and lot details in each record.
- Z-0509-2010 · Sep 23, 2009 · TerminatedIMPACT Catheters, Balloon Dilatation and Esophageal Dilatation Catheter, Model Number 307, STERILE, Rx only --- Common/Usual Name: PTA and Esophageal Catheter --- Manufactured f
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page