Mini Ghost™ 612542Recalls & reported problems
B Braun Interventional Systems, Inc · MG132 MINI GHOST BALLOON 4.5 X 2 X 80 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964341243), received between Nov 11, 2019 and Nov 11, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
- Material Rupture1
- Melted1
Patient problems
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 11, 2019Break; Melted; Material Rupture
AS PER THE REPORT FROM THE DISTRIBUTOR / HOSPITAL - WE WENT UP TO 30 ATM AND IT JUST BROKE. WE GOT HALF OF THE BALLOON OUT AND THEN CONTINUED TO PULL OUT PIECES AFTER. THE SURGEON GOT THE BIGGER PIECE OUT OF THE SUBCLAVIAN. IT LOOKED LIKE IT HAD MELTED ONTO THE GUIDEWIRE. THE BALLOON RUPTURED BOTH CIRCUMFERENTIALLY AND LONGITUDINALLY. THE PHYSICIAN WAS USING THE BALLOON FOR ANGIOPLASTY RIGHT PULMONARY ARTERY. AN INFLATION DEVICE WITH PRESSURE GAUGE WAS USED - MERIT GAUGE BASIC COMPACT INFLATION SYRINGE - 30ATM. THE SIZE AND TYPE OF INTRODUCER SHEATH USED WAS A 4F 45 CHB PERFORMER SHEATH; THE COOK ABBOT WORKHORSE .018 X 200 FLEX T GUIDEWIRE WAS USED. THE CATHETER SHAFT WAS NOT KINKED. THE PATIENT HAS A CONGENITAL DEFECT - TETRALOGY OF FALLOT. THE PATIENT CONDITION POST PROCEDURE WAS STABLE. A SECOND BALLOON WAS NOT USED TO COMPLETE THE PROCEDURE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9304814 ↗
FDA clearance and classification
Cleared through Special 510(k) K051343 on Jul 28, 2005 (decision: Substantially Equivalent). Submitted by NuMED, Inc. as “Mini Ghost Pta Catheter, Model 507”. The FDA review took 66 days. FDA record ↗
The FDA regulates product code DQY (catheter, percutaneous) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Mini Ghost™ 612542
Has the Mini Ghost™ 612542 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04046964341243), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Mini Ghost™ 612542?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 11, 2019. That report lists device problems “Break” (1), “Material Rupture” (1) and “Melted” (1); patient problem “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Mini Ghost™ 612542?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Mini Ghost™ 612542 FDA approved?
It was cleared through Special 510(k) K051343 on Jul 28, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Mini Ghost™ 612542, and is it still sold?
The FDA's device database lists B Braun Interventional Systems, Inc as the company responsible for it (brand Mini Ghost™, model 612542). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths651 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall362 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall31 deaths315 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths276 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS293 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS219 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record