Skip to content

AESCULAP FE953KRecalls & reported problems

Aesculap AG · AVM MICROCLIP PHYNOX STR.3MM STERILE · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964381485), received between Dec 28, 2023 and Dec 28, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Failure to Seal” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Seal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 28, 2023Failure to Seal; Failure to Seal

    IT WAS REPORTED TO AESCULAP INC. THAT AN AVM MICROCLIP PHYNOX STR.3MM STERILE (PART# FE953K) WAS USED DURING A PROCEDURE ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT THE CLIPS WERE STAYING OPEN AFTER BEING LOADED INTO THE PATIENT. REPORTEDLY THE PATIENT WAS RETURNED TO THE OPERATING ROOM AFTER THE PROCEDURE AND 3 MORE CLIPS WERE IMPLANTED FOR THE SECOND TIME. THERE WERE SEVEN ATTEMPTS AT IMPLANTING IN TOTAL WHERE THE CLIPS WERE STAYING OPEN. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18409051 ↗

FDA clearance and classification

The FDA regulates product code HCH (clip, aneurysm) under 21 CFR 882.5200 as a Class II device. 21 CFR 882.5200 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AESCULAP FE953K

Has the AESCULAP FE953K been recalled?

No FDA recall record in our data names this exact device identifier (DI 04046964381485), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AESCULAP FE953K?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 28, 2023. That report lists device problem “Failure to Seal” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AESCULAP FE953K?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the AESCULAP FE953K, and is it still sold?

The FDA's device database lists Aesculap AG as the company responsible for it (brand AESCULAP, model FE953K). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CLIP, ANEURYSM devices

All CLIP, ANEURYSM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page