PINNACLE® PC1000Recalls & reported problems

B. Braun Medical Inc. · PINNACLE® TPN Management System · Class 2

All models of PINNACLE (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04046964447198), received between Aug 11, 2021 and May 15, 2023 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Application Program Problem: Dose Calculation Error” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Application Program Problem: Dose Calculation Error9
  • Contamination of Device Ingredient or Reagent2

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2023Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY THE USER FACILITY: EVENT 2 CUSTOMER CALLED TO REPORT THAT THEY HAVE HAD 2 INSTANCES WHERE LIPIDS LEAKED INTO THE LINE WHILE ANOTHER INGREDIENT WAS PUMPING. THEY STATED THAT IT HAPPENED ONCE LAST WEEK AND THEN AGAIN TODAY ((B)(6) 2023).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16931020 ↗
  • MalfunctionMay 15, 2023Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    AS REPORTED BY THE USER FACILITY: EVENT 1 CUSTOMER CALLED TO REPORT THAT THEY HAVE HAD 2 INSTANCES WHERE LIPIDS LEAKED INTO THE LINE WHILE ANOTHER INGREDIENT WAS PUMPING. THEY STATED THAT IT HAPPENED ONCE LAST WEEK AND THEN AGAIN TODAY ((B)(6) 2023).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16931019 ↗
  • MalfunctionSep 16, 2022Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: CUSTOMER CALLED TO REPORT HAVING AN ISSUE WITH 35ML OF A SOLUTION BEING MISTAKENLY ADDED TO A FRESHLY HUNG FINAL CONTAINER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15442507 ↗
  • MalfunctionJun 13, 2022Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY USER FACILITY: THE PUMP WAS NOT PULLING UP SOLUTION AT THE BEGINNING OF COMPOUNDING. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14679602 ↗
  • MalfunctionSep 24, 2021Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: THE PINNACLE WILL RANDOMLY SKIP LINES WHILE COMPOUNDING AND THE MOST COMMON LINES SKIPPED ARE LINES 1 (LIPID) AND 2 (DEXTROSE) BUT OTHER LINES HAVE SKIPPED AS WELL. NO ERRORS ARE ALARMED. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12526865 ↗
  • MalfunctionSep 13, 2021Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: EVENT 1: THERE ARE ISSUES WITH BAGS OVER-DELIVERING ON THE PINNACLE COMPOUNDER. NO INJURIES WERE REPORTED. FINAL CONTAINERS WERE SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12459978 ↗
  • MalfunctionSep 13, 2021Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: EVENT 3. THERE ARE ISSUES WITH BAGS OVER-DELIVERING ON THE PINNACLE COMPOUNDER. NO INJURIES WERE REPORTED. FINAL CONTAINERS WERE SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12459972 ↗
  • MalfunctionSep 13, 2021Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: EVENT 2. THERE ARE ISSUES WITH BAGS OVER-DELIVERING ON THE PINNACLE COMPOUNDER. NO INJURIES WERE REPORTED. FINAL CONTAINERS WERE SCRAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12459968 ↗
  • MalfunctionAug 11, 2021Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: EVENT 3: PATIENTS REPORTED TO CUSTOMER THAT BAGS ARE RUNNING DRY DURING INFUSION. THE BAGS ARE COMPOUNDED ON THE PINNACLE COMPOUNDER. USER FACILITY HAS TRIED INCREASING OVERFILL AND CALIBRATING THE PUMP AFTER EVERY BAG. THEY REPORT THAT NO PINNACLE ERRORS HAVE OCCURRED DURING THIS TIME. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12306368 ↗
  • MalfunctionAug 11, 2021Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: EVENT 2: PATIENTS REPORTED TO CUSTOMER THAT BAGS ARE RUNNING DRY DURING INFUSION. THE BAGS ARE COMPOUNDED ON THE PINNACLE COMPOUNDER. USER FACILITY HAS TRIED INCREASING OVERFILL AND CALIBRATING THE PUMP AFTER EVERY BAG. THEY REPORT THAT NO PINNACLE ERRORS HAVE OCCURRED DURING THIS TIME. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12306364 ↗
  • MalfunctionAug 11, 2021Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS REPORTED BY THE USER FACILITY: EVENT 1: PATIENTS REPORTED TO CUSTOMER THAT BAGS ARE RUNNING DRY DURING INFUSION. THE BAGS ARE COMPOUNDED ON THE PINNACLE COMPOUNDER. USER FACILITY HAS TRIED INCREASING OVERFILL AND CALIBRATING THE PUMP AFTER EVERY BAG. THEY REPORT THAT NO PINNACLE ERRORS HAVE OCCURRED DURING THIS TIME. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12306363 ↗

FDA clearance and classification

The FDA regulates product code NEP (system/device, pharmacy compounding) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Common questions about the PINNACLE® PC1000

Has the PINNACLE® PC1000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04046964447198), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PINNACLE® PC1000?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Aug 11, 2021 and May 15, 2023. In those reports, the most common specific device problems are “Application Program Problem: Dose Calculation Error” (9) and “Contamination of Device Ingredient or Reagent” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PINNACLE® PC1000?

None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PINNACLE® PC1000 FDA approved?

Devices of this type (product code NEP) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by B. Braun Medical Inc.

Who makes the PINNACLE® PC1000, and is it still sold?

The FDA's device database lists B. Braun Medical Inc. as the company responsible for it (brand PINNACLE®, model PC1000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SYSTEM/DEVICE, PHARMACY COMPOUNDING devices