Quick summary

This B. Braun Easypump elastomeric infusion pump has 41 adverse event reports naming its identifier, including 1 death and 1 injury. Reports mostly describe foreign particles in the fluid path, plus a few leaks, air bubbles, or over-infusion. There are no recalls for this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
41
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 04046964448942), received between Mar 22, 2019 and Mar 24, 2026 — classified as 1 death, 1 injury, 39 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (35 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem35
  • Air/Gas in Device2
  • Excess Flow or Over-Infusion2
  • Adverse Event Without Identified Device or Use Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions33
  • No Known Impact Or Consequence To Patient5
  • Chemical Exposure1
  • Death1
  • Foreign Body Embolism1

Recent adverse event reports

The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2026Air/Gas in Device

    AS REPORTED BY THE COMPLAINANT: TINY AIR BUBBLES IN PATIENTS PICC LINE AFTER MEROPENEM ADMINISTRATION. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675112 ↗
  • MalfunctionFeb 11, 2026Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: 1 X FOREIGN PARTICLE FLOATING INSIDE THE MEMBRANE (FLUID PATH) - (12112025@4) - POST FILLING. 1 X FOREIGN PARTICLE FLOATING INSIDE THE TUBING (FLUID PATH) - (12112025@4) - POST FILLING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24317480 ↗
  • MalfunctionOct 23, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ACCORDING TO THE COMPLAINANT, DURING DISINFECTION, PARTICLES WERE OBSERVED IN THE INFUSION LINE OF THE PUMP. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23366286 ↗
  • MalfunctionJul 11, 2023Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: EASYPUMP LEAKED AT CONNECTION BETWEEN BOTTOM PORT AND TUBING. LEAK FOUND DURING COMPOUNDING. IS THERE CHEMO OR BLOOD EXPOSURE TO PRODUCT: (YES) FOR PHARMACY TECHNICIANS NO INJURY REPORTED.

    Patient problems: Chemical Exposure

    MAUDE report 17301958 ↗
  • MalfunctionMay 9, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: 1 X BLACK PARTICLE IN THE CLOSING CONE IN THE PATIENT CONNECTOR (IN THE FLUID PATH) (09292022@1) - PRIOR TO FILLING. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16898029 ↗
  • InjuryApr 12, 2023Air/Gas in Device; Air/Gas in Device

    AS REPORTED BY THE USER FACILITY: "THE FAMILY INITIATED THE EASYPUMP INFUSION AND THE PATIENT STOPPED BREATHING AND FACE COLOR CHANGED TO BLUE/PURPLE. THEY STOPPED THE PUMP AND THE PATIENT QUICKLY RECOVERED. THE FAMILY CALLED THEIR PEDIATRICIAN, AND THE PEDIATRICIAN ASSUMED THE PATIENT HAD CHOKED ON SOMETHING OR WAS VOMITING. THE FAMILY DID NOT WANT TO RESTART THE INFUSION, SO THEY CALLED EMS. EMS CAME AND RESTARTED THE EASYPUMP AND COMPLETED THE INFUSION WITH NO ISSUE. WHEN IT CAME TIME FOR THE NEXT INFUSION, THE PARENTS AGAIN DID NOT WANT TO START FOR FEAR OF THE EVENT OCCURRING AGAIN, SO THEY TOOK THE PATIENT TO THE ER. I'M UNCLEAR IF THE ER STARTED AN EASYPUMP THAT WE PREFILLED AND SENT TO THE PATIENT OR IF THEY USED ONE OF THEIR OWN, BUT THE NEXT DOSE WAS INITIATED IN THE ER. IT WAS THE ER DOCTOR THAT MENTIONED AIR EMBOLISM TO THE PARENTS.". "THE PATIENT HAD BEEN IN THE HOSPITAL AND ON THIS MEDICATION, CEFTRIAXONE, FOR A FEW DAYS AND HAD BEEN RECEIVING IT AT HOME VIA EASYPUMP FOR A FEW DOSES BEFORE THIS EVENT OCCURRED. THE INITIAL EASYPUMP INFUSION WAS GIVEN BY OUR VISITING NURSE WHO EDUCATED THE PARENTS ON HOW TO INITIATE THE REMAINING INFUSIONS. THIS EVENT WAS NOT THE FIRST

    Patient problems: Foreign Body Embolism

    MAUDE report 16724575 ↗
  • MalfunctionOct 7, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: IT WAS REPORTED THAT FOREIGN MATTER WAS OBSERVED IN THE FLUID PATH PAST THE FILTER OF THE DEVICE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15563469 ↗
  • MalfunctionOct 4, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: IT WAS REPORTED THAT FOREIGN MATTER WAS OBSERVED IN THE FLUID PATH PAST THE FILTER OF THE DEVICE. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538480 ↗
  • MalfunctionFeb 25, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: BLACK PARTICLE IN THE CLOSING CONE IN THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13599949 ↗
  • MalfunctionFeb 14, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: SMALL BLACK PARTICLE IN THE PATIENT CONNECTOR (IN FLUID PATH).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13523914 ↗
  • MalfunctionJan 26, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY: BLACK PARTICLE IN THE PATIENT CONNECTOR UNDER THE CLOSING CONE (IN THE FLUID PATH).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13358738 ↗
  • MalfunctionDec 6, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: ORANGE PARTICLE IN THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12934840 ↗
  • MalfunctionNov 16, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: 1 WITH ORANGE PARTICLE IN PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12823818 ↗
  • MalfunctionOct 19, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 1 BLUE PARTICLE IN THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12656525 ↗
  • MalfunctionOct 19, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 2, BLACK PARTICLES ON THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12656522 ↗
  • MalfunctionOct 12, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: SMALL BLACK PARTICLE ON THE PATIENT CONNECTOR UNDER THE CLOSING CONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12613629 ↗
  • MalfunctionOct 12, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: DARK RED PARTICLE IN THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12613628 ↗
  • MalfunctionSep 23, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 1 -1 WITH WHITE PARTICLE IN THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12521285 ↗
  • MalfunctionSep 23, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 2 -1 WITH SMALL SHINY PARTICLE IN THE TUBING POST FILTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12521214 ↗
  • MalfunctionAug 23, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    DURING THE EVALUATION OF THE RETURNED SAMPLES, A PARTICLE WAS OBSERVED AT THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12355816 ↗
  • MalfunctionJul 26, 2021Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    AS REPORTED BY THE USER FACILITY: WE HAVE HAD 4540040-02 INFUSE OVER 15 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12225105 ↗
  • MalfunctionJul 19, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY: ORANGE PARTICLE UNDER THE PATIENT SIDE CLOSING CONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12188386 ↗
  • MalfunctionJul 6, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY: EVENT 2 LARGE BLACK PARTICLE UNDER THE CLOSING CONE IN THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12122586 ↗
  • MalfunctionJul 6, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY: EVENT 1: SMALL ORANGE PARTICLE IN THE PATIENT SIDE CLOSING CONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12122584 ↗
  • MalfunctionJul 6, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY: EVENT 3. BLACK PARTICLE IN THE CLOSING CONE IN THE PATIENT CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12122579 ↗

Other PUMP, INFUSION, ELASTOMERIC devices

All PUMP, INFUSION, ELASTOMERIC devices →