PTS-X™ 613743Recalls & reported problems
B Braun Interventional Systems, Inc · PTSX304 PTSX BALLOON 30MM X 4CM X 80CM · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964512094), received between Mar 17, 2017 and Mar 17, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Material Rupture1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2017Material Rupture
AS PER THE REPORT FROM BIS: "WHILE INFLATING THE BALLOON AT THE MAXIMUM SIZE, THE BALLOON RUPTURED. IT RUPTURED AT THE RATED BURST PRESSURE. ACCOUNT THEN WENT TO A BIGGER SIZED BALLOON. A PTSX404 WAS USED TO COMPLETE THE PROCEDURE. NO HARM TO THE PATIENT. THE CATHETER SHAFT WAS NOT KINKED. THERE WAS NOTHING UNUSUAL ABOUT THE PATIENT ANATOMY." A LATER FOLLOW-UP STATED: "IN AN EMAIL FROM THE ACCOUNT: THEY WERE USING THE BALLOON FOR VESSEL SIZING. AN INFLATION DEVICE WITH PRESSURE GAUGE WAS NOT USED, THEY USED A 60CC SYRINGE. NO PATIENT INFORMATION WILL BE PROVIDED."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6412700 ↗
FDA clearance and classification
Cleared through Special 510(k) K041306 on Jun 10, 2004 (decision: Substantially Equivalent). Submitted by NuMED, Inc. as “Numed Pts X Sizing Balloon Catheter”. The FDA review took 24 days. FDA record ↗
The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PTS-X™ 613743
Has the PTS-X™ 613743 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04046964512094), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PTS-X™ 613743?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Mar 17, 2017. That report lists device problem “Material Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PTS-X™ 613743?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PTS-X™ 613743 FDA approved?
It was cleared through Special 510(k) K041306 on Jun 10, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the PTS-X™ 613743, and is it still sold?
The FDA's device database lists B Braun Interventional Systems, Inc as the company responsible for it (brand PTS-X™, model 613743). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, intravascular occluding, temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page