Skip to content

APEX™ 601224Recalls & reported problems

B. Braun Medical Inc. · APEX, CONTROL MODULE · Class 2

All models of APEX (17)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04046964801068), received between Apr 12, 2019 and Sep 3, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Application Program Problem: Dose Calculation Error” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Application Program Problem: Dose Calculation Error2

Patient problems

  • No Patient Involvement2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2019Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS PER USER FACILITY: CUSTOMER REPORTED THEY HAD ADDED THE NEW SMOFLIPID TO THEIR SCDM DATABASE EARLIER IN THE DAY BUT THE COMPOUNDER WAS NOT PUMPING THE CORRECT AMOUNT FOR THE PATIENT ORDER. THE COMPOUNDER WAS PUMPING WATER INSTEAD OF THE SMOFLIPID SOLUTION. NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8955177 ↗
  • MalfunctionApr 12, 2019Application Program Problem: Dose Calculation Error; Application Program Problem: Dose Calculation Error

    AS PER USER FACILITY: CUSTOMER REPORTED AN ENCOUNTER AN ISSUE WHERE WHEN THE UNIVERSAL INGREDIENT (UI) IS SELECTED TO BE SOMETHING OTHER THAN WATER AND THE CONTAINER IS CHANGED DURING A MANIFOLD FLUSH THE SOFTWARE CHANGES THE UI TO LINE 1 WATER. IN THIS CASE THE UI WAS LINE 3 TROPHAMINE AND IT CHANGED TO LINE 1 WATER WHEN THE TROPHAMINE CONTAINER WAS CHANGED. THE UI WAS CHANGED BACK TO TROPHAMINE BY THE OPERATOR AND COMPOUNDING WAS ABLE TO CONTINUE. NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8508402 ↗

FDA clearance and classification

The FDA regulates product code NEP (system/device, pharmacy compounding) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Common questions about the APEX™ 601224

Has the APEX™ 601224 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04046964801068), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the APEX™ 601224?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 12, 2019 and Sep 3, 2019. In those reports, the most common specific device problem is “Application Program Problem: Dose Calculation Error” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the APEX™ 601224?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the APEX™ 601224 FDA approved?

Devices of this type (product code NEP) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by B. Braun Medical Inc.

Who makes the APEX™ 601224, and is it still sold?

The FDA's device database lists B. Braun Medical Inc. as the company responsible for it (brand APEX™, model 601224). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SYSTEM/DEVICE, PHARMACY COMPOUNDING devices

All SYSTEM/DEVICE, PHARMACY COMPOUNDING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page