BIB® 614451Recalls & reported problems
B Braun Interventional Systems, Inc · BB015 BIB Catheter 24 4 11F · Class 2
All models of BIB (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04046964900280), received between Apr 23, 2019 and Apr 23, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K050857, K160889); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device Markings/Labelling Problem1
- Inadequacy of Device Shape and/or Size1
- Inadequate User Interface1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2019Inadequacy of Device Shape and/or Size; Device Markings/Labelling Problem; Inadequate User Interface
ON (B)(6) 2019 NUMED RECEIVED A MEDWATCH REPORT FROM THE FDA THAT WAS SENT VIA USPS (MW5085855). THIS MEDWATCH REPORT WAS FROM THE USER FACILITY BUT LACKED USER FACILITY INFORMATION, CONTACT INFORMATION, AND AN ACCURATE LOT NUMBER FOR THE DEVICE. BASED ON THE STERILIZATION LOT NUMBER REFERENCED ON THE REPORT, NUMED WAS ABLE TO DETERMINE THE ACCURATE LOT NUMBER. NUMED CONTACTED THE DISTRIBUTOR IN THE U.S. FOR MORE INFORMATION ON THE DISTRIBUTION OF THE DEVICE AND RECEIVED THE FOLLOWING ADDITIONAL INFORMATION: "ON (B)(6) 2019 FROM THE ACCOUNT VIA EMAIL: "I AM CONTACTING YOU BECAUSE OF AN ISSUE THAT WE HAD WITH A BIB BALLOON DURING A PROCEDURE THIS WEEK AND JUST WANTED TO TALK TO YOU AND DISCUSS IT TO SEE IF THERE IS SOMETHING WE CAN DO TO RESOLVE A POTENTIAL PROBLEM IN THE FUTURE." BIS MEDICAL DEVICE SPECIALIST (MDS) SPOKE TO THE ACCOUNT AND SENT THIS MESSAGE: "MORE OF AN INVENTORY QUESTION AND NOT A PIR." ON (B)(6) 2019 - PER A CONVERSATION WITH BIS MDS AND PETER FLOSDORF: THE TECH PULLED AN INCORRECT SIZE BECAUSE OF LACK OF TRAINING AT THE ACCOUNT. THE TECH USED THE INNER BALLOON SIZE INSTEAD OF THE OUTER BALLOON SIZE. THE ACCOUNT WILL CONDUCT INTERNAL TRAINING TO RECTIFY. THIS IS
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8541022 ↗
FDA clearance and classification
Cleared through 510(k) K160889 on May 20, 2016 (decision: Substantially Equivalent). Submitted by NuMED, Inc. as “BIB Stent Placement Catheter”. The FDA review took 50 days. FDA record ↗
Cleared through 510(k) K050857 on Jul 28, 2005. Submitted by NuMED, Inc. as “Bib Pta Balloon Catheter”. The FDA review took 114 days. FDA record ↗
The FDA regulates product code NVM (catheter, angioplasty, peripheral, transluminal, dual-balloon) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K050857, K160889. One submission often covers many models, so check the product and lot details in each record.
- Z-1754-2025 · Apr 21, 2025 · TerminatedBIB Balloon in Balloon Dilatation Catheter. Model Number: 420
Common questions about the BIB® 614451
Has the BIB® 614451 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04046964900280), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the BIB® 614451?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 23, 2019. That report lists device problems “Device Markings/Labelling Problem” (1), “Inadequacy of Device Shape and/or Size” (1) and “Inadequate User Interface” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BIB® 614451?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BIB® 614451 FDA approved?
It was cleared through 510(k) K050857 on Jul 28, 2005; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the BIB® 614451, and is it still sold?
The FDA's device database lists B Braun Interventional Systems, Inc as the company responsible for it (brand BIB®, model 614451). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, angioplasty, peripheral, transluminal, dual-balloon devices
- BIB® 614396B Braun Interventional Systems, Inc · Catheter, angioplasty, peripheral, transluminal, dual-balloon1 recall0 reports
- BIB® 614444B Braun Interventional Systems, Inc · Catheter, angioplasty, peripheral, transluminal, dual-balloon1 report
- BIB® 614446B Braun Interventional Systems, Inc · Catheter, angioplasty, peripheral, transluminal, dual-balloon1 report
- BIB® 614445B Braun Interventional Systems, Inc · Catheter, angioplasty, peripheral, transluminal, dual-balloon1 report
- BIB® 614440B Braun Interventional Systems, Inc · Catheter, angioplasty, peripheral, transluminal, dual-balloon1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page