Design Options® 552116Recalls & reported problems
B. Braun Medical Inc. · Continuous Epidural Tray · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04046964950254), received between May 2, 2019 and Sep 18, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Sep 18, 2025Open, Classified
Potential for the lid of the catheter connector to be in the incorrect position.
Z-0126-2026 · Root cause: Process control · FDA record ↗
- Jun 5, 2023Open, Classified
Kits were assembled with the incorrect Filter Straw.
Z-2003-2023 · Root cause: Mixed-up of materials/components · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Break2
- Device Contamination with Chemical or Other Material1
- Difficult to Advance1
- Difficult to Remove1
Patient problems
- No Known Impact Or Consequence To Patient4
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Device Contamination with Chemical or Other Material
When opening an epidural kit, a hair was noticed under the gauze and on the sterile drape. The kit was closed and the patient did not come in contact with it. A new kit was obtained, and the patient was prepped and draped. Epidural was placed without incident. The contaminated kit was saved for reporting.No Pt Harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817653 ↗MalfunctionMay 12, 2025Break; Difficult to Advance; Break; Difficult to Advance
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: CATHETER SHEARING. DETAILED INQUIRY DESCRIPTION UPON CATHETER REMOVAL THE CATHETER APPEARED SHEARED, AND TIP IS MISSING. NO INJURY REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 22005515 ↗MalfunctionOct 30, 2024Break; Difficult to Remove; Break; Difficult to Remove
AS REPORTED BY USER FACILITY: REGISTERED NURSE (RN) ATTEMPTED TO REMOVE EPIDURAL CATHETER FROM PATIENT'S BACK WHILE PATIENT ON HER SIDE AND MET SIGNIFICANT RESISTANCE. HAD PATIENT CURL AROUND ABDOMEN TO OPEN THE SPINAL SPACE. STILL UNABLE TO REMOVE EPIDURAL CATHETER. RN CALLED ANESTHESIA PROVIDER TO COME EVALUATE. ANESTHESIA PROVIDER'S ATTEMPT TO REMOVE EPIDURAL CATHETER AND ALSO MET SIGNIFICANT RESISTANCE. DUE TO PATIENT NOT HAVING MUCH FEELING IN HER LEGS ANESTHESIA PROVIDER DETERMINED THAT MORE TIME SHOULD BE GIVEN FOR SENSATION TO COME BACK TO PATIENT'S LEGS. 30 MIN LATER ANESTHESIA PROVIDER RETURNED TO PATIENT'S BEDSIDE TO REASSESS PATIENT'S SENSATION IN LEGS. SENSATION HAD IMPROVED AND ANESTHESIA PROVIDER ATTEMPTED TO REMOVE EPIDURAL CATHETER WHILE PATIENT WAS SITTING UP HUNCHED OVER ABDOMEN. ANESTHESIA PROVIDER UNABLE TO REMOVE EPIDURAL CATHETER. HAS PATIENT GET ONTO HANDS AND KNEES AND WAS ABLE TO REMOVE EPIDURAL CATHETER AT THIS POINT, BUT THE TIP OF THE CATHETER DID NOT COME OUT. THE CATHETER APPEARED TO HAVE UNCOILED WHERE IT BROKE OFF AND ANESTHESIA PROVIDER ATTEMPTED TO REMOVE THE REST OF THE CATHETER BUT WAS UNSUCCESSFUL. ANESTHESIA PROVIDER DETERMINED THAT PATIENT WA
Patient problems: Foreign Body In Patient
MAUDE report 20567771 ↗MalfunctionMay 2, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
AS PER USER FACILITY, EVENT 4: FACILITY REPORTED CATHETER CONNECTORS ARE DECLASPING. ABOUT 4 TRAYS HAVE BEEN AFFECTED. NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8574593 ↗MalfunctionMay 2, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
AS PER USER FACILITY, EVENT 2: FACILITY REPORTED CATHETER CONNECTORS ARE ENCLASPING. ABOUT 4 TRAYS HAVE BEEN AFFECTED. NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8574592 ↗MalfunctionMay 2, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
AS PER USER FACILITY, EVENT 1: FACILITY REPORTED CATHETER CONNECTORS ARE ENCLASPING. ABOUT 4 TRAYS HAVE BEEN AFFECTED. NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8574590 ↗MalfunctionMay 2, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
AS PER USER FACILITY, EVENT 3: FACILITY REPORTED CATHETER CONNECTORS ARE ENCLASPING. ABOUT 4 TRAYS HAVE BEEN AFFECTED. NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8574588 ↗
Other Epidural anesthesia kit devices
- Design Options® 552177B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552179B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Design Options® 552126B. Braun Medical Inc. · Epidural anesthesia kit2 recalls0 reports
- Perifix® 332079B. Braun Medical Inc. · Epidural anesthesia kit1 recall45 reports
- Perifix® 332097B. Braun Medical Inc. · Epidural anesthesia kit1 recall39 reports
- Perifix® 332220B. Braun Medical Inc. · Epidural anesthesia kit1 recall21 reports
- Perifix® 332078B. Braun Medical Inc. · Epidural anesthesia kit1 recall12 reports
- Design Options® 552155B. Braun Medical Inc. · Epidural anesthesia kit1 recall7 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.