Skip to content

n.a. 27290MRecalls & reported problems

Karl Storz GmbH & Co. KG · Forceps · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04048551082508), received between Jan 9, 2020 and Jun 26, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Material Separation1

Patient problems

  • Insufficient Information3
  • Foreign Body In Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE INFORMATION RECEIVED, THE GRASPER, FROM ONE OF THE PCNL SETS HAD BROKEN DURING USE. REPORTEDLY, THE INSTRUMENT WAS FUNCTIONING NORMALLY AT THE START OF THE CASE, BUT LATER THE GRASPING MECHANISM STOPPED WORKING. UPON INSPECTION, IT WAS CLEAR THE RIVET / SCREW THAT SECURED ONE SIDE OF THE JAW MECHANISM WAS MISSING. THE SCRUB TEAM HAD ENDEAVORED TO FIND THE MISSING PIECE, BUT WERE UNABLE TO DO SO, LATER ON, IN THE POUCH OF THE OPERATION DRAPE WE HAVE FOUND A METAL PIECE WHICH COULD POSSIBLE BE THE MISSING PART OF THE FORCEPS, X RAY TAKEN. SINCE ADDITIONAL X RAY TAKEN, THIS CASE IS REPORTABLE

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 22340857 ↗
  • InjuryNov 3, 2023Break; Break

    IT WAS REPORTED THAT THE FORCEPS BECAME DEFECTIVE DURING THE VERY FIRST PCNL SURGERY THE FORCEPS WERE USED FOR. ONE OF THE GRASPER JAWS HAD BECOME LOOSE FROM THE JOINT AFTER ONE STONE HAS BEEN REMOVED DURING A PCNL PROCEDURE. THE GRASPING FORCEPS HAD BEEN REMOVED FROM THE SURGERY AT ONCE. IT WAS REPORTED THAT THE FORCEPS SHAFT CAME OFF DURING A PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. AS THE PIN OF THE JOINT IS MISSING THIS CASE DEEMED AS REPORTABLE AS PRECAUTIONARY ACTION.

    Patient problems: Insufficient Information

    MAUDE report 18067971 ↗
  • InjuryNov 2, 2023Break; Break

    IT WAS REPORTED THAT THE FORCEPS BECAME DEFECTIVE DURING THE VERY FIRST PCNL SURGERY THE FORCEPS WERE USED FOR. ONE OF THE GRASPER JAWS HAD BECOME LOOSE FROM THE JOINT AFTER ONE STONE HAS BEEN REMOVED DURING A PCNL PROCEDURE. THE GRASPING FORCEPS HAD BEEN REMOVED FROM THE SURGERY AT ONCE. IT WAS REPORTED THAT THE FORCEPS SHAFT CAME OFF DURING A PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. AS THE PIN OF THE JOINT IS MISSING THIS CASE DEEMED AS REPORTABLE AS PRECAUTIONARY ACTION.

    Patient problems: Insufficient Information

    MAUDE report 18060105 ↗
  • MalfunctionJan 9, 2020Material Separation; Material Separation

    PER THE FACTORY IN (B)(6), ALLEGEDLY, THERE WAS AN EVENT OCCURRED IN (B)(6) DURING A PROCEDURE, A PIECE BROKE OFF AND FELL INSIDE THE PATIENT; IT IS UNCLEAR IF THE BROKEN PIECE REMAINED IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 9570073 ↗

Other Kit, nephroscope devices

All Kit, nephroscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page