Karl Storz GmbH & Co. KG · Fiber Optic Light Cable, 230 cm, Ø3.5 mm · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04048551168868), received between Mar 16, 2023 and Feb 4, 2026 — classified as 9 injuries.

The most frequently reported specific device problem is “Excessive Heating” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Excessive Heating3
  • Detachment of Device or Device Component2
  • Insufficient Device Problem Information2
  • Mechanical Jam1
  • Overheating of Device1
  • Temperature Problem1

Patient problems

  • Burn(s)7
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 4, 2026Insufficient Device Problem Information; Excessive Heating

    THE FOLLOWING WAS REPORTED: "A MEMBER OF STAFF BURNT THEMSELVES ON THE PROXIMAL END OF THE LIGHT GUIDE CABLE BY HOLDING ONTO THE INSULATED SECTION, CAUSING THE PROXIMAL END TO SWING BACK AND TOUCH THEIR ARM. THIS HAPPENED AROUND 4 WEEKS AGO BUT THE EXACT DATE AND GUIDE CABLE IN USE IS UNKNOWN.

    Patient problems: Burn(s)

    MAUDE report 24260327 ↗
  • InjuryJan 30, 2026Excessive Heating

    THE FOLLOWING WAS REPORTED: "A MEMBER OF STAFF BURNT THEMSELVES ON THE PROXIMAL END OF THE LIGHT GUIDE CABLE BY HOLDING ONTO THE INSULATED SECTION, CAUSING THE PROXIMAL END TO SWING BACK AND TOUCH THEIR ARM. THIS HAPPENED AROUND 4 WEEKS AGO BUT THE EXACT DATE AND GUIDE CABLE IN USE IS UNKNOWN.

    Patient problems: Burn(s)

    MAUDE report 24220194 ↗
  • InjuryJan 29, 2026Excessive Heating

    IT WAS REPORTED THAT THE LIGHT CABLE WAS RESTING ON THE SURGICAL DRAPE WHILE THE SCOPES WERE BEING CHANGED. HEAT FROM THE CABLE BURNED THROUGH THE DRAPES, RESULTING IN A BURN TO THE PATIENT'S SKIN. SINCE THE PATIENT GOT INJURED, THE CASE IS DEEMED AS REPORTABLE.

    Patient problems: Burn(s)

    MAUDE report 24215230 ↗
  • InjuryDec 5, 2025Mechanical Jam; Mechanical Jam

    ACCORDING TO THE INFORMATION RECEIVED, A BILATERAL FLEXIBLE URETEROSCOPY WAS PERFORMED USING A DISPOSABLE FLEXIBLE URETEROSCOPE. DURING INSERTION, THE CABLE ON THE EQUIPMENT STOPPED DEFLECTING AND THE SURGERY HAD TO BE SUSPENDED. ANOTHER PIECE OF EQUIPMENT WAS BROUGHT TO THE TABLE, BUT THE CABLE ON THAT EQUIPMENT ALSO BROKE. THIS DELAYED THE PROCEDURE AND ANOTHER PIECE OF EQUIPMENT HAD TO BE SET UP. ON (B)(6) 2025, IT WAS MENTIONED BY THE CUSTOMER THAT THE WHOLE PROCEDURE HAD TO BE CANCELLED AFTER THE ABOVE-MENTIONED EVENT. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23729889 ↗
  • InjuryOct 29, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT DURING A CYSTOSCOPY ON (B)(6) 2024, THE PATIENT WAS BURNED BY THE LIGHT CORD. THE LIGHT SOURCE WAS ON BUT NOT ATTACHED TO THE SCOPE. THE LIGHT CORD WAS DRAPED ACROSS THE PATIENT AND THEY SUSTAINED A BURN.

    Patient problems: Burn(s)

    MAUDE report 20562204 ↗
  • InjuryJun 6, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A HYSTEROSCOPY AND IUD REMOVAL PROCEDURE ON (B)(6) 2024, THE CABLE WAS FOUND DETACHED FROM THE SCOPE AND WAS LAYING ON TOP OF THE PATIENT. AFTER THE PROCEDURE WAS COMPLETED, A SUPERFICIAL BURN WAS FOUND ON THE PATIENT'S ANTERIOR SIDE. A TOPICAL SOLUTION WAS ADMINISTERED TO TREAT THE SUPERFICAL BURN. THE SCOPE THAT WAS BEING USED WAS A GYNECARE SCOPE, NOT A KARL STORZ SCOPE. THE CABLE WILL NOT BE SENT IN FOR EVALUATION AT THIS TIME PER THE CUSTOMER.

    Patient problems: Burn(s); Superficial (First Degree) Burn

    MAUDE report 19477391 ↗
  • InjuryJun 6, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A HYSTEROSCOPY AND IUD REMOVAL PROCEDURE ON APRIL-30-2024, THE CABLE WAS FOUND DETACHED FROM THE SCOPE AND WAS LAYING ON TOP OF THE PATIENT. AFTER THE PROCEDURE WAS COMPLETED, A SUPERFICIAL BURN WAS FOUND ON THE PATIENT'S ANTERIOR SIDE. A TOPICAL SOLUTION WAS ADMINISTERED TO TREAT THE SUPERFICAL BURN. THE SCOPE THAT WAS BEING USED WAS A GYNECARE SCOPE, NOT A KARL STORZ SCOPE. THE CABLE WILL NOT BE SENT IN FOR EVALUATION AT THIS TIME PER THE CUSTOMER.

    Patient problems: Burn(s)

    MAUDE report 19477387 ↗
  • InjuryJun 15, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT AN EVENT WITH THE 495NA FIBER OPTIC CABLE. ACCORDING TO THE INFORMATION RECEIVED, WHILE USING THE LIGHT CABLE ON A PATIENT, THE SCOPE ATTACHED TO THE LIGHT CABLE OVERHEATED INJURING THE PATIENT. PATIENT SUFFERED WHAT APPEARED TO BE THERMAL BURN 1ST DEGREE. PER THE FACILITY THE SCOPE AND LIGHT CABLE WERE USED DURING THE ENTIRE SURGICAL PROCEDURE, APPROXIMATELY 5 HOURS. THE PATIENT SUFFERED WHAT APPEARED TO BE A THERMAL BURN 1ST DEGREE.

    Patient problems: Insufficient Information

    MAUDE report 17136878 ↗
  • InjuryMar 16, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    AS PER MEDWATCH REPORT MW5114850 RECEIVED VIA EMAIL: A LAPAROSCOPIC CASE WITH FIBER OPTIC LIGHT CABLE RESULTED IN BURN TO PATIENT. LIGHT CABLE HEATED UP AND TRANSFERRED HEAT TO THE LAPAROSCOPE, RESULTING IN A BURN. THE RECEIVED MEDWATCH REPORT CONTAINED MINIMAL INFORMATION. THE ITEM NUMBER, LOT CODE, FACILITY NAME/LOCATION AND ADDITIONAL PATIENT INFORMATION WAS NOT PROVIDED. WE ARE FILING THIS REPORT REFERENCING A COMMONLY USED FIBER OPTIC LIGHT CABLE, MODEL NUMBER 495NA.

    Patient problems: Burn(s)

    MAUDE report 16553123 ↗

Other Light source, fiberoptic, routine devices

All Light source, fiberoptic, routine devices →