Skip to content

n.a. 495NACRecalls & reported problems

Karl Storz GmbH & Co. KG · Fiber Optic Light Cable, 230cm, Ø3.5mm · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04048551290705), received between Jun 2, 2022 and Sep 7, 2023 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Misassembly by Users” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Misassembly by Users2
  • Use of Device Problem1

Patient problems

  • Burn(s)2
  • Superficial (First Degree) Burn2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 7, 2023Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE LIGHT WAS SWITCHED ON WITHOUT THE OTHER END OF THE CORD BEING PLUGGED INTO THE SCOPE. THIS RESULTED IN A SMALL HOLE IN THE DRAPE NEAR THE PATIENT'S LOWER EXTREMITIES. THE PATIENT WAS ASSESSED, AND NO HARM HAD REACHED THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17703269 ↗
  • InjuryJun 7, 2022Misassembly by Users; Misassembly by Users

    AT THE BEGINNING OF A HYSTERECTOMY PROCEDURE THE SURGEON NOTICED THE LIGHT CABLE WAS NOT ATTACHED TO THE SCOPE. THE SCOPE WAS ATTACHED AND AFTER THE PROCEDURE IN RECOVERY A NOTICEABLE "BLISTER LOOKING "MARK WAS DISCOVERED ON PATIENT'S RIGHT THIGH. BURN WAS ASSESSED BY PLASTIC SURGEON TO BE A "SUPERFICIAL BURN" AND RECOMMENDED ANTIBIOTIC OINTMENT.

    Patient problems: Burn(s); Superficial (First Degree) Burn

    MAUDE report 14626157 ↗
  • InjuryJun 2, 2022Misassembly by Users; Misassembly by Users

    AT THE START OF A HYSTERECTOMY PROCEDURE WITH DILATION AND CURETTAGE, THE LIGHT CABLE WAS NOTICED NOT TO BE CONNECTED TO THE SCOPE BUT LAID ON THE DRAPE. A BURN HOLE WAS FOUND ON THE DRAPE WHICH HAD BURNED THROUGH TO THE PATIENT'S SKIN ON RIGHT INNER THIGH.

    Patient problems: Burn(s); Superficial (First Degree) Burn

    MAUDE report 14576496 ↗

Other Light source, fiberoptic, routine devices

All Light source, fiberoptic, routine devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page