Karl Storz GmbH & Co. KG · 24/26FR BIPOLAR CUTTING LOOP · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Karl Storz bipolar electrosurgical cutting loop is used to cut and coagulate tissue during hysteroscopic surgery inside the uterus. The 30 reports mostly describe the loop's metal tip breaking, cracking or shedding fragments during procedures, sometimes leaving a piece in the uterine cavity, and one report describes the cable burning through and briefly shocking the patient. Reports mention 7 injuries and no deaths, and there have been no recalls of this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 04048551392720), received between Jul 14, 2017 and Aug 3, 2026 — classified as 7 injuries, 23 malfunctions.

The most frequently reported specific device problem is “Material Separation” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 4 reportsSep 2025: 0 reportsOct 2025: 5 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation12
  • Break7
  • Detachment of Device or Device Component5
  • Material Fragmentation5
  • Fracture2
  • Sparking2
  • Battery Problem1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Insufficient Information7
  • No Consequences Or Impact To Patient6
  • Electric Shock1
  • Fibrosis1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 3, 2026Insufficient Device Problem Information

    It was reported that on (b)(6) 2026, according to the standard surgical procedure, the patient was scheduled to undergo hysteroscopic endometrial lesion resection + hysteroscopic cervical lesion resection + fractional diagnostic curettage + cervical biopsy. During the operation, while using the high-frequency electrosurgical device, the working handpiece suddenly burned through the bipolar electrode cable, causing the patient to experience a brief electric shock. The use of the device was immediately discontinued, and a backup working handpiece and electrode cable were replaced. After confirming no issues, the surgery continued.New additional information was received on July 24, 2026. Based on the new information there are 3 additional potentially causal KARL STORZ products involved in the event and additional reported error description: "The sealing ring leaks, and water flows through the gap to the electrode cable during the procedure."Cross reference MDR:9610617-2026-01599,9610617-2026-01798,9610617-2026-01799.

    Patient problems: Electric Shock

    MAUDE report 50122170 ↗
  • MalfunctionMay 7, 2026Break; Fracture

    IT WAS REPORTED THAT "DURING HYSTEROSCOPIC SURGERY, THE DOCTOR USED AN ELECTROSURGICAL CUTTING LOOP TO SEPARATE INTRAUTERINE TISSUE AND COAGULATE BLEEDING POINTS. WHILE THE CUTTING LOOP WAS IN USE INSIDE THE UTERINE CAVITY, A CIRCUMFERENTIAL FRACTURE OCCURRED AT THE FRONT 1.5 CM × 1.5 CM SECTION OF THE LOOP." ACCORDING TO THE AVAILABLE INFORMATION THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A PROLONGATION OF 10 MINUTES AND THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25115813 ↗
  • MalfunctionJan 29, 2026Break

    IT WAS REPORTED THAT THE TIP BROKE OFF WHILE RESECTING FIBROID IN THE UTERUS. NO PATIENT WAS HARMED. ADDITIONALLY, THEY WERE USING A COVIDIEN FT10 GENERATOR FOR THIS PROCEDURE AND THEY WERE ABLE TO FLUSH THE BROKEN PIECE OUT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24211548 ↗
  • MalfunctionDec 19, 2025Fracture; Sparking; Fracture; Sparking

    IT WAS REPORTED THAT THE PATIENT WAS ADMITTED WITH THE CHIEF COMPLAINT OF "PROLONGED MENSTRUAL PERIODS FOR OVER 9 YEARS." ON NOVEMBER 20, UNDER INTRAVENOUS COMBINED ANESTHESIA, THE PATIENT UNDERWENT HYSTEROSCOPIC RESECTION OF A UTERINE LESION (ENDOMETRIAL POLYP RESECTION) + DIAGNOSTIC DILATION AND CURETTAGE + REMOVAL OF AN INTRAUTERINE CONTRACEPTIVE DEVICE. AFTER ROUTINE DRAPING, THE HYSTEROSCOPE WAS INSERTED. WHILE PREPARING FOR ELECTROSURGICAL RESECTION DURING THE PROCEDURE, THE RESECTOSCOPE LOOP SUDDENLY SPARKED. THE HYSTEROSCOPE WAS IMMEDIATELY WITHDRAWN FOR INSPECTION, WHICH REVEALED A FRACTURE AT THE CONNECTOR OF THE DISPOSABLE HYSTEROSCOPIC RESECTOSCOPE. THE PROCEDURE CONTINUED AS PLANNED AFTER REPLACING THE DISPOSABLE HYSTEROSCOPIC RESECTOSCOPE. IT WAS CONFIRMED THAT THE PRODUCT CONCERNED HAS BEEN SCRAPPED BY THE HOSPITAL AFTER BEING DAMAGED AND CANNOT BE RETURNED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23851366 ↗
  • InjuryOct 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT "A HYSTEROSCOPIC RESECTOSCOPE AND ITS MATCHING RESECTOSCOPE LOOP WERE USED DURING THE PROCEDURE. WHILE RESECTING THE INTRAUTERINE LESION, THE SURGEON OBSERVED A METALLIC STRIP-LIKE OBJECT WITHIN THE UTERINE CAVITY UNDER THE VISUAL FIELD. IMMEDIATE INSPECTION OF THE RESECTOSCOPE LOOP REVEALED THAT APPROXIMATELY 5 MM OF THE FRONT-END METAL RING HAD DETACHED, THOUGH THE INTERNAL WIRE REMAINED CONNECTED. THE DAMAGED LOOP REMNANT WAS PROMPTLY REMOVED ON THE OPERATING TABLE. HOWEVER, DURING RETRIEVAL NEAR THE VAGINAL FORNIX, THE REMNANT DISAPPEARED FROM THE VISUAL FIELD. IT WAS INITIALLY SUSPECTED THAT THE REMNANT HAD BEEN FLUSHED OUT BY THE HYSTEROSCOPIC DISTENSION FLUID, BUT REPEATED EXAMINATIONS OF THE EXTERNAL FLUID COLLECTION BUCKET DID NOT LOCATE THE RESIDUAL FRAGMENT. A C-ARM WAS IMMEDIATELY USED FOR FLUOROSCOPIC EXAMINATION, WHICH CONFIRMED THAT THE METALLIC REMNANT REMAINED WITHIN THE PATIENT'S VAGINA. A SECONDARY SURGICAL INSPECTION WAS IMMEDIATELY PERFORMED, AND THE METALLIC STRIP WAS EVENTUALLY FOUND ON A STERILE GAUZE SPONGE IN THE VAGINAL FORNIX. AFTER SUCCESSFUL RETRIEVAL, COMPARISON CONFIRMED THAT ITS LENGTH AND EDGES MATCHED THE DAMAGED SECTION OF TH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23438651 ↗
  • InjuryOct 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT "A HYSTEROSCOPIC RESECTOSCOPE AND ITS MATCHING RESECTOSCOPE LOOP WERE USED DURING THE PROCEDURE. WHILE RESECTING THE INTRAUTERINE LESION, THE SURGEON OBSERVED A METALLIC STRIP-LIKE OBJECT WITHIN THE UTERINE CAVITY UNDER THE VISUAL FIELD. IMMEDIATE INSPECTION OF THE RESECTOSCOPE LOOP REVEALED THAT APPROXIMATELY 5 MM OF THE FRONT-END METAL RING HAD DETACHED, THOUGH THE INTERNAL WIRE REMAINED CONNECTED. THE DAMAGED LOOP REMNANT WAS PROMPTLY REMOVED ON THE OPERATING TABLE. HOWEVER, DURING RETRIEVAL NEAR THE VAGINAL FORNIX, THE REMNANT DISAPPEARED FROM THE VISUAL FIELD. IT WAS INITIALLY SUSPECTED THAT THE REMNANT HAD BEEN FLUSHED OUT BY THE HYSTEROSCOPIC DISTENSION FLUID, BUT REPEATED EXAMINATIONS OF THE EXTERNAL FLUID COLLECTION BUCKET DID NOT LOCATE THE RESIDUAL FRAGMENT. A C-ARM WAS IMMEDIATELY USED FOR FLUOROSCOPIC EXAMINATION, WHICH CONFIRMED THAT THE METALLIC REMNANT REMAINED WITHIN THE PATIENT'S VAGINA. A SECONDARY SURGICAL INSPECTION WAS IMMEDIATELY PERFORMED, AND THE METALLIC STRIP WAS EVENTUALLY FOUND ON A STERILE GAUZE SPONGE IN THE VAGINAL FORNIX. AFTER SUCCESSFUL RETRIEVAL, COMPARISON CONFIRMED THAT ITS LENGTH AND EDGES MATCHED THE DAMAGED SECTION OF TH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23438649 ↗
  • MalfunctionOct 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LOOP BROKE OFF DURING THE CASE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDRS: 9610617-2025-01829. 9610617-2025-01830.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351836 ↗
  • MalfunctionOct 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LOOP BROKE OFF DURING THE CASE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDRS: 9610617-2025-01830. 9610617-2025-01831.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350626 ↗
  • MalfunctionOct 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LOOP BROKE OFF DURING THE CASE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDRS: 9610617-2025-01829. 9610617-2025-01831.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350625 ↗
  • InjuryAug 11, 2025Material Separation; Material Separation

    IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2025 TWO (2) ELECTRODES BROKE. ACCORDING TO THE DOCTOR, A PIECE WAS RETAINED IN THE PATIENT AND THEY DID NOT SAVE THE BROKEN ELECTRODES. CROSS REFERENCE MEDWATCH NUMBER 2020550-2025-01400.

    Patient problems: Insufficient Information

    MAUDE report 22777209 ↗
  • InjuryAug 11, 2025Material Separation; Material Separation

    IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2025 TWO (2) ELECTRODES BROKE. ACCORDING TO THE DOCTOR, A PIECE WAS RETAINED IN THE PATIENT AND THEY DID NOT SAVE THE BROKEN ELECTRODES. CROSS REFERENCE MEDWATCH NUMBER 2020550-2025-01399.

    Patient problems: Insufficient Information

    MAUDE report 22777207 ↗
  • InjuryAug 11, 2025Material Separation; Material Separation

    IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2025 TWO (2) ELECTRODES BROKE. ACCORDING TO THE DOCTOR, A PIECE WAS RETAINED IN THE PATIENT AND THEY DID NOT SAVE THE BROKEN ELECTRODES. CROSS REFERENCE MEDWATCH NUMBER 9610617-2025-01400.

    Patient problems: Insufficient Information

    MAUDE report 22777161 ↗
  • InjuryAug 11, 2025Material Separation; Material Separation

    IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2025 TWO (2) ELECTRODES BROKE. ACCORDING TO THE DOCTOR, A PIECE WAS RETAINED IN THE PATIENT AND THEY DID NOT SAVE THE BROKEN ELECTRODES. CROSS REFERENCE MEDWATCH NUMBER 9610617-2025-01399.

    Patient problems: Insufficient Information

    MAUDE report 22777151 ↗
  • MalfunctionApr 21, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT EQUIPMENT BROKE, IDENTIFIED DURING PROCEDURE BY ATTENDING WHILE OPERATING, NOTICED ON CAMERA SCOPE BIPOLAR CUTTING LOOP NOT WORKING PROPERLY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21873927 ↗
  • MalfunctionNov 18, 2024Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2024, THE DOCTOR WAS REMOVING BLADDER TUMOR AND NOTICED THERE WAS A BROKEN PIECE OF THE SHEATH INSIDE OF THE PATIENT. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. NO HARM WAS CAUSED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20712258 ↗
  • MalfunctionSep 18, 2024Sparking; Battery Problem; Sparking; Battery Problem

    IT WAS REPORTED DURING A MEDICAL PROCEDURE, THE UNIPOLAR HIGH FREQUENCY CORD POPPED AND SPARKED. NO INJURT WAS NOTED TO THE PHYSICIAN OR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20246919 ↗
  • MalfunctionAug 14, 2024Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED DURING A TURBT PROCEDURE ON (B)(6) 2024, THE END OF THE ELECTRODE FELL INTO THE PATIENT'S BLADDER, THE DOCTOR WAS ABLE TO FLUSH THE ELECTRODE OUT. THE PROCEDURE COMPLETED WITH A NEW STERILE SINGLE USE BIPOLAR CUTTING LOOP. NO INJURIES WERE REPORTED AND NO SURGICAL DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19985902 ↗
  • MalfunctionJun 27, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A TRANSURETHRAL RESECTION OF PROSTATE PROCEDURE ON (B)(6) 2024, THE CUTTING LOOP BROKE OFF INSIDE OF THE PATIENT. THEY WERE ABLE TO REMOVE THE BROKEN PIECE USING A BLADDER EVACUATOR, AND WERE ABLE TO COMPLETE THE PROCEDURE USING A BACK-UP WITHOUT ANY PATIENT INJURY. THIS ISSUE CAUSED A DELAY OF ABOUT 10 MINUTES TO THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19629562 ↗
  • MalfunctionJun 27, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A BLADDER TUMOR REMOVAL PROCEDURE ON (B)(6) 2024, THE CUTTING LOOP BROKE OFF. THEY WERE ABLE TO COMPLETE THE PROCEDURE WITHOUT ANY PATIENT INJURY, AND WERE ALSO ABLE TO RETRIEVE THE BROKEN PIECE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19629241 ↗
  • MalfunctionMay 13, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A HYSTEROSCOPIC POLYPECTOMY OR MYOMECTOMY PROCEDURE ON (B)(6) 2024, THE BIPOLAR CUTTING LOOP BROKE INSIDE OF THE PATIENT. THE YELLOW PIECE WAS RETRIEVED, HOWEVER THE METAL TIP WAS NOT RECOVERED AND WAS PRESUMED LEFT INSIDE THE PATIENT. NO PATIENT INJURY WAS REPORTED. PATIENT IS HOME AND DOING WELL. THE NEXT STEP IS BEING DECIDED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19299392 ↗
  • MalfunctionNov 24, 2021Break

    DURING PROCEDURE, BIPOLAR LOOP BROKE AT SHAFT.

    Patient problems: Insufficient Information

    MAUDE report 12868561 ↗
  • MalfunctionNov 12, 2020Material Separation; Material Separation

    ALLEGEDLY, DURING A TURP PROCEDURE, THE BIPOLAR ELECTRODE LOOP BROKE INSIDE THE PATIENT. THE SURGEON REPORTED THAT THE FRAGMENTED PIECES WERE REMOVED. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED REGARDING THE OUTCOME TO THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 10830451 ↗
  • MalfunctionOct 1, 2020Break

    WHILE PERFORMING TURP PROCEDURE, THE SURGEON REPORTED THAT THE BIPOLAR CUTTING LOOP BROKE INTO PIECES INSIDE THE PATIENT. THE FRAGMENTED PIECES WERE REMOVED. MANUFACTURER RESPONSE FOR COAGULATOR-CUTTER, ENDOSCOPIC, BIPOLAR (AND ACCESSORIES), N.A. (PER SITE REPORTER). THE REPRESENTATIVE WAS NOTIFIED VIA TELEPHONE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10612661 ↗
  • MalfunctionJul 23, 2020Material Separation; Material Separation

    ALLEGEDLY, DURING A HYSTEROSCOPIC MYOMECTOMY PROCEDURE, 2-3 MM OF THE BIPOLAR ELECTRODE LOOP BROKE DUE TO THE RESISTANCE OF THE LARGE FIBROMA. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED REGARDING THE OUTCOME TO THE PATIENT.

    Patient problems: Fibrosis

    MAUDE report 10317093 ↗
  • MalfunctionJan 17, 2019Material Separation; Material Separation

    ALLEGEDLY, DURING A SEPTOPLASTY PROCEDURE, ONE OF THE JAWS BROKE OFF INSIDE THE PATIENT. THE PIECE WAS RETRIEVED AND THERE WAS NOT HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8259125 ↗

Other COAGULATOR-CUTTER, ENDOSCOPIC, BIPOLAR (AND ACCESSORIES) devices

All COAGULATOR-CUTTER, ENDOSCOPIC, BIPOLAR (AND ACCESSORIES) devices →