NA 40712035Recalls & reported problems
Karl Storz GmbH & Co. KG · DRILLCUT-X II-35 Shaver Handpiece · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Karl Storz shaver handpiece is a powered surgical drill used in ear, nose and throat procedures. The 24 reports naming its identifier, with no deaths and no injuries, mostly describe the handpiece jamming or freezing during a procedure, shutting down unexpectedly, or blade rotation becoming unstable or stopping. No recalls name this device. Reports rose
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 04048551409954), received between Aug 19, 2025 and Sep 30, 2026 — classified as 28 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam13
- Output Problem6
- Unexpected Shutdown4
- Intermittent Energy Output2
- Appropriate Device Problem Term/Code Not Available1
- Defective Component1
- Electrical /Electronic Property Problem1
- Erratic or Intermittent Display1
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- Insufficient Information3
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954093 ↗MalfunctionSep 30, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954090 ↗MalfunctionSep 30, 2026Mechanical Jam
It was reported "Getting jammed during surgery". No negative impact in state of health reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942505 ↗MalfunctionSep 30, 2026Output Problem
It was reported "When the rotation speed was set to 6,000 rpm and the device was run idle 5 to 6 times as instructed, the error message "E19: Motor stopped M1" was displayed, and we were able to confirm the phenomenon where the motor rotation stopped.". No negative impact in state of health reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942358 ↗MalfunctionAug 13, 2026Mechanical Jam
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50236881 ↗MalfunctionAug 5, 2026Mechanical Jam
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50150328 ↗MalfunctionJul 30, 2026Mechanical Jam
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50091721 ↗MalfunctionJul 30, 2026Mechanical Jam
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50088014 ↗MalfunctionJul 27, 2026Mechanical Jam
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50017321 ↗MalfunctionJun 26, 2026Mechanical Jam
IT WAS REPORTED THAT THE DRILLCUT INTERMITTENTLY FREEZES DURING PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25641663 ↗MalfunctionJun 5, 2026Intermittent Energy Output
IT WAS REPORTED MOTOR DEFECTIVE. CONTACT KARL STORZ SERVICE, FUNCTION OUTAGES AT HIGHER SPEEDS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25416059 ↗MalfunctionFeb 4, 2026Mechanical Jam
IT WAS REPORTED THAT "WHEN THE BLADE IS CONNECTED, THE ROTATION SOMETIMES BECOMES UNSTABLE OR STOPS AT BOTH LOW AND HIGH SPEED SETTINGS." IT WAS CONFIRMED THAT NO PATIENT WAS INVOLVED OR AFFECTED BY THIS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24262161 ↗MalfunctionJan 22, 2026Output Problem
THE FOLLOWING WAS REPORTED: "THE MOTOR IS NOT ACTIVATING CONSISTENTLY WHEN THE FOOTSWITCH IS PRESSED. IN SOME INSTANCES, MULTIPLE PRESSES ARE REQUIRED TO START THE MOTOR, AND IN OTHER CASES, THE MOTOR STOPS DURING SURGERY." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2026-00093.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24157710 ↗MalfunctionJan 8, 2026Mechanical Jam
IT WAS REPORTED THAT THE MOTOR KEEPS STOPPING AND THE MESSAGE MOTOR DEFECTIVE APPEARS, FOLLOWED BY CONNECT MOTOR EVEN THOUGH THE MOTOR IS STILL CONNECTED. THE ERROR CAN BE REPRODUCED IN SHAVER BLADES MODE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24007844 ↗MalfunctionJan 8, 2026Mechanical Jam
IT WAS REPORTED THAT THE DEVICE MOTOR IS DEFECTED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24007841 ↗MalfunctionJan 7, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING CASE SHAVER GOT STUCK AND SHOWS MOTOR DEFECTIVE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23991550 ↗MalfunctionDec 29, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DEVICE STOPPED WORKING DURING USE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23919175 ↗MalfunctionDec 12, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT WHENEVER THEY SELECT THE SHAVER MODE ON THE UNIDRIVE AND PUSH THE FOOTSWITCH, THE HANDPIECE WORKS FOR 1 SECOND THEN IT STOPS WORKING. THE SCREEN IS TURNING ORANGE ON THE UNIDRIVE INDICATING THAT THE CONNECTION IS WORKING BETWEEN THE FOOTSWITCH AND THE UNIDRIVE BUT ONLY FOR A SECOND BEFORE TURNING BACK TO WHITE INDICATING THE CONNECTION IS SEVERED. THEY TRIED WITH ANOTHER NEW HANDPIECE AND IT WORKED IN GOOD CONDTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23796395 ↗MalfunctionDec 12, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THE HANDPIECE IS DETECTED BY THE MOTOR CONTROL UNIT BUT THE MOTOR DOESN'T MOVE OR MOVE AND STUCK AFTER FEW ROUNDS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23796393 ↗MalfunctionDec 8, 2025Intermittent Energy Output; Intermittent Energy Output
IT WAS REPORTED: "MOTOR WITH A LOUD SOUND AND SPORADIC STOPS. - E20 MOTOR STOPPED M2". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23752275 ↗MalfunctionNov 20, 2025Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT THE SHAVER HANDPIECE JAMMED DURING PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23606876 ↗MalfunctionNov 13, 2025Output Problem; Output Problem; Appropriate Device Problem Term/Code Not Available
THE FOLLOWING WAS REPORTED: THE MOTOR DOES NOT START UNLESS THE FOOTSWITCH IS PRESSED 2-3 TIMES. THERE IS RESISTANCE, AND IT FEELS LIKE IT GETS STUCK. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 23544055 ↗MalfunctionNov 13, 2025Output Problem; Output Problem
THE FOLLOWING WAS REPORTED: WHEN CONNECTED TO THE UM600 AND OPERATED AT A ROTATION SPEED OF 1,500 RPM (WITH THE BLADE ATTACHED), WE OBSERVED THAT THE MOTOR OCCASIONALLY FAILED TO ROTATE OR STOPPED UNEXPECTEDLY. AT THOSE TIMES, THE EVENT LOG RECORDED 'ID 116: MOTOR IS BLOCKED. A SIMILAR PHENOMENON WAS ALSO OBSERVED WHEN THE ROTATION SPEED WAS SET TO 10,000 RPM. HOWEVER, THIS ISSUE WAS NOT OBSERVED WHEN TESTING WITH OTHER HANDPIECES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 23544054 ↗MalfunctionNov 12, 2025Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT THE SHAVER HAND PIECE IS GETTING STUCK OR DID NOT START AT ALL. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 23533966 ↗MalfunctionSep 18, 2025Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT WHEN USING THE DRILLCUT-X II-35 MOTOR MAY COME TO A STANDSTILL WHEN USED IN SHAVER MODE AT SPEEDS ABOVE 8,000 RPM. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23091984 ↗
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page