n.a. 091361-06Recalls & reported problems
Karl Storz GmbH & Co. KG · FIVE S 3.5x65, sterile for single use · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04048551422519), received between Jun 13, 2023 and Sep 17, 2026 — classified as 4 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Material Separation3
- Positioning Problem2
- Detachment of Device or Device Component1
- No Display/Image1
- Optical Distortion1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information3
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50809170 ↗MalfunctionMar 11, 2026Material Separation
IT WAS REPORTED: CCOPE HAD SHAFT SEPARATION DURING PROCEDURE AND DISTAL TIP. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24569347 ↗MalfunctionMar 10, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE TIP OF THE SCOPE DETACHED AT THE DEFLECTION POINT DURING USE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24561356 ↗InjuryJul 1, 2025Positioning Problem; Positioning Problem
IT WAS REPORTED THAT DURING PROCEDURE WHILE INTUBATING THE PATIENT USING THE FIVE S, THEY DEFLECTED THE SCOPE AND COULDN'T PUT IT BACK IN THE RIGHT POSITION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. FURTHER INFORMATION HAS BEEN RECEIVED ON JUNE 24, 2025. THE FOLLOWING WAS REPORTED: DURING THE PROCEDURE, AFTER SUCCESSFUL ENTRY INTO THE TRACHEA WITH THE BRONCHOSCOPE AND ADVANCING THE ENDOTRACHEAL TUBE OVER IT, WE ATTEMPTED TO WITHDRAW THE BRONCHOSCOPE FROM THE ETT (ENDOTRACHEAL TUBE). UNFORTUNATELY, WE WERE UNABLE TO DO SO. DESPITE MULTIPLE ATTEMPTS AND MANIPULATION OF THE CONTROL LEVER IN BOTH UPWARD, DOWNWARD, AND NEUTRAL POSITIONS, THE BRONCHOSCOPE REMAINED LOCKED IN A FIXED CURVED CONFIGURATION. THIS CREATED A HIGHLY CHALLENGING SITUATION. AFTER SEVERAL FORCEFUL ATTEMPTS, WE WERE EVENTUALLY ABLE TO REMOVE THE ENTIRE UNIT, BRONCHOSCOPE AND ETT TOGETHER WITHOUT CAUSING HARM TO THE PATIENT. HOWEVER, THIS NECESSITATED THE EXTUBATION OF A DIFFICULT AIRWAY, AND WE HAD TO REINTUBATE THE PATIENT USING AN ALTERNATIVE APPROACH. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. HOWEVER, DUE TO THE REPORTED UNPLANNED REINTUBATION, THIS CASE IS DEEMED REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22373479 ↗InjuryJul 1, 2025Positioning Problem; Positioning Problem
IT WAS REPORTED THAT DURING PROCEDURE WHILE INTUBATING THE PATIENT USING THE FIVE S, THEY DEFLECTED THE SCOPE AND COULDN'T PUT IT BACK IN THE RIGHT POSITION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. FURTHER INFORMATION HAS BEEN RECEIVED ON JUNE 24, 2025. THE FOLLOWING WAS REPORTED: DURING THE PROCEDURE, AFTER SUCCESSFUL ENTRY INTO THE TRACHEA WITH THE BRONCHOSCOPE AND ADVANCING THE ENDOTRACHEAL TUBE OVER IT, WE ATTEMPTED TO WITHDRAW THE BRONCHOSCOPE FROM THE ETT (ENDOTRACHEAL TUBE). UNFORTUNATELY, WE WERE UNABLE TO DO SO. DESPITE MULTIPLE ATTEMPTS AND MANIPULATION OF THE CONTROL LEVER IN BOTH UPWARD, DOWNWARD, AND NEUTRAL POSITIONS, THE BRONCHOSCOPE REMAINED LOCKED IN A FIXED CURVED CONFIGURATION. THIS CREATED A HIGHLY CHALLENGING SITUATION. AFTER SEVERAL FORCEFUL ATTEMPTS, WE WERE EVENTUALLY ABLE TO REMOVE THE ENTIRE UNIT, BRONCHOSCOPE AND ETT TOGETHER WITHOUT CAUSING HARM TO THE PATIENT. HOWEVER, THIS NECESSITATED THE EXTUBATION OF A DIFFICULT AIRWAY, AND WE HAD TO REINTUBATE THE PATIENT USING AN ALTERNATIVE APPROACH. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. HOWEVER, DUE TO THE REPORTED UNPLANNED REINTUBATION, THIS CASE IS DEEMED REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22373477 ↗MalfunctionJul 8, 2024Break; Break
IT WAS REPORTED THAT DURING DOUBLE-LUMEN PLACEMENT PROCEDURE ON (B)(6) 2024, THE DISTAL TIP OF THE SCOPE BROKE. NO PATIENT INJURY WAS REPORTED, AND THEY WERE ABLE TO COMPLETE THE PROCEDURE USING A BACK-UP STETHOSCOPE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19697788 ↗MalfunctionJul 8, 2024Break; Break
IT WAS REPORTED THAT DURING DOUBLE-LUMEN PLACEMENT PROCEDURE ON JUNE-6-2024, THE DISTAL TIP OF THE SCOPE BROKE. NO PATIENT INJURY WAS REPORTED, AND THEY WERE ABLE TO COMPLETE THE PROCEDURE USING A BACK-UP STETHOSCOPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19697780 ↗MalfunctionMar 18, 2024No Display/Image; Optical Distortion; No Display/Image; Optical Distortion
IT WAS REPORTED ON FEBURARY 27, 2023 THAT :DURING A VETERINARY INTERVENTION, IMAGE FAILURE 5 MINUTES INTO THE PROCEDURE. NO HARM TO PATIENT, USER OR THIRD PARTIES. FURTHER INFORMATION RECEIVED ON THE MARCH 14, 2024 THAT THE BRONCHOSCOPY HAD TO BE ABORTED AND THE LUNG WASH WAS PERFORMED BLIND. DUE TO THE ADDITIONAL INFORMATION AND THE MENTIONED CHANGE OF THE SURGICAL PROCEDURE, THE RISK FOR SERIOUS DETERIORATION IN STATE OF HEALTH IS NOT NEGLIGIBLE. THUS, A REPORT IS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18926516 ↗InjuryJun 13, 2023Material Separation; Material Separation
IT WAS REPORTED THAT THE SCOPE WAS USED FOR AN AWAKE FIBRE-OPTIC INTUBATION THE CAMERA TIP BROKE OFF AND WENT MISSING DURING BRONCHOSCOPY. THE BRONCHOSCOPE WAS ALREADY IN THE TRACHEA, BUT THE ET TUBE STRUGGLED TO RAILROAD THROUGH THE MASS LESION AT THE VOCAL CORDS. THE SCOPE WAS PULLED BACK TO CONFIRM THE POSITION OF THE ET TUBE, BUT THE PICTURE WENT BLURRY AND DID NOT CLEAR WITH SUCTIONING. THE PATIENT STARTED TO DESATURATE. THE SCOPE WAS RETRACTED, AND THE VENTILATION WAS INITIATED. THE SCOPE WAS CHECKED FOR COMPLETENESS, AND IT WAS DISCOVERED THAT THE CAMERA TIP WAS ABSENT. AFTER THE PROCEDURE, THE TIP WAS FOUND ON THE FLOOR. THE INTUBATION WAS UNSUCCESSFULLY, AND A TRACHEOSTOMY NEEDS TO BE PERFORMED. THE CAUSE OF THE TRACHEOSTOMY IS UNKNOWN, BUT DUE TO THE STEPS TAKEN AND THE ADDITIONAL TREATMENT, THIS CASE WAS DEEMED REPORTABLE.
Patient problems: Insufficient Information
MAUDE report 17117788 ↗InjuryJun 13, 2023Material Separation; Material Separation
IT WAS REPORTED THAT THE SCOPE WAS USED FOR AN AWAKE FIBRE-OPTIC INTUBATION THE CAMERA TIP BROKE OFF AND WENT MISSING DURING BRONCHOSCOPY. THE BRONCHOSCOPE WAS ALREADY IN THE TRACHEA, BUT THE ET TUBE STRUGGLED TO RAILROAD THROUGH THE MASS LESION AT THE VOCAL CORDS. THE SCOPE WAS PULLED BACK TO CONFIRM THE POSITION OF THE ET TUBE, BUT THE PICTURE WENT BLURRY AND DID NOT CLEAR WITH SUCTIONING. THE PATIENT STARTED TO DESATURATE. THE SCOPE WAS RETRACTED, AND THE VENTILATION WAS INITIATED. THE SCOPE WAS CHECKED FOR COMPLETENESS, AND IT WAS DISCOVERED THAT THE CAMERA TIP WAS ABSENT. AFTER THE PROCEDURE, THE TIP WAS FOUND ON THE FLOOR. THE INTUBATION WAS UNSUCCESSFULLY, AND A TRACHEOSTOMY NEEDS TO BE PERFORMED. THE CAUSE OF THE TRACHEOSTOMY IS UNKNOWN, BUT DUE TO THE STEPS TAKEN AND THE ADDITIONAL TREATMENT, THIS CASE WAS DEEMED REPORTABLE.
Patient problems: Insufficient Information
MAUDE report 17117787 ↗
Other LARYNGOSCOPE, NON-RIGID devices
- NA 091361-01Karl Storz GmbH & Co. KG · LARYNGOSCOPE, NON-RIGID2 reports
- C-MAC 11301BNXKarl Storz GmbH & Co. KG · LARYNGOSCOPE, NON-RIGID2 reports
- n.a. 11303BNXKarl Storz GmbH & Co. KG · LARYNGOSCOPE, NON-RIGID2 reports
- n.a. 11301BN1Karl Storz GmbH & Co. KG · LARYNGOSCOPE, NON-RIGID2 reports
- n.a. 11302BD2Karl Storz GmbH & Co. KG · LARYNGOSCOPE, NON-RIGID1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page