Skip to content

TI500 Globetrotter Neonatal Transport System MU20508Recalls & reported problems

Draeger Medical Systems, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04049098000192), received between Oct 28, 2024 and Oct 28, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Key or Button Unresponsive/not Working” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K001019); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Key or Button Unresponsive/not Working1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED BY THE CUSTOMER THAT ON (B)(6) 2024 THE TRANSPORT INCUBATOR 500 (TI500) GLOBE TROTTER HAD AN ISSUE WITH THE TEMPERATURE ADJUST BUTTON AND THE VENTILATOR HAD AN ISSUE MAINTAINING PRESSURE. THERE WAS NO ADVERSE PATIENT EFFECTS REPORTED. THE DEVICE WAS INSPECTED AND SERVICED BY DRAEGER ENGINEERING AND FOUND THAT THE REPORTED PROBLEMS ONLY OCCURRED INTERMITTENTLY. THE ISSUE RELATED TO THE UNRESPONSIVE TEMPERATURE ADJUST BUTTON WAS CONFIRMED DURING DRAEGER'S INSPECTION OF THE DEVICE. DRAGER SERVICE FOLLOWED SERVICE SOFTWARE FOR REPAIR INSTRUCTIONS AND REPLACED THE TEMPERATURE CONTROLLER TO RESOLVE THE ISSUE. THE CUSTOMER'S CONCERN WITH THE VENTILATOR NOT HOLDING PRESSURE WAS NOT ABLE TO BE REPRODUCED DURING SERVICE. THE DEVICE WAS TESTED AND CONFIRMED TO HAVE BEEN OPERATING PER THE MANUFACTURER'S SPECIFICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20545945 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001019. One submission often covers many models, so check the product and lot details in each record.

  • Z-1053-2008 · Jan 31, 2008 · Terminated
    Draeger Globe Trotter IC, Catalog Number: MU20509, Draeger Medical, Inc., Telford, PA 18969
  • Z-1052-2008 · Jan 31, 2008 · Terminated
    Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA 18969