Infinity Vista XL Monitor MS18986Recalls & reported problems
Draeger Medical Systems, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04049098003858), received between Jan 12, 2022 and Aug 1, 2024 — classified as 1 death, 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
6 other recalls cover devices cleared under the same FDA submission (K070566); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
- Device Fell1
- Fire1
- Incorrect, Inadequate or Imprecise Result or Readings1
- No Audible Prompt/Feedback1
- Unable to Obtain Readings1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Cyanosis1
- Insufficient Information1
- Low Oxygen Saturation1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 1, 2024Unable to Obtain Readings; Unable to Obtain Readings
THE CUSTOMER REPORTED THAT ON (B)(6) 2024, AFTER THE PATIENT'S SURGERY AND EXTUBATION, THE OXYGEN SATURATION READING SHOWED SIGNIFICANT INTERFERENCE AND THE DEVICE REPORTEDLY DID NOT TRIGGER ANY ALARMS. THE DOCTOR NOTICED CYANOSIS AROUND THE PATIENT¿S MOUTH AND LIPS, INTERVENED PROMPTLY BY PROVIDING ASSISTED BREATHING TREATMENT, AND THE PATIENT'S CONDITION IMPROVED IMMEDIATELY. THE ISSUE WAS IDENTIFIED IN TIME AND DID NOT RESULT IN SIGNIFICANT HARM TO THE PATIENT. IT WAS REPORTED THAT THE SPRING OF THE PULSE OXIMETER HAD MALFUNCTIONED.
Patient problems: Cyanosis
MAUDE report 19883125 ↗MalfunctionAug 30, 2023Crack; Incorrect, Inadequate or Imprecise Result or Readings; Device Fell; Crack; Incorrect, Inadequate or Imprecise Result or Readings; Device Fell
IT WAS REPORTED THAT TWO PATIENT MONITORING SYSTEMS THAT WERE MOUNTED TO THE WALL WITH A DEDICATED HOLDER FELL DOWN TO THE FLOOR. THERE WAS NO DETAIL IN THE REPORT WHICH WOULD REASONABLY SUGGEST THAT A PATIENT, USER OR THIRD PERSON MAY HAVE BEEN HURT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Low Oxygen Saturation
MAUDE report 17649557 ↗DeathSep 12, 2022No Audible Prompt/Feedback; No Audible Prompt/Feedback
ON 5 SEP 2022 A COMPLAINT WAS RECEIVED REGARDING AN EVENT WHERE THE INVOLVED PATIENT DIED. THERE WAS AN ALLEGED REPORT OF A MALFUNCTION INITIALLY, BUT IT WAS VERIFIED THAT ALARMS WERE MANUALLY TURNED OFF FOR THE EVENT. ON 9 SEP 2022 NEW INFORMATION WAS RECEIVED STATING THAT THE CUSTOMER WAS ALLEGING THAT THE DEVICE NOT TRIGGERING AN ALARM DURING THE EVENT PLAYED A PART IN THE PATIENT OUTCOME.
Patient problems: Insufficient Information
MAUDE report 15405519 ↗MalfunctionJan 12, 2022Fire; Fire
IT WAS REPORTED THAT A FIRE INCIDENT OCCURRED WITH A DRAEGER VISTA XL MONITOR BEING USED WITH A COUNTERFEIT DRAEGER BATTERY (WITH DRÄGER LOGO STAMPED ON THE LABEL) THAT WAS INSTALLED BY A NON AUTHORIZED 3RD PARTY. NO ADVERSE PATIENT IMPACT WAS REPORTED. NOTE THAT THE FOLLOWING IS PROVIDED IN THE VISTA XL INSTRUCTIONS FOR USE: WARNING ONLY THE ITEMS INDICATED ON THE LIST OF ACCESSORIES IN THE ¿ACCESSORIES¿ CHAPTER HAVE BEEN TESTED AND APPROVED TO BE USED WITH THE DEVICE. ACCORDINGLY IT IS STRONGLY RECOMMENDED THAT ONLY THESE ACCESSORIES BE USED IN CONJUNCTION WITH THE SPECIFIC DEVICE. OTHERWISE THE CORRECT FUNCTIONING OF THE DEVICE MAY BE COMPROMISED. CAUTION: IT IS STRONGLY RECOMMENDED THAT YOU USE BATTERIES THAT ARE PROVIDED BY DRÄGER MEDICAL. THE USE OF NON-APPROVED BATTERIES MAY DAMAGE THE DEVICE. NOTE: CABLES AND ACCESSORIES NOT SPECIFIED WITHIN THE INSTRUCTIONS FOR USE ARE NOT AUTHORIZED. USING OTHER CABLES AND/OR ACCESSORIES MAY ADVERSELY IMPACT SAFETY, PERFORMANCE AND ELECTROMAGNETIC COMPATIBILITY (INCREASED EMISSION AND DECREASED IMMUNITY). UNDER APPROVED ACCESSORIES: EXTERNAL BATTERY PACK (SEALED LEAD ACID) 5592097.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13238574 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070566. One submission often covers many models, so check the product and lot details in each record.
- Z-0281-2012 · Oct 31, 2011 · TerminatedDraeger Infinity(R) Delta, Delta XL, Gamma XXL and Kappa Monitors are used to monitor, physiologically, the vital signs of patients (with Arrhythmia Detection or Alarms).
- Z-3047-2011 · Jul 29, 2011 · TerminatedInfinity Delta Portable Patient Monitor. Draeger Medical Systems, Inc. Portable monitor used to monitor vital signs of neonatal, pediatric, and adult patients for bedside and tr
- Z-3048-2011 · Jul 29, 2011 · TerminatedInfinity Gamma X XL Portable Patient Monitor. Draeger Medical Systems, Inc. Portable monitor used to monitor vital signs of neonatal, pediatric, and adult patients for bedside a
- Z-2283-2010 · Jun 4, 2010 · TerminatedInfinity Delta/Gamma X XL Patient Monitor. Portable monitor used to monitor vital signs of neonatal, pediatric, and adult patients for bedside and transport applications.
- Z-2284-2010 · Jun 4, 2010 · TerminatedInfinity Delta/Gamma X XL Patient Monitor. Portable monitor used to monitor vital signs of neonatal, pediatric, and adult patients for bedside and transport applications.
- Z-0582-2010 · Nov 9, 2009 · TerminatedDelta/Delta XL Monitors with Omega S Package
Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page