INFINITY M500 WITH POWER MS20407Recalls & reported problems
Draeger Medical Systems, Inc. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04049098054447), received between Jun 28, 2024 and Sep 16, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Electrical Shorting” (1 report).
3 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 13, 2024Open, Classified
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.
Z-2189-2024 · Root cause: Device Design · FDA record ↗
- Mar 28, 2018Terminated
Software anomaly resulting in the loss of patient settings and stored patient data.
Z-1813-2018 · Root cause: Under Investigation by firm · FDA record ↗
- Feb 21, 2018Terminated
The invasive pressure visual and audible alarms may not activate.
Z-1382-2018 · Root cause: Software Manufacturing/Software Deployment · FDA record ↗
Most reported problems
Device problems
- Electrical Shorting1
- Inappropriate Audible Prompt/Feedback1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2024Electrical Shorting; Insufficient Device Problem Information; Electrical Shorting; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT THE KEYBOARD CONNECTION CABLE WAS "SCORCHED" NEAR WHERE THE CABLE MEETS THE KEYBOARD. THE PICTURES PROVIDED CONFIRMED THE DAMAGE TO THE KEYBOARD CABLE. IT WAS REPORTED THAT NO ENTRY COULD BE MADE ON THE KEYBOARD RIGHT BEFORE THE DAMAGE TO THE CABLE WAS NOTICED. IT WAS ALSO CONFIRMED THERE WAS NO INJURY TO THE USER OR A PATIENT AND NO IMPACT ON ANY OTHER FUNCTION OF ICS MONITORING. THERE WAS NO ADVERSE PATIENT OR USER IMPACT OR DEATH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20225077 ↗MalfunctionJun 28, 2024Inappropriate Audible Prompt/Feedback; Inappropriate Audible Prompt/Feedback
THE CUSTOMER REPORTED THAT THERE WERE SEVERAL ALARMS BUT NO EVENTS SAVED. ALL CURVES ON THE ENTIRE CENTRAL MONITORING UNIT ARE CLOSED. CANNOT MAKE SETTINGS OR CLOSE ALARMS (FOR EXAMPLE). THE CENTRAL MONITORING UNIT STILL HAS A CURRENT DISPLAY OF PARAMETERS AND WAS REBOOTED AND CURRENTLY WORKS. THERE WAS NO REPORT OF ADVERSE EVENT OR PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19632840 ↗
Other Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.