LifeStar® Vascular & Biliary Stent VIUS10060Recalls & reported problems

Bard Peripheral Vascular, Inc. · LifeStar™ Vascular and Biliary Stent System 10 mm x 60 mm (80 cm delivery catheter) · Class 3 · Implant

All models of LifeStar Vascular & Biliary Stent (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04049519005553), received between Mar 21, 2017 and Jun 24, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Component Misassembled” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Misassembled2
  • Device Dislodged or Dislocated2
  • Activation Failure1
  • Difficult or Delayed Positioning1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2025Device Dislodged or Dislocated; Component Misassembled; Device Dislodged or Dislocated; Component Misassembled

    ON (B)(6) 2025, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING THE LIFESTAR VASCULAR STENT. DURING THE PROCEDURE, IT CAME TO NOTICE THAT THE RADIOPAQUE MARKER WAS NOT IN THE CORRECT LOCATION ON THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22319029 ↗
  • MalfunctionJun 24, 2025Device Dislodged or Dislocated; Component Misassembled; Device Dislodged or Dislocated; Component Misassembled

    ON (B)(6) 2025, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING THE LIFESTAR VASCULAR STENT. DURING THE PROCEDURE, IT CAME TO NOTICE THAT THE RADIOPAQUE MARKER WAS NOT IN THE CORRECT LOCATION ON THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22319000 ↗
  • MalfunctionApr 26, 2017Difficult or Delayed Positioning; Difficult or Delayed Positioning

    IT WAS REPORTED THAT DURING A STENTING PROCEDURE FOR TREATMENT OF A PRE-DILATED UPPER AXILLARY A/V GRAFT STENOSIS, IT WAS DIFFICULT TO DEPLOY THE STENT. DURING THE ATTEMPT TO RETRACT THE Y ADAPTER RESISTANCE WAS EXPERIENCED; HOWEVER, THE STENT COULD BE DEPLOYED SUCCESSFULLY AT THE TARGET SITE WITHOUT FURTHER ISSUES. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6522362 ↗
  • InjuryMar 21, 2017Activation Failure; Activation Failure

    IT WAS REPORTED THAT ONE END OF THE STENT DID NOT SELF-EXPAND AFTER DEPLOYMENT IN THE BRACHIAL VEIN FOR TREATMENT OF A STENOSIS IN AN A/V FISTULA. A BALLOON CATHETER WAS USED TO EXPAND THE STENT COMPLETELY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6422228 ↗