Dornier MedTech 270 micron Holmium Laser Fiber Cable, Reusable, RFIDRecalls & reported problems
Dornier Medtech America, Inc. · 270 micron Holmium Laser Fiber Cable, Reusable, RFID · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 04049958003158), received between Oct 22, 2019 and Dec 12, 2025 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event10
- Fracture4
- Connection Problem2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient2
- Burn(s)1
- Insufficient Information1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2025Break; Break
DMTA RECEIVED A REPORT THAT A PHYSICIAN OBSERVED A BREAK IN THE HOLMIUM FIBER UPON REMOVING IT FROM THE PACKAGING, PRIOR TO USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23794561 ↗MalfunctionSep 10, 2024No Apparent Adverse Event; No Apparent Adverse Event
DORNIER RECEIVED COMPLAINT INFORMATION REGARDING A LASER FIBER WHICH HAD REPORTEDLY BURNED. THE CUSTOMER STATED, "WHEN PLACED IN THE LASER, IT SMELLED LIKE BURNING".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20186101 ↗MalfunctionApr 23, 2024No Apparent Adverse Event; No Apparent Adverse Event
HOLMIUM FIBER REPORTEDLY DEFECTIVE AND OBSERVED WITH THE "TOP OF THE SMA PIN CHARRED", POSSIBLY INDICATING A FIBER BURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19166430 ↗MalfunctionSep 18, 2023No Apparent Adverse Event; No Apparent Adverse Event
A NOTIFICATION WAS RECEIVED REGARDING A HOLMIUM FIBER UNIT WHICH WAS REPORTED TO HAVE BROKEN DURING USAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17764021 ↗MalfunctionSep 12, 2023No Apparent Adverse Event; No Apparent Adverse Event
HOLMIUM FIBER REPORTEDLY OVERHEATED/BURNED DURING USAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17732680 ↗MalfunctionAug 23, 2023No Apparent Adverse Event; No Apparent Adverse Event
TWO HOLMIUM LASER FIBERS REPORTEDLY BURNED DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17607281 ↗MalfunctionApr 2, 2021No Apparent Adverse Event; No Apparent Adverse Event
A NOTIFICATION WAS RECEIVED REGARDING A HOLMIUM FIBER UNIT WHICH WAS REPORTED TO HAVE BURNED DURING USE. NO PATIENT HARM OR IMPACT WAS REPORTED. ADDITIONAL COMPLAINT EXPERIENCES WERE REPORTED BY THE SAME CUSTOMER, LASER DEVICES NOT RECOGNIZING FIBERS, FIBERS NOT TRANSMITTING POWER PROPERLY, LOOSE CONNECTORS, FIBERS NOT WORKING. A TOTAL OF 7 UNITS WERE REPORTED FOR THREE SEPARATE PRODUCTION LOTS OF FIBERS UNDER TWO DIFFERENT PRODUCT ID'S. K2013151, LOT F0120R - 3 UNITS. K2013152 LOT B1018R - 3 UNITS. K2013151 LOT F0320R - 1 UNIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11609281 ↗MalfunctionMar 23, 2021Fracture; Fracture
COMPLAINT INFORMATION WAS RECEIVED FOR A HOLIUM FIBER WHICH WAS IDENTIFIED TO HAVE BROKE. NO REPORTS OF INJURY OR PATIENT COMPLICATION WERE REPORTED TO DORNIER MEDTECH AMERICA DUE TO PRODUCT COMPLAINT. THIS COMPLAINT WAS INITIALLY REPORTED TO FDA UNDER 1037955-2020-00040 ON 09/14/2020.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11549586 ↗MalfunctionMar 22, 2021Fracture; Fracture
COMPLAINT INFORMATION WAS RECEIVED FOR A HOLMIUM FIBER WHICH WAS IDENTIFIED TO HAVE BROKE INSIDE THE PATIENT DURING OPERATION. NO REPORTS OF INJURY OR PATIENT COMPLICATION WERE REPORTED TO DORNIER MEDTECH AMERICA DUE TO THIS PRODUCT COMPLAINT. THIS COMPLAINT WAS INITIALLY REPORTED TO FDA UNDER 1037955-2020-00036 ON (B)(6) 2020.
Patient problems: Insufficient Information
MAUDE report 11541910 ↗MalfunctionMar 15, 2021Connection Problem; No Apparent Adverse Event; Connection Problem; No Apparent Adverse Event
A NOTIFICATION WAS RECEIVED REGARDING HOLMIUM FIBERS REPORTED TO HAVE BURNED. THE COMPLAINT REPORTED "4 THERAPEUTIC OPTIVAL FIBERS WERE BURNED, BATCH F5019R". NO PATIENT HARM WAS REPORTED. THIS COMPLAINT WAS INITIALLY REPORTED TO FDA UNDER 1037955-2020-00019 ON (B)(6) 2020.
Patient problems: Burn(s)
MAUDE report 11483390 ↗MalfunctionOct 14, 2020Fracture; Fracture
COMPLAINT INFORMATION WAS RECEIVED FOR A HOLMIUM FIBER WHICH WAS IDENTIFIED TO HAVE BROKE. NO REPORTS OF INJURY OR PATIENT COMPLICATION WERE REPORTED TO DORNIER MEDTECH AMERICA DUE TO THIS PRODUCT COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10680722 ↗MalfunctionOct 8, 2020No Apparent Adverse Event; No Apparent Adverse Event
A NOTIFICATION WAS RECEIVED REGARDING HOLMIUM FIBER UNITS (5 UNITS), WHICH WERE REPORTED TO HAVE BURNED DURING USE WITH 2 CUSTOMER LASERS. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10651859 ↗MalfunctionOct 1, 2020Fracture; Fracture
COMPLAINT INFORMATION WAS RECEIVED FOR A HOLMIUM FIBER WHICH WAS IDENTIFIED TO HAVE BROKE INSIDE THE PATIENT DURING OPERATION. NO REPORTS OF INJURY OR PATIENT COMPLICATION WERE REPORTED TO DORNIER MEDTECH AMERICA DUE TO THIS PRODUCT COMPLAINT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10616136 ↗MalfunctionJul 8, 2020Connection Problem; No Apparent Adverse Event; Connection Problem; No Apparent Adverse Event
A NOTIFICATION WAS RECEIVED REGARDING HOLMIUM FIBERS REPORTED TO HAVE BURNED. THE COMPLAINANT REPORTED, "4 THERAPEUTIC OPTICAL FIBES WERE BURNED, BATCH F5019R". NO PATIENT HARM WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10246226 ↗MalfunctionOct 22, 2019No Apparent Adverse Event; No Apparent Adverse Event
COMPLAINT DETAILS PROVIDED IN THE INITIAL NOTIFICATION STATED, "WE RECEIVED A CUSTOMER COMPLAINT ABOUT BURNING OF THE NEW 270 MICROMETER REUSABLE FIBERS. THEY HAVE DAMAGED 3 UNUSED FIBERS".
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9220729 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page