uCentum™ comprehensive posterior system CS 3804-065-045Recalls & reported problems
Ulrich GmbH & Co. KG · Screw, pre-fixation, Ø 6.5 mm, length 45 mm perforated · Class 2 · Implant
All models of uCentum comprehensive posterior system (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04052536079551), received between May 15, 2018 and Dec 6, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K123717); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component2
Patient problems
- Failure of Implant3
- Discomfort2
- Pain2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 6, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORT: DURING THE IMPLANTATION OF A PEDICLE SCREW, AN UNNOTICED DISLOCATION OF THE INLAY OF THE SCREW OCCURRED WITHOUT ANY APPLICATION OF FORCE. A LONGITUDINAL ROD WAS ABLE TO BE PROPERLY BLOCKED WITH THE HEAD SCREW WITHOUT ANY PROBLEMS. THE DISLOCATED PART WAS NOT NOTICEABLE IN THE INTRAOPERATIVE IMAGE. THE DISLOCATED PART OF THE SCREW COULD ONLY BE DETECTED IN THE POSTOPERATIVE X-RAY AND THE PATIENT HAD TO BE REVISED. THE PATIENT INSISTED ON RETAINING THE SCREW.
Patient problems: Failure of Implant
MAUDE report 20864232 ↗InjuryDec 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORT: DURING THE IMPLANTATION OF A PEDICLE SCREW, AN UNNOTICED DISLOCATION OF THE INLAY OF THE SCREW OCCURRED WITHOUT ANY APPLICATION OF FORCE. A LONGITUDINAL ROD WAS ABLE TO BE PROPERLY BLOCKED WITH THE HEAD SCREW WITHOUT ANY PROBLEMS. THE DISLOCATED PART WAS NOT NOTICEABLE IN THE INTRAOPERATIVE IMAGE. THE DISLOCATED PART OF THE SCREW COULD ONLY BE DETECTED IN THE POSTOPERATIVE X-RAY AND THE PATIENT HAD TO BE REVISED. THE PATIENT INSISTED ON RETAINING THE SCREW AFTER THE REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 20858238 ↗InjuryJun 21, 2018Break; Break
TWO PRE-FIXATION SCREWS IN S1 BROKE (ONE DIRECTILY UNDERNEATH THE SCREW HEAD AND THE OTHER AT THE MIDDLE OF THE THREAD). A REVISION WAS DONE TO REPLACE THE BROKEN SCREWS.
Patient problems: Discomfort; Failure of Implant; Pain
MAUDE report 7626966 ↗InjuryMay 15, 2018Break; Break
PRE-FIXATION SCREW IN S1 BROKE DIRECTLY UNDERNEATH THE SCREW HEAD AND HAD TO BE REPLACED.
Patient problems: Discomfort; Pain
MAUDE report 7516478 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123717. One submission often covers many models, so check the product and lot details in each record.
- Z-1126-2019 · Mar 8, 2019 · TerminatedULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page