POWER DRIVE ART1 2304007Recalls & reported problems
Richard Wolf GmbH · MOTOR CONTROL UNIT 2304 for Orthopedics, Spine surgery and Bronchoscopy, for connection of handpieces M4, M5/0, M5/3, S1 and M1, connection for two handpieces and a footswitch, 5.7 inch monochrome Display, compatible with core, U: 120VAC, 50/60Hz, Dim. (wxhxd): 330x155x365mm · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04055207012246), received between Feb 16, 2022 and Feb 18, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Device Displays Incorrect Message2
- Display or Visual Feedback Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 18, 2022Display or Visual Feedback Problem; Device Displays Incorrect Message; Appropriate Device Problem Term/Code Not Available; Display or Visual Feedback Problem; Device Displays Incorrect Message; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO RICHARD WOLF: "(B)(6) CALLED (B)(4), THE LOCAL REP, DURING THE PROCEDURE. WHEN THE FOOTSWITCH FOR THE POWERDRIVE ART1 WAS ACTIVATED, THE CONTROLLER DISPLAYED A "FOOTPEDAL DEFECTIVE" ERROR MESSAGE. THE REP ASKED THEM TO DISCONNECT AND RE-CONNECT THE FOOTSWITCH, AND THE MESSAGE WAS NO LONGER DISPLAYED. WHEN THEY STARTED THE PROCEDURE, THE HANDPIECE WASN'T ROTATING THE BURR. THEY DISCONNECTED THE CABLE FROM THE HANDPIECE AND GENERATOR, AND RE-CONNECTED. THE HANDPIECE STILL WOULD NOT ROTATE THE BURR. THEN THE REP CALLED ME TO CONFERENCE WITH THE OPERATING ROOM, AND THAT'S WHEN THE PHYSICIAN STOPPED THE CASE. HE HAD MADE AN INCISION, STOPPED THE CASE, CLOSED THE INCISION, AND WILL RE-SCHEDULE. CHRIS, THE LOCAL SALES REP WENT TO THE ACCOUNT AND DETERMINED THAT THE (2) 89955.0000 M4 MOTOR HANDPIECES AND THE 8564.851 CONNECTING CABLE WERE FUNCTIONING PROPERLY, AND WILL NOT BE RETURNED TO US FOR EVALUATION." WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE ANY INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED IS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13559175 ↗InjuryFeb 16, 2022Display or Visual Feedback Problem; Device Displays Incorrect Message; Appropriate Device Problem Term/Code Not Available; Display or Visual Feedback Problem; Device Displays Incorrect Message; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO RICHARD WOLF: "(B)(6) CALLED (B)(6) , THE LOCAL REP, DURING THE PROCEDURE. WHEN THE FOOTSWITCH FOR THE POWERDRIVE ART1 WAS ACTIVATED, THE CONTROLLER DISPLAYED A "FOOTPEDAL DEFECTIVE" ERROR MESSAGE. THE REP ASKED THEM TO DISCONNECT AND RE-CONNECT THE FOOTSWITCH, AND THE MESSAGE WAS NO LONGER DISPLAYED. WHEN THEY STARTED THE PROCEDURE, THE HANDPIECE WASN'T ROTATING THE BURR. THEY DISCONNECTED THE CABLE FROM THE HANDPIECE AND GENERATOR, AND RE-CONNECTED. THE HANDPIECE STILL WOULD NOT ROTATE THE BURR. THEN THE REP CALLED ME TO CONFERENCE WITH THE OPERATING ROOM, AND THAT'S WHEN THE PHYSICIAN STOPPED THE CASE. HE HAD MADE AN INCISION, STOPPED THE CASE, CLOSED THE INCISION, AND WILL RE-SCHEDULE. (B)(6) , THE LOCAL SALES REP WENT TO THE ACCOUNT AND DETERMINED THAT THE (2) 89955.0000 M4 MOTOR HANDPIECES AND THE 8564.851 CONNECTING CABLE WERE FUNCTIONING PROPERLY, AND WILL NOT BE RETURNED TO US FOR EVALUATION." WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE ANY INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13542837 ↗
Other MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices
- NA 05.001.201Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED1,902 reports
- NA 530.710Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED1,800 reports
- NA 530.705Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED1,492 reports
- NA 530.715Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED390 reports
- NA 530.730Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED267 reports
- NA 05.001.240Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED262 reports
- KINCISE™ 1000-00-101Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED242 reports
- NA 530.791Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED84 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page